510(k) Report

510(k) · Cleared

Recalled · Class II · 2026

Becton Dickinson Syringe

K980987510(k) TraditionalCleared Jun 11, 199886-day reviewClass II21 CFR 880.5860Recalls 15

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Bd Becton Dickinson Vacutainer Systems Preanalytic · FMF Syringe, Piston

Cleared on Jun 11, 1998 after an 86-day review. The FDA summary names its predicate without a number. 2 of the 114 devices in its family, counting itself, have a Class I recall. 113 later devices descend from it. It has been recalled 15 times.

Decided
Jun 11, 1998
Review
86 days
Class
Class II
Product code
FMF
Recalls
15
Predicates
0

113 later devices descend from it, 18 citing it directly. 2 of the 114 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2022: 14 descendants cleared2007: 1 descendant cleared2005: 4 descendants cleared2020: 10 descendants cleared2011: 5 descendants cleared2019: 5 descendants cleared, 1 with a Class I recall12025: 6 descendants cleared2013: 3 descendants cleared2018: 5 descendants cleared2012: 2 descendants cleared2009: 2 descendants cleared2015: 4 descendants cleared2006: 1 descendant cleared2017: 10 descendants cleared, 1 with a Class I recall12014: 3 descendants cleared2016: 8 descendants cleared2021: 10 descendants cleared2024: 5 descendants cleared2023: 9 descendants cleared2026: 6 descendants cleared1995200020052010201520202025
  1. This deviceK980987 Becton Dickinson SyringeBd Becton Dickinson Vacutainer Systems Preanalytic · 1998
  2. Descendant · cites it directly · 35 of its ownK070264 50 ML Terumo Syringe for Administration of UV Sensitive Medicines…Terumo Europe N.V. · 2007
  3. Descendant · 2 generations on · 34 of its ownK083514 Terumo Syringe with/Without NeedleTerumo Europe N.V. · 2009
  4. Descendant · cites it directly · 33 of its ownK112434 BD Oral/Enteral Syringe with BD Univia ConnectionBd Medical Surgical · 2011
  5. Descendant · 3 generations on · 32 of its ownK170651 Sterile Disposable Syringe with Safety Needle, Sterile Disposable…Yangzhou Medline Industry, Co., Ltd. · 2017
  6. Descendant · 2 generations on · 31 of its ownK142128 Kangaroo Enteral Feeding Syringes with Enfit ConnectorCovidien · 2015

Showing 5 of 113 descendants. See all descendants

Descendants cleared per year

2005: 42006: 12007: 12009: 22011: 52012: 22013: 32014: 32015: 42016: 82017: 102018: 52019: 52020: 102021: 102022: 142023: 92024: 52025: 62026: 6
2005high 14 (2022) · low 0 (2008)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 113 · See all
  • K07026450 ML Terumo Syringe for Administration of UV Sensitive Medicines, Model BS-50LBTerumo Europe N.V. · 2007 · cites it directly35descendants
  • K083514Terumo Syringe with/Without NeedleTerumo Europe N.V. · 2009 · 2 generations on34descendants
  • K112434BD Oral/Enteral Syringe with BD Univia ConnectionBd Medical Surgical · 2011 · cites it directly33descendants
  • K170651Sterile Disposable Syringe with Safety Needle, Sterile Disposable Syringe with Needle…Yangzhou Medline Industry, Co., Ltd. · 2017 · 3 generations on32descendants
  • K142128Kangaroo Enteral Feeding Syringes with Enfit ConnectorCovidien · 2015 · 2 generations on31descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K980987Becton Dickinson SyringeBd Becton Dickinson Vacutainer Systems Preanalytic1998—
+1K07026450 ML Terumo Syringe for Administration of UV Sensitive Medicines, Model BS-50LBTerumo Europe N.V.2007—
+2K083514Terumo Syringe with/Without NeedleTerumo Europe N.V.2009—
+1K112434BD Oral/Enteral Syringe with BD Univia ConnectionBd Medical Surgical2011—
+3K170651Sterile Disposable Syringe with Safety Needle, Sterile Disposable Syringe with…Yangzhou Medline Industry, Co., Ltd.2017—
+2K142128Kangaroo Enteral Feeding Syringes with Enfit ConnectorCovidien2015—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200540
200610
200710
200920
201150
201220
201330
201430
201540
201680
2017101
201850
201951
2020100
2021100
2022140
202390
202450
202560
202660
FIG. 1 — Lineage of K980987, 1998–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FMF device

  1. #1 K241821 Luer Lock Syringe with Safety Needle; Luer…score 16.2 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—Becton Dickinson Single Use Hypodermic Syringes
Named without a submission number
—

What descends from it? (113)

See all 113

18 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K07026450 ML Terumo Syringe for Administration of UV Sensitive Medicines, Model BS-50LBTerumo Europe N.V.200735
K112434BD Oral/Enteral Syringe with BD Univia ConnectionBd Medical Surgical201133
K110771BD Single Use, Hypodermic SyringeBecton, Dickinson and Company201117
K110053Correctinject Safety System, Catheter Connector, Filter, Infusion Adaptor, Syringe KitSmiths Medical Asd, Inc.20125
K042500Duopross SyringesDuopross Meditech Corporation20054
K151766BD Single Use, Hypodermic SyringeBecton, Dickinson and Company20152
K113241BD Emerald Single Use, Hypodermic SyringeBecton Dickinson20112
K182589BD Plastipak SyringeBecton, Dickinson and Company20191
K090929Sterile Hypodermic Syringe and Needle for Single Use, Sterile Insuling Syringe for…Shanghai Double Dove Industrial Co., Ltd.20091
K042486Procedur 10 Syringe DeviceAvanca Medical Devices, Inc.20051

