510(k) · Cleared
Recalled · Class II · 2022Schneider Guider Softip Guiding Catheters
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Boston Scientific Scimed, Inc. · DQY Catheter, Percutaneous
Cleared on Aug 11, 1998 as substantially equivalent to K961999 (Boston Scientific Scimed, Inc., 1996) after a 187-day review. Its family tree reaches back 1 generation to 1996. 26 later devices descend from it. It has been recalled 20 times.
- Decided
- Aug 11, 1998
- Review
- 187 days
- Class
- Class II
- Product code
- DQY
- Recalls
- 20
- Predicates
- 1
Its longest line of descent runs back 1 generation to K961999 (1996). 26 later devices descend from it, 3 citing it directly.
- Root · 1 generation backK961999 Schneider Guider Softip Guiding CathetersBoston Scientific Scimed, Inc. · 1996
- This deviceK980453 Schneider Guider Softip Guiding CathetersBoston Scientific Scimed, Inc. · 1998
- Descendant · cites it directly · 21 of its ownK071226 8FR Turbo Booster, Guide Catheter, 7FR Turbo Booster, Model 518-063…Spectranetics Corp. · 2007 · Recalled
- Descendant · 2 generations on · 20 of its ownK091299 Spectranetics Turbo-Tandem SystemSpectranetics Corp. · 2009
- Descendant · 3 generations on · 19 of its ownK094036 Turbo-Tandem System Model 472-110Spectranetics Corp. · 2010
- Descendant · 3 generations on · 17 of its ownK140775 Turbo-Elite Atherectomy Catheters, Turbo Tandem SystemSpectranetics Corporation · 2014 · Recalled
- Descendant · 4 generations on · 8 of its ownK181642 B-Laser Atherectomy SystemEximo Medical, Ltd. · 2018
Showing 5 of 26 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 26 · See all- K0712268FR Turbo Booster, Guide Catheter, 7FR Turbo Booster, Model 518-063, 518-043Spectranetics Corp. · 2007 · cites it directly21descendants
- K091299Spectranetics Turbo-Tandem SystemSpectranetics Corp. · 2009 · 2 generations on20descendants
- K094036Turbo-Tandem System Model 472-110Spectranetics Corp. · 2010 · 3 generations on19descendants
- K140775Turbo-Elite Atherectomy Catheters, Turbo Tandem SystemSpectranetics Corporation · 2014 · 3 generations on17descendants
- K181642B-Laser Atherectomy SystemEximo Medical, Ltd. · 2018 · 4 generations on8descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K961999 | Schneider Guider Softip Guiding Catheters | Boston Scientific Scimed, Inc. | 1996 | — |
| 0 (this device) | K980453 | Schneider Guider Softip Guiding Catheters | Boston Scientific Scimed, Inc. | 1998 | — |
| +1 | K071226 | 8FR Turbo Booster, Guide Catheter, 7FR Turbo Booster, Model 518-063, 518-043 | Spectranetics Corp. | 2007 | 2 |
| +2 | K091299 | Spectranetics Turbo-Tandem System | Spectranetics Corp. | 2009 | — |
| +3 | K094036 | Turbo-Tandem System Model 472-110 | Spectranetics Corp. | 2010 | — |
| +3 | K140775 | Turbo-Elite Atherectomy Catheters, Turbo Tandem System | Spectranetics Corporation | 2014 | 3 |
| +4 | K181642 | B-Laser Atherectomy System | Eximo Medical, Ltd. | 2018 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2007 | 1 | 0 |
| 2009 | 1 | 0 |
| 2010 | 1 | 0 |
| 2011 | 1 | 0 |
| 2013 | 1 | 0 |
| 2014 | 1 | 0 |
| 2015 | 2 | 0 |
| 2017 | 3 | 0 |
| 2018 | 2 | 0 |
| 2020 | 1 | 0 |
| 2021 | 1 | 0 |
| 2022 | 5 | 0 |
| 2023 | 2 | 0 |
| 2024 | 1 | 0 |
| 2025 | 2 | 0 |
| 2026 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in DQY? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new DQY device
- #1 K234115 APRO 55 Catheterscore 17.5 · 6 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K961999 | Schneider Guider Softip Guiding CathetersSummary page 1…(612) 550-5771 Ronald W. Bennett Senior Regulatory Affairs Specialist SCHNEIDER GUIDER Softip® Guiding Catheters Class I SCHNEIDER GUIDER Softip® Guiding Catheters (X961999) Cordis Vista brite tip™ guiding catheter The SCHNEIDER GUIDER Softip® Guiding Catheter in this 510(k) is avajlable in four Fren… | 1996 |
What descends from it? (26)
See all 263 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Aug 9, 2022 | Class II | Z-1582-2022 | Process control | The impacted products were distributed with an MPXF tip curve shape instead of the 40XF tip curve shape for the Guider/40XF/8FR/90CM guide catheter. |
| Dec 11, 2017 | Class II | Z-2735-2018 | Process change control | Potential polymer material degradation. |
| Oct 5, 2016 | Class II | Z-0338-2017 | Employee error | Firm became aware of potentially defective Guider Guide Catheters due to a manufacturing non-conformance that could lead to hub leaks in the resulting product. |
| Oct 5, 2016 | Class II | Z-0340-2017 | Employee error | Firm became aware of potentially defective Guider Guide Catheters due to a manufacturing non-conformance that could lead to hub leaks in the resulting product. |
| Oct 5, 2016 | Class II | Z-0341-2017 | Employee error | Firm became aware of potentially defective Guider Guide Catheters due to a manufacturing non-conformance that could lead to hub leaks in the resulting product. |
