510(k) · Cleared
Recalled · 2009Transfx Intermediate External Fixation System,Transfx Small External Fixation System, Transfx Mini External Fixation
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Immedica, Inc. · LXT Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component, Metal Composite
Cleared on May 17, 1999 after a 63-day review. FDA’s file for it is a 510(k) statement, which names no predicate. 4 later devices descend from it. It has been recalled 8 times.
- Decided
- May 17, 1999
- Review
- 63 days
- Class
- Class II
- Product code
- LXT
- Recalls
- 8
- Predicates
- 0
4 later devices descend from it, 3 citing it directly.
- This deviceK990848 Transfx Intermediate External Fixation System,Transfx Small External…Immedica, Inc. · 1999
- Descendant · cites it directly · 2 of its ownK131413 Xtrafix Small External Fixation SystemZimmer, Inc. · 2014
- Descendant · cites it directly · 1 of its ownK141697 Xtrafix External Fixation System, Xtrafix Small External Fixation…Zimmer, Inc. · 2014
- Descendant · 2 generations on · 0 of its ownK152484 FastFrame External Fixation System - Distal RadiusZimmer, Inc. · 2015
- Descendant · cites it directly · 0 of its ownK061759 Reprocessed External Fixation DeviceAscent Healthcare Solutions · 2006
Most influential descendants
Showing 4 of 4 · See all- K131413Xtrafix Small External Fixation SystemZimmer, Inc. · 2014 · cites it directly2descendants
- K141697Xtrafix External Fixation System, Xtrafix Small External Fixation SystemZimmer, Inc. · 2014 · cites it directly1descendant
- K152484FastFrame External Fixation System - Distal RadiusZimmer, Inc. · 2015 · 2 generations on0descendants
- K061759Reprocessed External Fixation DeviceAscent Healthcare Solutions · 2006 · cites it directly0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| 0 (this device) | K990848 | Transfx Intermediate External Fixation System,Transfx Small External Fixation… | Immedica, Inc. | 1999 | — |
| +1 | K131413 | Xtrafix Small External Fixation System | Zimmer, Inc. | 2014 | — |
| +1 | K141697 | Xtrafix External Fixation System, Xtrafix Small External Fixation System | Zimmer, Inc. | 2014 | — |
| +2 | K152484 | FastFrame External Fixation System - Distal Radius | Zimmer, Inc. | 2015 | — |
| +1 | K061759 | Reprocessed External Fixation Device | Ascent Healthcare Solutions | 2006 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2006 | 1 | 0 |
| 2014 | 2 | 0 |
| 2015 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in LXT? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new LXT device
- #1 K230067 Para-Fix External Fixation Systemscore 6.4 · 2 generations · no Class I recall
What was it cleared against?
FDA’s file for it is a 510(k) statement, which names no predicate.
What descends from it? (4)
3 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| May 1, 2009 | Not classified | Z-1533-2009 | Process control | Drill tip is out of specifications, reducing the efficiency of drilling action, requiring more force and potentially heating the bone surface during operation. |
| May 1, 2009 | Not classified | Z-1532-2009 | Process control | Drill tip is out of specifications, reducing the efficiency of drilling action, requiring more force and potentially heating the bone surface during operation. |
| May 1, 2009 | Not classified | Z-1538-2009 | Process control | Drill tip is out of specifications, reducing the efficiency of drilling action, requiring more force and potentially heating the bone surface during operation. |
| May 1, 2009 | Not classified | Z-1534-2009 | Process control | Drill tip is out of specifications, reducing the efficiency of drilling action, requiring more force and potentially heating the bone surface during operation. |
| May 1, 2009 | Not classified | Z-1535-2009 | Process control | Drill tip is out of specifications, reducing the efficiency of drilling action, requiring more force and potentially heating the bone surface during operation. |
| May 1, 2009 | Not classified | Z-1536-2009 | Process control | Drill tip is out of specifications, reducing the efficiency of drilling action, requiring more force and potentially heating the bone surface during operation. |
| May 1, 2009 | Not classified | Z-1537-2009 | Process control | Drill tip is out of specifications, reducing the efficiency of drilling action, requiring more force and potentially heating the bone surface during operation. |
| Feb 10, 2003 | Not classified | Z-0907-03 | Other | Mislabeled; wrong identification number on pin. |
The 5 latest of 8 recalls; the web page lists all.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- Of its descendants, K131413 (Zimmer, Inc., 2014) has gone furthest, with 2 descendants of its own.
How does its category compare?
All LXT devicesLXT Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component, Metal Composite: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · LXT
1in 2023
2026 to Oct 25, outlined.
Median 510(k) review, days · LXT
Too few LXT 510(k)s in any year since 2007 to give a median.
Recalls per year · LXT
2in 2018
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · LXT
7in 2025
MAUDE reports naming product code LXT, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 1 | — | 1 | 26 |
| 2008 | 1 | — | 0 | 58 |
| 2009 | 1 | — | 7 | 41 |
| 2010 | 0 | — | 0 | 33 |
| 2011 | 1 | — | 0 | 85 |
| 2012 | 0 | — | 0 | 42 |
| 2013 | 2 | — | 0 | 39 |
| 2014 | 0 | — | 0 | 36 |
| 2015 | 0 | — | 0 | 35 |
| 2016 | 0 | — | 0 | 38 |
| 2017 | 0 | — | 0 | 66 |
| 2018 | 0 | — | 2 | 73 |
| 2019 | 0 | — | 0 | 40 |
| 2020 | 0 | — | 0 | 32 |
| 2021 | 0 | — | 0 | 24 |
| 2022 | 0 | — | 0 | 21 |
| 2023 | 1 | — | 0 | 18 |
| 2024 | 0 | — | 0 | 27 |
| 2025 | 0 | — | 0 | 7 |
| 2026 | 0 | — | 0 | — |
What is it indicated for?
Unstable fractures of the distal radius: intra-articular, extra-articular, and preliminary fixation before internal fixation. Fractures and fracture dislocations of the hand, wrist, forearm, mandible, foot, ankle, and long bones; joint fusion, osteotomies, and peri-articular fractures.
Indications for use read by OCR from a scanned FDA 510(k) statement, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K990848 · printed Nov 5, 2026
Cite this page
510(k) Report. “K990848: Transfx Intermediate External Fixation System,Transfx Small External Fixation System, Transfx Mini External Fixation.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K990848
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026