510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

IL Test Heparin

K980242510(k) TraditionalCleared Mar 19, 199855-day reviewClass II21 CFR 864.7525Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Instrumentation Laboratory CO · KFF Assay, Heparin

Cleared on Mar 19, 1998 as substantially equivalent to K935212 (Instrumentation Laboratory CO, 1994) after a 55-day review. Its family tree reaches back 1 generation to 1994. 7 later devices descend from it. It has been recalled once.

Decided
Mar 19, 1998
Review
55 days
Class
Class II
Product code
KFF
Recalls
1
Predicates
1

Its longest line of descent runs back 1 generation to K935212 (1994). 7 later devices descend from it, 2 citing it directly.

Descendants, by year cleared 2009: 1 descendant cleared1998: 1 descendant cleared2022: 1 descendant cleared2017: 2 descendants cleared2023: 1 descendant cleared2024: 1 descendant clearedK935212 · IL Test Heparin (XA), PN 84699-15 · Instrumentation Laboratory CO · 19941990200020102020
Descendants, by generation 1 generation forward: 2 descendants2 generations forward: 3 descendants3 generations forward: 2 descendants K935212 · IL Test Heparin (XA), PN 84699-15 · Instrumentation Laboratory CO · 19940
  1. Root · 1 generation backK935212 IL Test Heparin (XA), PN 84699-15Instrumentation Laboratory CO · 1994
  2. This deviceK980242 IL Test HeparinInstrumentation Laboratory CO · 1998
  3. Descendant · cites it directly · 4 of its ownK090209 Hemosil Liquid Heparin, Heparin Calibrators and LMW and UF Heparin…Instrumentation Laboratory CO · 2009 · Recalled
  4. Descendant · 2 generations on · 2 of its ownK213464 HemosIL Liquid Anti-XaInstrumentation Laboratory CO · 2022 · Recalled
  5. Descendant · 3 generations on · 1 of its ownK223187 HemosIL Liquid Anti-XaInstrumentation Laboratory CO · 2023
  6. Descendant · cites it directly · 1 of its ownK983178 Coamatic HeparinInstrumentation Laboratory CO · 1998
  7. Descendant · 3 generations on · 0 of its ownK240315 INNOVANCE Anti-Xa (OPPU05); INNOVANCE Apixaban Standards (OPPW05)…Siemens Healthcare Diagnostics Products GmbH · 2024

Showing 5 of 7 descendants. See all descendants

Descendants cleared per year

1998: 12009: 12017: 22022: 12023: 12024: 1
1998high 2 (2017) · low 0 (1999)2026

Most influential descendants

Showing 5 of 7 · See all
  • K090209Hemosil Liquid Heparin, Heparin Calibrators and LMW and UF Heparin ControlsInstrumentation Laboratory CO · 2009 · cites it directly4descendants
  • K213464HemosIL Liquid Anti-XaInstrumentation Laboratory CO · 2022 · 2 generations on2descendants
  • K223187HemosIL Liquid Anti-XaInstrumentation Laboratory CO · 2023 · 3 generations on1descendant
  • K983178Coamatic HeparinInstrumentation Laboratory CO · 1998 · cites it directly1descendant
  • K240315INNOVANCE Anti-Xa (OPPU05); INNOVANCE Apixaban Standards (OPPW05); INNOVANCE Apixaban…Siemens Healthcare Diagnostics Products GmbH · 2024 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K935212IL Test Heparin (XA), PN 84699-15Instrumentation Laboratory CO1994—
0 (this device)K980242IL Test HeparinInstrumentation Laboratory CO1998—
+1K090209Hemosil Liquid Heparin, Heparin Calibrators and LMW and UF Heparin ControlsInstrumentation Laboratory CO20095
+2K213464HemosIL Liquid Anti-XaInstrumentation Laboratory CO20222
+3K223187HemosIL Liquid Anti-XaInstrumentation Laboratory CO2023—
+1K983178Coamatic HeparinInstrumentation Laboratory CO1998—
+3K240315INNOVANCE Anti-Xa (OPPU05); INNOVANCE Apixaban Standards (OPPW05); INNOVANCE…Siemens Healthcare Diagnostics Products GmbH2024—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
199810
200910
201720
202210
202310
202410
FIG. 1 — Lineage of K980242, 1994–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KFF? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KFF device

  1. #1 K260214 STA® - Liquid Anti-Xascore 12.4 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K935212IL Test Heparin (XA), PN 84699-15
Summary page 1
…ry 22, 1998 Name of the device: IL Test™ Heparin Classification name(s): 864.7525 Heparin Assay Class II 81KFF Assay, Heparin Identification of predicate device(s): K935212 IL Test™ Heparin Xa Description of the device/intended use(s): IL Test™ Heparin is an in vitro diagnostic test for the quantit…
Instrumentation Laboratory CO1994—

What descends from it? (7)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K090209Hemosil Liquid Heparin, Heparin Calibrators and LMW and UF Heparin ControlsInstrumentation Laboratory CO20094
K983178Coamatic HeparinInstrumentation Laboratory CO19981

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 3, 2025Class IIZ-2323-2025Process controlMultiple complaints indicating lower than expected quality control (QC) results and in some instances QC recovery below the lower limit of the package insert acceptance range when testing is performed with HemosIL Liquid Anti-Xa and…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.
  • Of its descendants, K090209 (Instrumentation Laboratory CO, 2009) has gone furthest, with 4 descendants of its own.

How does its category compare?

All KFF devices

KFF Assay, Heparin: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · KFF

1in 2022

2009: 12011: 12017: 12022: 12026 (partial year): 1
2007high 1 (2009) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KFF

Too few KFF 510(k)s in any year since 2007 to give a median.

Recalls per year · KFF

1in 2024

2010: 12021: 12024: 1
2007high 1 (2010) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KFF

7in 2025

2007: 12013: 42014: 32017: 42018: 52019: 12021: 42022: 12025: 7
2007high 7 (2025) · low 0 (2008)2025

MAUDE reports naming product code KFF, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—01
20080—00
20091—00
20100—10
20111—00
20120—00
20130—04
20140—03
20150—00
20160—00
20171—04
20180—05
20190—01
20200—00
20210—14
20221—01
20230—00
20240—10
20250—07
20261—0—

What is it indicated for?

IL Test™ Heparin is an in vitro diagnostic test for the quantitative determination of

unfractionated heparin (UFH) and low molecular weight heparin (LMWH) activity in human

plasma. It is an automated heparin assay based on a synthetic chromogenic substrate and Factor

Xa inactivation.

Heparin is the most frequently used antithrombotic drug. The biological activity of this

sulphated glycosaminoglycan resides in its ability to accelerate (up to 2000-fold) the inhibitory

effect of antithrombin on coagulation proteases. In recent years, it has been shown that LMWH,

besides being as useful therapeutically as UFH, also has a longer half-life.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K980242: IL Test Heparin.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K980242

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026