| Oct 5, 2016 | Class II | Z-0342-2017 | Employee error | Firm became aware of potentially defective Guider Guide Catheters due to a manufacturing non-conformance that could lead to hub leaks in the resulting product. |
| Oct 5, 2016 | Class II | Z-0344-2017 | Employee error | Firm became aware of potentially defective Guider Guide Catheters due to a manufacturing non-conformance that could lead to hub leaks in the resulting product. |
| Oct 5, 2016 | Class II | Z-0337-2017 | Employee error | Firm became aware of potentially defective Guider Guide Catheters due to a manufacturing non-conformance that could lead to hub leaks in the resulting product. |
| Oct 5, 2016 | Class II | Z-0343-2017 | Employee error | Firm became aware of potentially defective Guider Guide Catheters due to a manufacturing non-conformance that could lead to hub leaks in the resulting product. |
| Oct 5, 2016 | Class II | Z-0333-2017 | Employee error | Firm became aware of potentially defective Guider Guide Catheters due to a manufacturing non-conformance that could lead to hub leaks in the resulting product. |
| Oct 5, 2016 | Class II | Z-0335-2017 | Employee error | Firm became aware of potentially defective Guider Guide Catheters due to a manufacturing non-conformance that could lead to hub leaks in the resulting product. |
| Oct 5, 2016 | Class II | Z-0346-2017 | Employee error | Firm became aware of potentially defective Guider Guide Catheters due to a manufacturing non-conformance that could lead to hub leaks in the resulting product. |
| Oct 5, 2016 | Class II | Z-0347-2017 | Employee error | Firm became aware of potentially defective Guider Guide Catheters due to a manufacturing non-conformance that could lead to hub leaks in the resulting product. |
| Oct 5, 2016 | Class II | Z-0334-2017 | Employee error | Firm became aware of potentially defective Guider Guide Catheters due to a manufacturing non-conformance that could lead to hub leaks in the resulting product. |
| Oct 5, 2016 | Class II | Z-0336-2017 | Employee error | Firm became aware of potentially defective Guider Guide Catheters due to a manufacturing non-conformance that could lead to hub leaks in the resulting product. |
| Oct 5, 2016 | Class II | Z-0339-2017 | Employee error | Firm became aware of potentially defective Guider Guide Catheters due to a manufacturing non-conformance that could lead to hub leaks in the resulting product. |
| Oct 5, 2016 | Class II | Z-0345-2017 | Employee error | Firm became aware of potentially defective Guider Guide Catheters due to a manufacturing non-conformance that could lead to hub leaks in the resulting product. |
| Oct 6, 2009 | Not classified | Z-0107-2010 | Environmental control | Product degradation-- a polymer portion of the distal section of the catheter may degrade prematurely, leading to potential systemic embolization of particulates with ischemic complications. |
| Jul 11, 2007 | Not classified | Z-0260-2008 | Labeling mix-ups | Mislabeling: Catheters were found to be mislabeled (100 cm labeled devices actually contains 90cm length catheter) |
| Jul 11, 2007 | Not classified | Z-0259-2008 | Labeling mix-ups | Mislabeling: Catheters were found to be mislabeled (90 cm labeled devices actually contains 100 cm length catheter) |
The 5 latest of 20 recalls; the web page lists all.
No ancestor or descendant has a Class I recall.
What does this lineage say?
How does its category compare?
All DQY devicesDQY Catheter, Percutaneous: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · DQY
21in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · DQY
115in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · DQY
1in 2025
Class I (10 since 2007) Class II/III not classified
Adverse event reports per year · DQY
3,056in 2025
MAUDE reports naming product code DQY, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 16 | 85 | 6 | 313 |
| 2008 | 31 | 66 | 12 | 496 |
| 2009 | 37 | 73 | 17 | 819 |
| 2010 | 22 | 74 | 32 | 1,293 |
| 2011 | 36 | 71 | 6 | 1,211 |
| 2012 | 30 | 87 | 2 | 1,191 |
| 2013 | 32 | 92 | 36 (1) | 1,352 |
| 2014 | 19 | 127 | 3 | 1,497 |
| 2015 | 27 | 118 | 3 | 2,337 |
| 2016 | 22 | 117 | 22 (7) | 2,252 |
| 2017 | 26 | 113 | 8 | 2,489 |
| 2018 | 25 | 106 | 2 | 2,855 |
| 2019 | 30 | 111 | 1 | 2,902 |
| 2020 | 20 | 120 | 2 | 2,889 |
| 2021 | 18 | 81 | 2 | 3,006 |
| 2022 | 18 | 74 | 0 | 2,643 |
| 2023 | 18 | 114 | 17 | 3,049 |
| 2024 | 16 | 87 | 1 | 3,190 |
| 2025 | 21 | 115 | 1 (1) | 3,056 |
| 2026 | 11 | 30 | 2 (1) | — |
What is it indicated for?
The SCHNEIDER GUIDER Softip® Guiding Catheters are designed for the introduction of interventional devices. They are intended to facilitate the placement of interventional devices into the peripheral and coronary systems and, in addition for the XF models, into the neurovascular system.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K980453 · printed Nov 6, 2026
Cite this page
510(k) Report. “K980453: Schneider Guider Softip Guiding Catheters.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K980453
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026