510(k) · Cleared
Recalled · Class I · 2022Sensititre 18-24 HR Susceptibility Plates (Addition of Meropenem)
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Trek Diagnostic Systems, Inc. · JTW System, Transport, Aerobic
Cleared on Nov 30, 1998 after a 75-day review. FDA’s file for it is a 510(k) statement, which names no predicate. It has a Class I recall. It has been recalled 9 times, including 9 Class I recalls.
- Decided
- Nov 30, 1998
- Review
- 75 days
- Class
- Class I
- Product code
- JTW
- Recalls
- 9 (9 Class I)
- Predicates
- 0
FDA’s file for it is a 510(k) statement, which names no predicate.
Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
- Candidate predicates for a device like this, rankedin the report
- Recall history across each candidate’s familyin the report
What was it cleared against?
FDA’s file for it is a 510(k) statement, which names no predicate.
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (1 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Oct 20, 2022 | Class I | Z-0327-2023 | Process control | An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when… |
| Oct 20, 2022 | Class I | Z-0329-2023 | Process control | An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when… |
| Oct 20, 2022 | Class I | Z-0334-2023 | Process control | An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when… |
| Oct 20, 2022 | Class I | Z-0330-2023 | Process control | An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when… |
| Oct 20, 2022 | Class I | Z-0328-2023 | Process control | An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when… |
| Oct 20, 2022 | Class I | Z-0331-2023 | Process control | An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when… |
| Oct 20, 2022 | Class I | Z-0332-2023 | Process control | An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when… |
| Oct 20, 2022 | Class I | Z-0333-2023 | Process control | An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when… |
| Oct 20, 2022 | Class I | Z-0335-2023 | Process control | An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when… |
The 5 latest of 9 recalls; the web page lists all.
No ancestor or descendant has a Class I recall.
How does its category compare?
All JTW devicesJTW System, Transport, Aerobic: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · JTW
1in 2012
2026 to Oct 25, outlined.
Median 510(k) review, days · JTW
Too few JTW 510(k)s in any year since 2007 to give a median.
Recalls per year · JTW
1in 2022
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · JTW
36in 2025
MAUDE reports naming product code JTW, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 0 | 0 |
| 2008 | 0 | — | 0 | 0 |
| 2009 | 0 | — | 1 | 0 |
| 2010 | 0 | — | 0 | 0 |
| 2011 | 0 | — | 0 | 2 |
| 2012 | 1 | — | 0 | 2 |
| 2013 | 0 | — | 0 | 1 |
| 2014 | 0 | — | 0 | 3 |
| 2015 | 0 | — | 0 | 2 |
| 2016 | 0 | — | 0 | 2 |
| 2017 | 0 | — | 0 | 14 |
| 2018 | 0 | — | 4 | 32 |
| 2019 | 0 | — | 0 | 8 |
| 2020 | 0 | — | 0 | 12 |
| 2021 | 0 | — | 0 | 7 |
| 2022 | 0 | — | 1 | 5 |
| 2023 | 0 | — | 0 | 22 |
| 2024 | 0 | — | 0 | 37 |
| 2025 | 0 | — | 0 | 36 |
| 2026 | 0 | — | 0 | — |
What is it indicated for?
The Sensititre 18 hour MIC or Breakpoint Susceptibility system is an in vitro diagnostic product for clinical susceptibility testing of gram negative and gram positive organisms.
This 510(k) is for the addition of Meropenem in the dilution range of 0.004 — 8 ug/ml to the Sensititre 18 — 24 hour MIC panel for testing gram negative isolates.
Indications for use read by OCR from a scanned FDA 510(k) statement, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K983244 · printed Nov 4, 2026
Cite this page
510(k) Report. “K983244: Sensititre 18-24 HR Susceptibility Plates (Addition of Meropenem).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K983244
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Page last updated Nov 3, 2026