510(k) Report

510(k) · Cleared

Recalled · 2004

Theratron Elite

K983917510(k) SpecialCleared Dec 4, 199830-day reviewClass II21 CFR 892.5750Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Theratronics, Inc. · IWB System, Radiation Therapy, Radionuclide

Cleared on Dec 4, 1998 as substantially equivalent to K964606 (Theratronics, Inc., 1997) and 1 other predicate after a 30-day review. Its family tree reaches back 2 generations to 1990. 1 later device descends from it. It has been recalled 2 times.

Decided
Dec 4, 1998
Review
30 days
Class
Class II
Product code
IWB
Recalls
2
Predicates
2

Its longest line of descent runs back 2 generations to K894165 (1990). 4 ancestors in all, the oldest K850543 (1985). 1 later device descends from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1997: 1 other ancestor1985: 1 other ancestor2006: 1 descendant clearedK894165 · Theratron 1000 · Theratronics, Inc. · 1990K964607 · Theraton 1000E · Theratronics, Inc. · 1997 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 1 other ancestor 1 generation forward: 1 descendant K964607 · Theraton 1000E · Theratronics, Inc. · 1997 · recalledK894165 · Theratron 1000 · Theratronics, Inc. · 19900
  1. Root · 2 generations backK894165 Theratron 1000Theratronics, Inc. · 1990
  2. 1 generation backK964607 Theraton 1000ETheratronics, Inc. · 1997 · Recalled
  3. This deviceK983917 Theratron EliteTheratronics, Inc. · 1998
  4. Descendant · cites it directly · 0 of its ownK053485 Theratron Equinox, Models 80 CM and 100 CMMds Nordion, Inc. · 2006

Most influential descendants

Showing 1 of 1 · See all
  • K053485Theratron Equinox, Models 80 CM and 100 CMMds Nordion, Inc. · 2006 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K894165Theratron 1000Theratronics, Inc.1990—
−1K964607Theraton 1000ETheratronics, Inc.19971
0 (this device)K983917Theratron EliteTheratronics, Inc.1998—
+1K053485Theratron Equinox, Models 80 CM and 100 CMMds Nordion, Inc.2006—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198510
199710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200610
FIG. 1 — Lineage of K983917, 1990–2006Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in IWB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IWB device

  1. #1 K250391 Leksell Gamma Knife® (Elekta Esprit); Leksell…score 11.2 · 13 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K964606Theratron 780E
Summary page 1
…v 16, 1998 Trade name: Theratron Elite Common name: Cobalt Teletherapy device Classif. name: Radionuclide Radiation Therapy System Equivalent device: Theratron 780E (K964606), Theratron 1000E (K964607). Description of device: The Theratron Elite consists of a source head, collimator, gantry, main fra…
Theratronics, Inc.19971
*K964607Theraton 1000E
Summary page 1
…ratron Elite Common name: Cobalt Teletherapy device Classif. name: Radionuclide Radiation Therapy System Equivalent device: Theratron 780E (K964606), Theratron 1000E (K964607). Description of device: The Theratron Elite consists of a source head, collimator, gantry, main frame, base, controls and a pe…
Theratronics, Inc.19971

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K053485Theratron Equinox, Models 80 CM and 100 CMMds Nordion, Inc.20060

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 8, 2004Not classifiedZ-1183-06OtherThe source locking device may not fit properly through the smaller opening in the new field light shield which could prevent the source from being returned manually under emergency conditions.
Jul 8, 2004Not classifiedZ-1184-06OtherThe source locking device may not fit properly through the smaller opening in the new field light shield which could prevent the source from being returned manually under emergency conditions.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 8 years and passes through 1 company; a new generation was cleared every 7.3 years on median.
  • 2 of the 4 ancestors have a recall of any class.
  • Of its descendants, K053485 (Mds Nordion, Inc., 2006) has gone furthest, with 0 descendants of its own.

How does its category compare?

All IWB devices

IWB System, Radiation Therapy, Radionuclide: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · IWB

1in 2025

2007: 12009: 12011: 12012: 12014: 32015: 32016: 12017: 12018: 32019: 12020: 12021: 22022: 12024: 12025: 1
2007high 3 (2014) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IWB

57in 2018

2014: 170 days median2015: 79 days median2018: 57 days median
2007high 170 (2014) · low 57 (2018)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IWB

1in 2023

2007: 32008: 32009: 12010: 12011: 22014: 12023: 1
2007high 3 (2007) · low 0 (2012)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · IWB

3in 2025

2007: 82008: 152009: 82010: 92011: 52012: 32013: 22014: 82015: 52016: 72017: 42018: 22019: 32020: 22023: 12025: 3
2007high 15 (2008) · low 0 (2021)2025

MAUDE reports naming product code IWB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—38
20080—315
20091—18
20100—19
20111—25
20121—03
20130—02
2014317018
201537905
20161—07
20171—04
201835702
20191—03
20201—02
20212—00
20221—00
20230—11
20241—00
20251—03
20260—0—

What is it indicated for?

The device is intended to be used in:

Delivering the intended dose of radiation at a specified position;

Delivering the radiation in accordance with the selected relationship of the radiation to the patient (fixed or moving beam therapy, beam modifying device, etc);

Delivering the radiation without causing unnecessary risk to the patient, the operator, other persons or the immediate environment.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K983917: Theratron Elite.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K983917

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026