Elecsys Afp

K981282510(k) TraditionalCleared May 1, 199823-day reviewClass II21 CFR 866.6010

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Roche Diagnostics GmbH · LOJ Kit, Test,Alpha-Fetoprotein For Testicular Cancer

Cleared on May 1, 1998 as substantially equivalent to K973351 (Boehringer Mannheim Corp., 1997) after a 23-day review. Its family tree reaches back 1 generation to 1997. 2 later devices descend from it.

Decided
May 1, 1998
Review
23 days
Class
Class II
Product code
LOJ
Recalls
None
Predicates
1

Its longest line of descent runs back 1 generation to K973351 (1997). 2 ancestors in all. 2 later devices descend from it, 2 citing it directly.

Descendants, by year cleared 2004: 1 descendant cleared2022: 1 descendant clearedK973351 · Elecsys Afp · Boehringer Mannheim Corp. · 19971995200020052010201520202025
Descendants, by generation 1 generation forward: 2 descendants K973351 · Elecsys Afp · Boehringer Mannheim Corp. · 19970
  1. Root · 1 generation backK973351 Elecsys AfpBoehringer Mannheim Corp. · 1997
  2. This deviceK981282 Elecsys AfpRoche Diagnostics GmbH · 1998
  3. Descendant · cites it directly · 0 of its ownK220176 Elecsys AFPRoche Diagnostics · 2022
  4. Descendant · cites it directly · 0 of its ownK043095 Roche Diagnostics Elecsys Afp Calset IIRoche Diagnostics Corp. · 2004

Most influential descendants

Showing 2 of 2 · See all
  • K220176Elecsys AFPRoche Diagnostics · 2022 · cites it directly0descendants
  • K043095Roche Diagnostics Elecsys Afp Calset IIRoche Diagnostics Corp. · 2004 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K973351Elecsys AfpBoehringer Mannheim Corp.1997—
0 (this device)K981282Elecsys AfpRoche Diagnostics GmbH1998—
+1K220176Elecsys AFPRoche Diagnostics2022—
+1K043095Roche Diagnostics Elecsys Afp Calset IIRoche Diagnostics Corp.2004—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200410
202210
FIG. 1 — Lineage of K981282, 1997–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LOJ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LOJ device

  1. #1 K220176 Elecsys AFPscore 4.6 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K973351Elecsys Afp
Summary page 2
…ys 1010 and 2010 immunoassay analyzers. The Roche Diagnostics/Boehringer Mannheim Elecsys® AFP Assay has been approved for use on the Elecsys 2010 immunoassay analyzer (K973351). The application of the Elecsys® AFP Assay on the Elecsys 1010 immunoassay analyzer is substantially equivalent to the same assa…
Boehringer Mannheim Corp.1997—

What descends from it? (2)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K220176Elecsys AFPRoche Diagnostics20220
K043095Roche Diagnostics Elecsys Afp Calset IIRoche Diagnostics Corp.20040

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.
  • Of its descendants, K220176 (Roche Diagnostics, 2022) has gone furthest, with 0 descendants of its own.

How does its category compare?

All LOJ devices

LOJ Kit, Test,Alpha-Fetoprotein For Testicular Cancer: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LOJ

2in 2022

2008: 22009: 22022: 2
2007high 2 (2008) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LOJ

Too few LOJ 510(k)s in any year since 2007 to give a median.

Recalls per year · LOJ

1in 2010

2007: 12009: 12010: 1
2007high 1 (2007) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LOJ

5in 2025

2007: 12008: 52009: 72010: 12011: 62012: 62013: 112014: 32015: 32016: 52017: 92018: 182019: 142020: 12021: 52022: 12023: 32024: 42025: 5
2007high 18 (2018) · low 1 (2007)2025

MAUDE reports naming product code LOJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—11
20082—05
20092—17
20100—11
20110—06
20120—06
20130—011
20140—03
20150—03
20160—05
20170—09
20180—018
20190—014
20200—01
20210—05
20222—01
20230—03
20240—04
20250—05
20260—0—

What is it indicated for?

Immunoassay for the in vitro quantitative determination of alpha-fetoprotein in human serum and plasma to aid in the management of patients with non-seminomatous germ cell tumors.

The electrochemiluminescence immunoassay “ECLIA” is intended for use on the Boehringer Mannheim Elecsys 1010 and 2010 immunoassay analyzers.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K981282: Elecsys Afp.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K981282

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.