510(k) Report

510(k) · Cleared

Recalled · Class II · 2018

Abbott Architect Ferritin

K983759510(k) TraditionalCleared Dec 7, 199842-day reviewClass II21 CFR 866.5340Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Abbott Laboratories · DBF Ferritin, Antigen, Antiserum, Control

Cleared on Dec 7, 1998 after a 42-day review. The FDA summary names its predicate without a number. It has been recalled 2 times.

Decided
Dec 7, 1998
Review
42 days
Class
Class II
Product code
DBF
Recalls
2
Predicates
0

The FDA summary names its predicate without a number.

FIG. 1 — Lineage of K983759Source: FDA 510(k) database · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DBF? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DBF device

  1. #1 K193650 LIAISON Ferritinscore 2.0 · 1 generation · no Class I recall
  2. #2 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—AxSYM Ferritin assay
Named without a submission number
—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 21, 2018Class IIZ-2208-2019Device DesignAbbott has become aware of a manufacturing issue that can result in damage to a bottle neck in some of the reagent cartridges found in the lots listed in Appendix A. If a bottle neck is damaged, it may cause a failure during the reagent…
Jun 6, 2006Not classifiedZ-1204-06OtherSome ARCHITECT Ferritin Reagent lots may not be meeting the accuracy by correlation to AxSYM claims as they are listed in the package insert (slope of 1.0 +/- 0.2).

No ancestor or descendant has a Class I recall.

How does its category compare?

All DBF devices

DBF Ferritin, Antigen, Antiserum, Control: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DBF

1in 2021

2009: 12010: 12011: 12012: 12014: 12017: 12020: 12021: 1
2007high 1 (2009) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DBF

Too few DBF 510(k)s in any year since 2007 to give a median.

Recalls per year · DBF

3in 2012

2012: 3
2007high 3 (2012) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DBF

1in 2025

2012: 12013: 12016: 12017: 22018: 72019: 12020: 12021: 32023: 32024: 12025: 1
2007high 7 (2018) · low 0 (2007)2025

MAUDE reports naming product code DBF, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20091—00
20101—00
20111—00
20121—31
20130—01
20141—00
20150—00
20160—01
20171—02
20180—07
20190—01
20201—01
20211—03
20220—00
20230—03
20240—01
20250—01
20260—0—

What is it indicated for?

The Abbott ARCHITECT™ Ferritin assay is a Chemiluminescent Microparticle Immunoassay (CMIA) for the quantitative determination of ferritin in human serum and plasma on the Abbott ARCHITECT i System. Measurements obtained by this device aid in the diagnosis of diseases affecting iron metabolism, such as hemochromatosis (iron overload) and iron deficiency anemia.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K983759: Abbott Architect Ferritin.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K983759

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)