510(k) Report

510(k) · Cleared

Cleared

Vapo-Solution for Use in MDT Harvey Chemiclave Sterilizers, Models 5000, 5500, and 6000

K984270510(k) TraditionalCleared May 20, 1999171-day reviewClass II21 CFR 880.6885

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Roger L. Goodman, D.D.S., P.C. · MED Sterilant, Medical Devices

Cleared on May 20, 1999 after a 171-day review. The FDA summary names its predicate without a number.

Decided
May 20, 1999
Review
171 days
Class
Class II
Product code
MED
Recalls
None
Predicates
0

The FDA summary names its predicate without a number.

FIG. 1 — Lineage of K984270Source: FDA 510(k) database · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MED? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MED device

  1. #1 K251666 SYSTEM 1E Liquid Chemical Sterilant Processing…score 13.8 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—MDT Harvey Vapo-Steril® solution
Named without a submission number
—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

How does its category compare?

All MED devices

MED Sterilant, Medical Devices: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MED

2in 2023

2007: 22008: 12010: 62012: 32013: 12014: 12016: 12017: 12018: 22019: 42021: 22022: 12023: 22026 (partial year): 2
2007high 6 (2010) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MED

84in 2019

2010: 281 days median2012: 111 days median2019: 84 days median
2007high 281 (2010) · low 84 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MED

2in 2011

2007: 32008: 22011: 2
2007high 3 (2007) · low 0 (2009)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MED

9in 2025

2007: 342008: 792009: 1582010: 892011: 1062012: 3372013: 1402014: 1052015: 1102016: 1352017: 742018: 422019: 352020: 322021: 312022: 162023: 92024: 82025: 9
2007high 337 (2012) · low 8 (2024)2025

MAUDE reports naming product code MED, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—334
20081—279
20090—0158
20106281089
20110—2106
201231110337
20131—0140
20141—0105
20150—0110
20161—0135
20171—074
20182—042
2019484035
20200—032
20212—031
20221—016
20232—09
20240—08
20250—09
20262—0—

What is it indicated for?

Vapo Solution™ is an alcohol solution intended to be used only in Harvey Chemiclave sterilizers to sterilize medical and dental surgical instruments and equipment. Vapo Solution™ is intended to be a direct substitute for Vapo-Steril®. Vapo Solution™ is recommended for use in the following Chemiclave models: 4000, 5000, 5500, 6000, and models covered under K924380 and K943654. Vapo Solution™ is compatible with Chemiclave cycle parameters as follows: Model…

Read the full statement (350 words)

Vapo Solution™ is an alcohol solution intended to be used only in Harvey Chemiclave sterilizers to sterilize medical and dental surgical instruments and equipment. Vapo Solution™ is intended to be a direct substitute for Vapo-Steril®. Vapo Solution™ is recommended for use in the following Chemiclave models: 4000, 5000, 5500, 6000, and models covered under K924380 and K943654.

Vapo Solution™ is compatible with Chemiclave cycle parameters as follows:

Model 4000, 5000, 5500, 6000, EMS500, EM5500, and EM6000

• Operating pressure: 20–40 psi (1.41–2.82 kg/cm²)

• Operating Temperature: 132 ± 2 °C (270 ± 5 °F)

• Maximum load: Bagged instruments:

6" Chamber (models 4000, 5000, EM5000) 0.6Kg (1.3lbs)

8" Chamber (models 5500, EM5500) 1.2Kg (2.6 lbs)

10" Chamber (models 6000, EM6000) 2.4Kg (5.3lbs)

• Maximum load: Unwrapped instruments:

6" Chamber (models 4000, 5000, EM5000) 0.75Kg (1.7 lbs)

8" Chamber (models 5500, EM5500) 1.4Kg (3.1lbs)

10" Chamber (models 6000, EM6000) 2.8Kg (6.2lbs)

Model 7000 and 8000

• Operating pressure: approximately 25 psi

• Operating Temperature: 132 ± 2°C (270 ± 5°F)

• Maximum load: Bagged instruments:

8" Chamber (model 7000) 1.2Kg (2.6lbs)

10" Chamber (model 8000) 2.4Kg (5.3lbs)

• Maximum load: Unwrapped instruments:

8" Chamber (model 7000) 1.4 Kg (3.1lbs)

10" Chamber (model 8000) 2.8Kg (6.2lbs)

Vapo Solution™ is compatible with but not limited to the following instruments and material types:

• Surgical and cutting instruments, handpieces, and similar items

• Plastics that can tolerate in excess of 132 °C, i.e. Polysulfone, Phenolic, Nylon, Teflon, Polyethylene, Polyester, Polypropylene. Note: Polycarbonate is not chemically compatible.

• Rubber components made of Silicone, fluorosilicone

Vapo Solution™ is not compatible with the following materials:

• Linens, textiles or fabrics - These are incompatible with the chemical vapor process.

• Liquids and agars - These are incompatible with high temperature processing (132°C).

• Items contained in tightly woven packs or wraps, or sealed containers - These cause poor vapor penetration.

• Plastics which cannot tolerate temperatures in excess of 132°C.

• Polycarbonate - Chemically incompatible.

• Nylon tubing and bags - Inhibits penetration of chemical vapors.

Vapo Solution™ is not recommended for use for “flash” sterilization cycles.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K984270: Vapo-Solution for Use in MDT Harvey Chemiclave Sterilizers, Models 5000, 5500, and 6000.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K984270

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)