# K980876: Asthma Monitor AM2

- Pathway: 510(k) (Traditional)
- Applicant: Erich Jaeger GmbH & Co. KG
- Decided: Aug 14, 1998 after 161 days
- Product code: BZG (Spirometer, Diagnostic)
- Device class: 2
- Regulation: 21 CFR 868.1840
- Recalls of this device: none
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
No predicates in its family tree.
15 later devices descend from it; 1 cites it directly.
0 of 16 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- None. FDA’s file for it is a 510(k) statement, which names no predicate.

## Most influential descendants
- K090486: Asthma Monitor Model, AM1 (2009), 14 descendants of its own
- K092324: Spiropro (2009), 9 descendants of its own
- K100687: Vitalograph Hand Held Spirometer, Model 2120 (2010), 6 descendants of its own
- K092890: Asthma Monitor AM3 (2010), 3 descendants of its own
- K200550: Vitalograph Model 2120 In2itive eDiary (2020), 2 descendants of its own

## Indications for use
The Asthma Monitor AM2 from JAEGER is an electronic measurement device to monitor the lung function (determination of the respiratory flows and volume) with high reproducibility wherever and whenever is a need of. The AM2 measures the flow during expiration serving for the calculation of further parameters as FEV1, FVC or FEF25-75. The AM2 is used to monitor the respiratory status of human beings in the areas asthma, chronic obstructive pulmonary disorder and in areas like occupational medicine, clinical trials and disease management. The patient is informed of the results by numeric values for selected parameters (e.g. PEF, FEV1). Furthermore a visual control unit, displayed as a kind of traffic lights, allows an immediate indication of the measurement based on criteria defined by the patient’s physician. The device saves the results of a measurement (always with date and time) automatically in an internal database. The memory capacity is designed to store up to 400 measurements. In addition, a questionnaire functionality can be called up by the use of a software package (AMOS) to record e.g. the “Quality of Life” status. When enabled, the AM2 can be programmed with up to 12 questions, where the patient can select then from up to 7 different answers. This information is also stored in the internal database and can be transmitted for evaluation to a standard PC using the software package AMOS. The AM2 is designed to replace ordinary peak flow meter, diary and pencil by a single system. Easy handling, sturdy and handy design allow the Asthma Monitor AM2 being used almost everywhere: at work, at home, in school, for experts opinion, research or clinical trial purposes and in occupational medicine.