510(k) Report

510(k) · Cleared

Recalled · 2008

Amm Cal, Bil Cal, CO2 Cal, HDL Cal, Iron/MG Cal, LDL Cal, MC Cal, Upro Cal, and Ise Calibrator Levels 1 and 2

K981706510(k) TraditionalCleared Jul 9, 199856-day reviewClass II21 CFR 862.1150Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Abbott Laboratories · JIX Calibrator, Multi-Analyte Mixture

Cleared on Jul 9, 1998 after a 56-day review. The FDA summary names no predicate by number. 1 later device descends from it. It has been recalled 3 times.

Decided
Jul 9, 1998
Review
56 days
Class
Class II
Product code
JIX
Recalls
3
Predicates
0

1 later device descends from it, 1 citing it directly.

Descendants, by year cleared 2011: 1 descendant cleared1995200020052010201520202025
  1. This deviceK981706 Amm Cal, Bil Cal, CO2 Cal, HDL Cal, Iron/MG Cal, LDL Cal, MC Cal…Abbott Laboratories · 1998
  2. Descendant · cites it directly · 0 of its ownK103403 Multiconstituent CalibratorAbbott Laboratories · 2011

Most influential descendants

Showing 1 of 1 · See all
  • K103403Multiconstituent CalibratorAbbott Laboratories · 2011 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K981706Amm Cal, Bil Cal, CO2 Cal, HDL Cal, Iron/MG Cal, LDL Cal, MC Cal, Upro Cal…Abbott Laboratories1998—
+1K103403Multiconstituent CalibratorAbbott Laboratories2011—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201110
FIG. 1 — Lineage of K981706, 1998–2011Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JIX? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against?

The FDA summary names no predicate by number.

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K103403Multiconstituent CalibratorAbbott Laboratories20110

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 24, 2008Not classifiedZ-0950-2009Error in labelingAn incorrect Magnesium calibrator 2 value (CAL 2) was provided for Clinical Chemistry Iron/Magnesium Calibrator lot number 54187M200. Use of the incorrect Magnesium CAL 2 value causes quality control and patient magnesium results to be…
Aug 10, 2007Not classifiedZ-0288-2008Material/Component ContaminationIncorrect Calibrator values: changes in a raw material used in the Bilirubin Calibrator value assignment process contributed to variability in the preparation of the standard.
Dec 23, 2005Not classifiedZ-0516-06OtherThe two Multiconstituent Calibrator (MCC) lot numbers 19906M200 and 20879M100 failed to meet product performance expectations throughout the labeled expiration period, as a result of the shortened expiration dating, two lots are now…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K103403 (Abbott Laboratories, 2011) has gone furthest, with 0 descendants of its own.

How does its category compare?

All JIX devices

JIX Calibrator, Multi-Analyte Mixture: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JIX

1in 2016

2007: 62008: 52009: 42010: 52011: 72012: 22013: 42014: 12015: 32016: 1
2007high 7 (2011) · low 0 (2017)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JIX

29in 2015

2007: 53 days median2008: 69 days median2009: 121 days median2010: 100 days median2011: 77 days median2013: 50 days median2015: 29 days median
2007high 121 (2009) · low 29 (2015)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JIX

7in 2023

2007: 22008: 32009: 22010: 12011: 22012: 12014: 12016: 12017: 12018: 12019: 12020: 22023: 72026 (partial year): 1
2007high 7 (2023) · low 0 (2013)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JIX

18in 2025

2007: 1402008: 1732009: 392011: 92012: 32013: 242014: 352015: 42016: 62017: 172018: 162019: 122020: 222021: 92022: 112023: 132024: 152025: 18
2007high 173 (2008) · low 0 (2010)2025

MAUDE reports naming product code JIX, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20076532140
20085693173
20094121239
2010510010
201177729
20122—13
2013450024
20141—135
201532904
20161—16
20170—117
20180—116
20190—112
20200—222
20210—09
20220—011
20230—713
20240—015
20250—018
20260—1—

What is it indicated for?

An Abbott Clinical Chemistry Calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K981706: Amm Cal, Bil Cal, CO2 Cal, HDL Cal, Iron/MG Cal, LDL Cal, MC Cal, Upro Cal, and Ise Calibrator Levels 1 and 2.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K981706

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026