510(k) Report

510(k) · Cleared

Cleared

Echoflow Doppler Blood Velocity Meter

K990642510(k) TraditionalCleared Sep 16, 1999202-day reviewClass II21 CFR 870.2880

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Echocath, Inc. · JOP Transducer, Ultrasonic

Cleared on Sep 16, 1999 as substantially equivalent to K861378 (Ultramed, Inc., 1986) and 1 other predicate after a 202-day review. Its family tree reaches back 1 generation to 1986. 5 later devices descend from it.

Decided
Sep 16, 1999
Review
202 days
Class
Class II
Product code
JOP
Recalls
None
Predicates
2

Its longest line of descent runs back 1 generation to K861378 (1986). 2 ancestors in all. 5 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1993: 1 other ancestor2002: 2 descendants cleared2003: 1 descendant cleared2005: 1 descendant cleared2004: 1 descendant clearedK861378 · Ultramed TF 122 Ultrason Vascu Imag & Doppler Syst · Ultramed, Inc. · 19861990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 1 descendant2 generations forward: 3 descendants3 generations forward: 1 descendant K861378 · Ultramed TF 122 Ultrason Vascu Imag & Doppler Syst · Ultramed, Inc. · 19860
  1. Root · 1 generation backK861378 Ultramed TF 122 Ultrason Vascu Imag & Doppler SystUltramed, Inc. · 1986
  2. This deviceK990642 Echoflow Doppler Blood Velocity MeterEchocath, Inc. · 1999
  3. Descendant · cites it directly · 4 of its ownK013803 Flowguard DeviceBiosonix, Ltd. · 2002
  4. Descendant · 2 generations on · 2 of its ownK030357 Quantix/or DeviceCardiosonix, Ltd. · 2003 · Recalled
  5. Descendant · 3 generations on · 1 of its ownK041180 Quantix or Vessel Stabilizer DeviceCardiosonix, Ltd. · 2004
  6. Descendant · 2 generations on · 0 of its ownK050088 Quantix/or Blood Flow Monitor; Flexible Probe with Optional Vessel…Neoprobe Corp. · 2005
  7. Descendant · 2 generations on · 0 of its ownK023431 Quantix/ND DeviceCardiosonix, Ltd. · 2002

Most influential descendants

Showing 5 of 5 · See all
  • K013803Flowguard DeviceBiosonix, Ltd. · 2002 · cites it directly4descendants
  • K030357Quantix/or DeviceCardiosonix, Ltd. · 2003 · 2 generations on2descendants
  • K041180Quantix or Vessel Stabilizer DeviceCardiosonix, Ltd. · 2004 · 3 generations on1descendant
  • K050088Quantix/or Blood Flow Monitor; Flexible Probe with Optional Vessel StabilizerNeoprobe Corp. · 2005 · 2 generations on0descendants
  • K023431Quantix/ND DeviceCardiosonix, Ltd. · 2002 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K861378Ultramed TF 122 Ultrason Vascu Imag & Doppler SystUltramed, Inc.1986—
0 (this device)K990642Echoflow Doppler Blood Velocity MeterEchocath, Inc.1999—
+1K013803Flowguard DeviceBiosonix, Ltd.2002—
+2K030357Quantix/or DeviceCardiosonix, Ltd.20032
+3K041180Quantix or Vessel Stabilizer DeviceCardiosonix, Ltd.2004—
+2K050088Quantix/or Blood Flow Monitor; Flexible Probe with Optional Vessel StabilizerNeoprobe Corp.2005—
+2K023431Quantix/ND DeviceCardiosonix, Ltd.2002—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200220
200310
200410
200510
FIG. 1 — Lineage of K990642, 1986–2005Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JOP? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JOP device

  1. #1 K242662 Falcon/Xpress (Falcon/Xpress)score 8.8 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K861378Ultramed TF 122 Ultrason Vascu Imag & Doppler Syst
Summary page 1
…ood-flow measuring system Classification Name: Flowmeter, blood, ultrasonic, w/wo calibration, and Transducer, Ultrasonic (3) Predicate devices: 1. Ultramed Ultraplex, K861378 2. Zertl K921377, P/Meter (4) Description The EchoFlow velocity system is a computer-based ultrasonic Doppler blood velocity me…
Ultramed, Inc.1986—
*K921377P/Meter
Summary page 1
…g system Classification Name: Flowmeter, blood, ultrasonic, w/wo calibration, and Transducer, Ultrasonic (3) Predicate devices: 1. Ultramed Ultraplex, K861378 2. Zertl K921377, P/Meter (4) Description The EchoFlow velocity system is a computer-based ultrasonic Doppler blood velocity measuring system us…
Zertl Medical, Inc.1993—

What descends from it? (5)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K013803Flowguard DeviceBiosonix, Ltd.20024

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • None of the 2 ancestors has a recall.
  • Of its descendants, K013803 (Biosonix, Ltd., 2002) has gone furthest, with 4 descendants of its own.

How does its category compare?

All JOP devices

JOP Transducer, Ultrasonic: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JOP

2in 2024

2007: 12011: 12013: 12014: 12016: 12024: 2
2007high 2 (2024) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JOP

Too few JOP 510(k)s in any year since 2007 to give a median.

Recalls per year · JOP

No recall names JOP since 2007.

Adverse event reports per year · JOP

2in 2025

2014: 12017: 22020: 22021: 32022: 12024: 12025: 2
2007high 3 (2021) · low 0 (2007)2025

MAUDE reports naming product code JOP, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—00
20080—00
20090—00
20100—00
20111—00
20120—00
20131—00
20141—01
20150—00
20161—00
20170—02
20180—00
20190—00
20200—02
20210—03
20220—01
20230—00
20242—01
20250—02
20260—0—

What is it indicated for?

The EchoFlow BVM-1 is an ultrasonic Doppler system that can be used to measure blood flow velocities. It can be used either for measurement of vessels below the skin, or for intraoperative measurements. It is not intended for cardiac or fetal use. A 10 MHz probe is supplied. The unit maintains its specified accuracy even if the angle between the probe face and vessel axis is as much as £15°, unlike conventional Doppler systems where the angle between the…

Read the full statement (263 words)

The EchoFlow BVM-1 is an ultrasonic Doppler system that can be used to measure blood flow velocities. It can be used either for measurement of vessels below the skin, or for intraoperative measurements. It is not intended for cardiac or fetal use. A 10 MHz probe is supplied. The unit maintains its specified accuracy even if the angle between the probe face and vessel axis is as much as £15°, unlike conventional Doppler systems where the angle between the flow direction and the probe must be accurately known. The unit can detect large blood vessels at depths up to 12 mm and small blood vessels at depths up to 10 mm in normal tissue. The system will not detect any signals from vessels more than 15.5 mm from the probe.

The IOP10 transducer for the EchoFlow BVM-1 is intended to be used with that unit to measure blood flow velocities. It can be used either for measurement of vessels below the skin, or for intraoperative measurements. It is not intended for cardiac or fetal use. A 10 MHz probe is supplied. The unit maintains its specified accuracy even if the angle between the probe face and vessel axis is as much as +15°, unlike conventional Doppler systems where the angle between the flow direction and the probe must be accurately known. The unit can detect large blood vessels at depths up to 12 mm and small blood vessels at depths up to 10 mm in normal tissue. The system will not detect any signals from vessels more than 15.5 mm from the probe.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K990642: Echoflow Doppler Blood Velocity Meter.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K990642

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026