510(k) · Cleared
Recalled · 2006Stratus (R) CS Myoglobin (Myo) Testpak
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Dade Behring, Inc. · DDR Myoglobin, Antigen, Antiserum, Control
Cleared on Dec 14, 1998 as substantially equivalent to K981102 (Dade Behring, Inc., 1998) after a 28-day review. Its family tree reaches back 1 generation to 1998. 1 later device descends from it. It has been recalled 2 times.
- Decided
- Dec 14, 1998
- Review
- 28 days
- Class
- Class II
- Product code
- DDR
- Recalls
- 2
- Predicates
- 1
Its longest line of descent runs back 1 generation to K981102 (1998). 1 later device descends from it, 1 citing it directly.
- Root · 1 generation backK981102 Stratus CS Myoglobin (Myo) TestpakDade Behring, Inc. · 1998
- This deviceK984065 Stratus (R) CS Myoglobin (Myo) TestpakDade Behring, Inc. · 1998
- Descendant · cites it directly · 0 of its ownK051650 Stratus CS Acute Care MB Isoenzyme of Creatine Kinase (CKMB) Test…Dade Behring, Inc. · 2005 · Recalled
Most influential descendants
Showing 1 of 1 · See all- K051650Stratus CS Acute Care MB Isoenzyme of Creatine Kinase (CKMB) Test Pak; Stratus CS Acute…Dade Behring, Inc. · 2005 · cites it directly0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K981102 | Stratus CS Myoglobin (Myo) Testpak | Dade Behring, Inc. | 1998 | — |
| 0 (this device) | K984065 | Stratus (R) CS Myoglobin (Myo) Testpak | Dade Behring, Inc. | 1998 | — |
| +1 | K051650 | Stratus CS Acute Care MB Isoenzyme of Creatine Kinase (CKMB) Test Pak; Stratus… | Dade Behring, Inc. | 2005 | 11 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2005 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in DDR? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new DDR device
- #1 K231832 Access Myoglobinscore 6.3 · 2 generations · no Class I recall
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K981102 | Stratus CS Myoglobin (Myo) TestpakSummary page 1…/6/98 Device Name: Stratus® CS Myoglobin (MYQO) TestPak Classification Name: Myoglobin Immunological Test System Predicate Device: Stratus® CS Myoglobin (MYO) TestPak (K981102) Device Description: The MYO TestPak consists of a plastic cartridge with five wells and a small square of glass fiber paper emb… | 1998 |
What descends from it? (1)
1 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K051650 | Stratus CS Acute Care MB Isoenzyme of Creatine Kinase (CKMB) Test Pak; Stratus CS Acute… | 2005 | 0 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Apr 3, 2006 | Not classified | Z-0878-06 | Other | Foil sealing of the TestPak may occlude the pipet tip resulting in insufficient aspiration of fluids. Erroneous test results may be generated without an associated error code. Test results may be falsely elevated or depressed, and the… |
| Apr 3, 2006 | Not classified | Z-0884-06 | Other | Foil sealing of the DilPak may occlude the pipet tip resulting in insufficient aspiration of fluids. Erroneous test results may be generated without an associated error code. Test results may be falsely elevated or depressed, and the… |
No ancestor or descendant has a Class I recall.
What does this lineage say?
How does its category compare?
All DDR devicesDDR Myoglobin, Antigen, Antiserum, Control: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · DDR
1in 2023
2026 to Oct 25, outlined.
Median 510(k) review, days · DDR
Too few DDR 510(k)s in any year since 2007 to give a median.
Recalls per year · DDR
2in 2015
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · DDR
5in 2025
MAUDE reports naming product code DDR, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 2 | 0 |
| 2008 | 0 | — | 0 | 0 |
| 2009 | 2 | — | 1 | 3 |
| 2010 | 0 | — | 0 | 0 |
| 2011 | 0 | — | 0 | 5 |
| 2012 | 1 | — | 0 | 2 |
| 2013 | 1 | — | 0 | 3 |
| 2014 | 0 | — | 0 | 1 |
| 2015 | 0 | — | 2 | 0 |
| 2016 | 0 | — | 0 | 3 |
| 2017 | 0 | — | 0 | 5 |
| 2018 | 0 | — | 0 | 3 |
| 2019 | 0 | — | 0 | 6 |
| 2020 | 0 | — | 0 | 2 |
| 2021 | 0 | — | 0 | 1 |
| 2022 | 0 | — | 0 | 1 |
| 2023 | 1 | — | 0 | 7 |
| 2024 | 0 | — | 0 | 31 |
| 2025 | 0 | — | 0 | 5 |
| 2026 | 0 | — | 0 | — |
What is it indicated for?
The MYO TestPak used on the Stratus® CS STAT Fluorometric Analyzer is an in vitro diagnostic product for the measurement of Myoglobin in heparinized plasma. Measurements of Myoglobin are used as aids in the rapid diagnosis of renal or heart disease, e.g. acute myocardial infarction.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K984065 · printed Nov 6, 2026
Cite this page
510(k) Report. “K984065: Stratus (R) CS Myoglobin (Myo) Testpak.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K984065
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026