510(k) Report

510(k) · Cleared

Recalled · Class II · 2020

Cook Mini-Incubator

K983642510(k) TraditionalCleared Feb 1, 1999108-day reviewClass II21 CFR 884.6120Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Cook Ob/Gyn · MQG Accessory, Assisted Reproduction

Cleared on Feb 1, 1999 after a 108-day review. The FDA summary names no predicate by number. 14 later devices descend from it. It has been recalled 3 times.

Decided
Feb 1, 1999
Review
108 days
Class
Class II
Product code
MQG
Recalls
3
Predicates
0

14 later devices descend from it, 2 citing it directly.

Descendants, by year cleared 2011: 3 descendants cleared2012: 1 descendant cleared2017: 1 descendant cleared2015: 1 descendant cleared2020: 1 descendant cleared2014: 1 descendant cleared2018: 3 descendants cleared2025: 1 descendant cleared2022: 1 descendant cleared2024: 1 descendant cleared1995200020052010201520202025
  1. This deviceK983642 Cook Mini-IncubatorCook Ob/Gyn · 1999
  2. Descendant · cites it directly · 11 of its ownK092183 Embryoscope and EmbryoslideUnisense Fertilitech A/S · 2011
  3. Descendant · 2 generations on · 9 of its ownK111715 EmbryoscopeUnisense Fertilitech A/S · 2011
  4. Descendant · 2 generations on · 8 of its ownK113075 Embryo Viewer SoftwareUnisense Fertilitech A/S · 2011
  5. Descendant · 2 generations on · 4 of its ownK150961 EmbryoSlide Culture DishVitrolife A/S · 2015
  6. Descendant · 3 generations on · 4 of its ownK133712 Embryoviewer SoftwareUnisense Fertilitech A/S · 2014

Showing 5 of 14 descendants. See all descendants

Descendants cleared per year

2011: 32012: 12014: 12015: 12017: 12018: 32020: 12022: 12024: 12025: 1
2011high 3 (2011) · low 0 (2013)2026

Most influential descendants

Showing 5 of 14 · See all
  • K092183Embryoscope and EmbryoslideUnisense Fertilitech A/S · 2011 · cites it directly11descendants
  • K111715EmbryoscopeUnisense Fertilitech A/S · 2011 · 2 generations on9descendants
  • K113075Embryo Viewer SoftwareUnisense Fertilitech A/S · 2011 · 2 generations on8descendants
  • K150961EmbryoSlide Culture DishVitrolife A/S · 2015 · 2 generations on4descendants
  • K133712Embryoviewer SoftwareUnisense Fertilitech A/S · 2014 · 3 generations on4descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K983642Cook Mini-IncubatorCook Ob/Gyn1999—
+1K092183Embryoscope and EmbryoslideUnisense Fertilitech A/S2011—
+2K111715EmbryoscopeUnisense Fertilitech A/S2011—
+2K113075Embryo Viewer SoftwareUnisense Fertilitech A/S2011—
+2K150961EmbryoSlide Culture DishVitrolife A/S2015—
+3K133712Embryoviewer SoftwareUnisense Fertilitech A/S2014—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201130
201210
201410
201510
201710
201830
202010
202210
202410
202510
FIG. 1 — Lineage of K983642, 1999–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MQG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MQG device

  1. #1 K261083 Wallace® Aspiration Pump (WAPV1); Wallace®…score 12.2 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

The FDA summary names no predicate by number.

What descends from it? (14)

See all 14

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K092183Embryoscope and EmbryoslideUnisense Fertilitech A/S201111
K121566Planer BT37 IncubatorPlaner Plc20121

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 16, 2020Class IIZ-0809-2021Under Investigation by firmA small percentage of the Disposable Humidification Flasks may have a manufacturing defect that can cause the silicone stoppers to dislodge or gas to leak from the flasks, impeding gas flow to the incubation chambers of the Mini Incubator.
Jan 12, 2018Class IIZ-1377-2018Device DesignThe temperature of the incubation lids of the MINC Bench top Incubators may become too high or too low for proper gamete/embryo incubation due to fatigue failure of the flat flex lid heater.
Jan 12, 2018Class IIZ-1378-2018Device DesignThe temperature of the incubation lids of the MINC Bench top Incubators may become too high or too low for proper gamete/embryo incubation due to fatigue failure of the flat flex lid heater.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K092183 (Unisense Fertilitech A/S, 2011) has gone furthest, with 11 descendants of its own.

How does its category compare?

All MQG devices

MQG Accessory, Assisted Reproduction: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MQG

1in 2025

2011: 42012: 12014: 32016: 22017: 22018: 32024: 12025: 12026 (partial year): 2
2007high 4 (2011) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MQG

135in 2018

2011: 137 days median2014: 236 days median2018: 135 days median
2007high 236 (2014) · low 135 (2018)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MQG

1in 2023

2019: 12020: 12023: 1
2007high 1 (2019) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MQG

1in 2025

2016: 12019: 22020: 32023: 32024: 12025: 1
2007high 3 (2020) · low 0 (2007)2025

MAUDE reports naming product code MQG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
2011413700
20121—00
20130—00
2014323600
20150—00
20162—01
20172—00
2018313500
20190—12
20200—13
20210—00
20220—00
20230—13
20241—01
20251—01
20262—0—

What is it indicated for?

The COOK Mini-Incubator is intended to be used to store and

preserve gametes and/or embryos at or near body temperature.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K983642: Cook Mini-Incubator.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K983642

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026