510(k) · Cleared
ClearedCable Screw with Integral Crimp
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Pioneer Surgical Technology · JDQ Cerclage, Fixation
Cleared on May 22, 1998 after an 88-day review. The FDA summary names no predicate by number.
- Decided
- May 22, 1998
- Review
- 88 days
- Class
- Class II
- Product code
- JDQ
- Recalls
- None
- Predicates
- 0
The FDA summary names no predicate by number.
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in JDQ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new JDQ device
- #1 K261271 Cable-Ready® Cable Grip Systemscore 14.7 · 11 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against?
The FDA summary names no predicate by number.
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
How does its category compare?
All JDQ devicesJDQ Cerclage, Fixation: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · JDQ
2in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · JDQ
130in 2024
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · JDQ
5in 2019
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · JDQ
84in 2025
MAUDE reports naming product code JDQ, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 1 | — | 0 | 5 |
| 2008 | 2 | — | 2 | 25 |
| 2009 | 2 | — | 0 | 19 |
| 2010 | 0 | — | 2 | 28 |
| 2011 | 6 | 130 | 0 | 30 |
| 2012 | 3 | 134 | 0 | 130 |
| 2013 | 1 | — | 1 | 321 |
| 2014 | 2 | — | 0 | 223 |
| 2015 | 5 | 92 | 1 | 205 |
| 2016 | 4 | 129 | 0 | 224 |
| 2017 | 1 | — | 0 | 138 |
| 2018 | 6 | 101 | 0 | 182 |
| 2019 | 0 | — | 5 | 158 |
| 2020 | 4 | 93 | 0 | 125 |
| 2021 | 0 | — | 0 | 141 |
| 2022 | 2 | — | 0 | 134 |
| 2023 | 3 | 141 | 0 | 98 |
| 2024 | 3 | 130 | 0 | 115 |
| 2025 | 2 | — | 0 | 84 |
| 2026 | 1 | — | 0 | — |
What is it indicated for?
The Cable Screw with Integral Crimp is indicated where a fracture may not be securely held by either a screw or a cerclage cable alone The device can be used in conjunction with plating:
humeral fractures
niedial malleolus fractures
butterfly fractures requiring interfragmentary fixation with cerclage cabling
femur fractures
pelvic fractures
forearm shaft fractures
patellar fractures
olecranon fractures
proximal tibial and tibial shaft fractures
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K980690 · printed Nov 6, 2026
Cite this page
510(k) Report. “K980690: Cable Screw with Integral Crimp.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K980690
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Page last updated Nov 3, 2026