510(k) Report

510(k) · Cleared

Recalled · 2006

Invasive Blood Pressure (Ibp) Function, Model 9500 Multigas Monitor

K983250510(k) TraditionalCleared Dec 12, 1999452-day reviewClass II21 CFR 870.1110Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Magnetic Resonance Equipment Corp. · DSK Computer, Blood-Pressure

Cleared on Dec 12, 1999 after a 452-day review. The FDA summary names its predicate without a number. It has been recalled once.

Decided
Dec 12, 1999
Review
452 days
Class
Class II
Product code
DSK
Recalls
1
Predicates
0

The FDA summary names its predicate without a number.

FIG. 1 — Lineage of K983250Source: FDA 510(k) database · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DSK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DSK device

  1. #1 K231586 Endophys Blood Pressure Monitor model BPM-30score 8.7 · 9 generations · no Class I recall
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—BCI Model 9100 Multigas Monitor with Invasive Blood Pressure function
Named without a submission number
—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 29, 2006Not classifiedZ-0258-2007OtherInterference problems: certain invasive blood pressure transducers connected to these cables are susceptible to interference if exposed to static magnetic field strengths exceeding 300 gauss. In certain orientations, the transducer causes…

No ancestor or descendant has a Class I recall.

How does its category compare?

All DSK devices

DSK Computer, Blood-Pressure: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DSK

1in 2023

2007: 12008: 12015: 22016: 12023: 1
2007high 2 (2015) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DSK

Too few DSK 510(k)s in any year since 2007 to give a median.

Recalls per year · DSK

No recall names DSK since 2007.

Adverse event reports per year · DSK

26in 2025

2010: 102011: 52012: 692013: 52014: 82015: 72016: 132017: 62018: 932019: 1482020: 292021: 312022: 292023: 142024: 122025: 26
2007high 148 (2019) · low 0 (2007)2025

MAUDE reports naming product code DSK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—00
20081—00
20090—00
20100—010
20110—05
20120—069
20130—05
20140—08
20152—07
20161—013
20170—06
20180—093
20190—0148
20200—029
20210—031
20220—029
20231—014
20240—012
20250—026
20260—0—

What is it indicated for?

Invasive blood pressure function is an option for the Model 9500 Multigas monitor. This function would be used when a clinician determined the need for continuous monitoring of arterial or venous blood pressure for adult patients undergoing MRI examination.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K983250: Invasive Blood Pressure (Ibp) Function, Model 9500 Multigas Monitor.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K983250

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Page last updated Nov 3, 2026