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 27, 2026Class IIZ-1496-2026Under Investigation by firmDuring the bulk packaging process, some Luer Slip syringes were incorrectly included in a batch designated for Luer Lok syringes.
Dec 23, 2020Class IIZ-1109-2021Process controlThe product packaging for the 10 ml eccentric tip syringe may exhibit open seals. An open seal impacts package integrity and potentially compromises sterility of the syringe.
Feb 1, 2019Class IIZ-2007-2019Under Investigation by firmThe syringes have scale markings that were printed incorrectly. The scale is skewed to varying degrees, resulting in missing and/or partially printed scale numbers and scale lines.
Oct 2, 2012Class IIZ-0564-2013Mixed-up of materials/componentsBD received a complaint report for mixed syringe tips (oral and non-oral syringe tips) within the same lot.
Apr 9, 2007Not classifiedZ-0813-2007OtherSmall cracks in plastic trays of one lot of BD 5mL Syringe Luer Lok Tip, Bulk Sterile Convenience Paks that may adversely affect tray integrity and steriity.
Feb 15, 2007Not classifiedZ-0719-2007OtherThe unit label on a limited number of syringes is incorrect, citing a 20 ml size. All other levels of labeling are correct, identifying the product as 30 ml size.
May 25, 2005Not classifiedZ-1054-05OtherThe firm believes that some of the syringes in the affected lots may not have received the intended sterilization dose.
May 25, 2005Not classifiedZ-1058-05OtherThe firm believes that some of the syringes in the affected lots may not have received the intended sterilization dose.
May 25, 2005Not classifiedZ-1055-05OtherThe firm believes that some of the syringes in the affected lots may not have received the intended sterilization dose.
May 25, 2005Not classifiedZ-1056-05OtherThe firm believes that some of the syringes in the affected lots may not have received the intended sterilization dose.
May 25, 2005Not classifiedZ-1057-05OtherThe firm believes that some of the syringes in the affected lots may not have received the intended sterilization dose.
Aug 4, 2003Not classifiedZ-1231-03OtherMislabeled kit: Prefilled Lidocaine syringe is labeled 'sterile fluid path only', outer label states kit content is sterile
Aug 4, 2003Not classifiedZ-1232-03OtherMislabeled kit: Prefilled Lidocaine syringe is labeled 'sterile fluid path only', outer label states kit content is sterile
Aug 4, 2003Not classifiedZ-1233-03OtherMislabeled kit: Prefilled Lidocaine syringe is labeled 'sterile fluid path only', outer label states kit content is sterile
Aug 4, 2003Not classifiedZ-1230-03OtherMislabeled kit: Prefilled Lidocaine syringe is labeled 'sterile fluid path only', outer label states kit content is sterile

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K162219CADD® Infusion Disposables Portfolio with NRFit™ connectorsSmiths Medical Asd, Inc. · 2017 · 2 generations on · ◆ in the 2017 descendants bar of FIG. 11 Class I recall
  • K190502ENFit Oral / Enteral SyringeJiangsu Caina Medical Co.,Ltd · 2019 · 5 generations on · ◆ in the 2019 descendants bar of FIG. 14 Class I recalls

What does this lineage say?

  • Of its descendants, K070264 (Terumo Europe N.V., 2007) has gone furthest, with 35 descendants of its own.

How does its category compare?

All FMF devices

FMF Syringe, Piston: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FMF

13in 2025

2007: 232008: 92009: 142010: 112011: 162012: 92013: 122014: 132015: 112016: 122017: 132018: 82019: 112020: 152021: 202022: 242023: 222024: 142025: 132026 (partial year): 9
2007high 24 (2022) · low 8 (2018)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FMF

114in 2025

2007: 90 days median2008: 74 days median2009: 94 days median2010: 130 days median2011: 109 days median2012: 86 days median2013: 154 days median2014: 84 days median2015: 87 days median2016: 146 days median2017: 135 days median2018: 87 days median2019: 146 days median2020: 107 days median2021: 175 days median2022: 211 days median2023: 173 days median2024: 130 days median2025: 114 days median2026 (partial year): 239 days median
2007high 239 (2026) · low 74 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FMF

19in 2023

2007: 62008: 112009: 152010: 122011: 32012: 62013: 12014: 12015: 22017: 32018: 22019: 12020: 22021: 12022: 22023: 19, 18 Class I2026 (partial year): 2, 1 Class I
2007high 19 (2023) · low 0 (2016)2026

Class I (19 since 2007) Class II/III not classified

Adverse event reports per year · FMF

4,181in 2025

2007: 6742008: 6912009: 9132010: 6792011: 1862012: 3072013: 3262014: 2442015: 2732016: 6842017: 1,8582018: 3,9522019: 5,8632020: 5,3342021: 5,8742022: 5,2332023: 5,2142024: 5,1222025: 4,181
2007high 5,874 (2021) · low 186 (2011)2025

MAUDE reports naming product code FMF, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200723906674
200897411691
2009149415913
20101113012679
2011161093186
20129866307
2013121541326
201413841244
201511872273
2016121460684
20171313531,858
201888723,952
20191114615,863
20201510725,334
20212017515,874
20222421125,233
20232217319 (18)5,214
20241413005,122
20251311404,181
202692392 (1)—

What is it indicated for?

These syringes are intended for use by health care professionals for general purpose fluid aspiration/injection.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K980987: Becton Dickinson Syringe.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K980987

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026