510(k) · Cleared
Recalled · Class III · 2022Vacutainer Brand Multiple Sample Luer Adapter, Blood Collection Sets and Safety-Lok Blood Collection, Safety Blood Colle
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Becton Dickinson Vacutainer Systems · JKA Tubes, Vials, Systems, Serum Separators, Blood Collection
Cleared on May 28, 1999 as substantially equivalent to K931367 (Bd Becton Dickinson Vacutainer Systems Preanalytic, 1993) and 1 other predicate after a 57-day review. Its family tree reaches back 2 generations to 1993. 14 later devices descend from it. It has been recalled 4 times.
- Decided
- May 28, 1999
- Review
- 57 days
- Class
- Class II
- Product code
- JKA
- Recalls
- 4
- Predicates
- 2
Its longest line of descent runs back 2 generations to K931367 (1993). 14 later devices descend from it, 6 citing it directly.
- Root · 2 generations backK931367 Vacutainer Brand Luer AdapterBd Becton Dickinson Vacutainer Systems Preanalytic · 1993
- 1 generation backK983490 Terumo Venoject,Luer AdapterTerumo Corp. · 1998
- This deviceK991088 Vacutainer Brand Multiple Sample Luer Adapter, Blood Collection Sets…Becton Dickinson Vacutainer Systems · 1999
- Descendant · cites it directly · 6 of its ownK200027 Blood Collection Needle with/without Holder, Safety Blood Collection…Jiangsu Caina Medical Co.,Ltd · 2020
- Descendant · 2 generations on · 1 of its ownK222478 BD Vacutainer® Luer-Lok™ Access Device, BD Vacutainer® Blood…Becton, Dickinson and Company · 2023
- Descendant · cites it directly · 1 of its ownK190485 Kawasumi Multiple Sample Adapter with Pre-Attached Holder (MBCH)Kawasumi Laboratories, Inc. · 2019
- Descendant · cites it directly · 1 of its ownK160532 Greiner HoldexGreiner Bio-One Na, Inc. · 2016
- Descendant · 3 generations on · 0 of its ownK252899 Female Culture Device; Male Culture Device; Transfer Device; Access…Zhejiang Kindly Medical Device Co., Ltd. · 2026
Showing 5 of 14 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 14 · See all- K200027Blood Collection Needle with/without Holder, Safety Blood Collection Needle with/without…Jiangsu Caina Medical Co.,Ltd · 2020 · cites it directly6descendants
- K222478BD Vacutainer® Luer-Lok™ Access Device, BD Vacutainer® Blood Transfer DeviceBecton, Dickinson and Company · 2023 · 2 generations on1descendant
- K190485Kawasumi Multiple Sample Adapter with Pre-Attached Holder (MBCH)Kawasumi Laboratories, Inc. · 2019 · cites it directly1descendant
- K160532Greiner HoldexGreiner Bio-One Na, Inc. · 2016 · cites it directly1descendant
- K252899Female Culture Device; Male Culture Device; Transfer Device; Access DeviceZhejiang Kindly Medical Device Co., Ltd. · 2026 · 3 generations on0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −2 | K931367 | Vacutainer Brand Luer Adapter | Bd Becton Dickinson Vacutainer Systems Preanalytic | 1993 | — |
| −1 | K983490 | Terumo Venoject,Luer Adapter | Terumo Corp. | 1998 | — |
| 0 (this device) | K991088 | Vacutainer Brand Multiple Sample Luer Adapter, Blood Collection Sets and… | Becton Dickinson Vacutainer Systems | 1999 | — |
| +1 | K200027 | Blood Collection Needle with/without Holder, Safety Blood Collection Needle… | Jiangsu Caina Medical Co.,Ltd | 2020 | — |
| +2 | K222478 | BD Vacutainer® Luer-Lok™ Access Device, BD Vacutainer® Blood Transfer Device | Becton, Dickinson and Company | 2023 | — |
| +1 | K190485 | Kawasumi Multiple Sample Adapter with Pre-Attached Holder (MBCH) | Kawasumi Laboratories, Inc. | 2019 | — |
| +1 | K160532 | Greiner Holdex | Greiner Bio-One Na, Inc. | 2016 | — |
| +3 | K252899 | Female Culture Device; Male Culture Device; Transfer Device; Access Device | Zhejiang Kindly Medical Device Co., Ltd. | 2026 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2016 | 1 | 0 |
| 2018 | 1 | 0 |
| 2019 | 3 | 0 |
| 2020 | 1 | 0 |
| 2022 | 2 | 0 |
| 2023 | 3 | 0 |
| 2024 | 1 | 0 |
| 2025 | 1 | 0 |
| 2026 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in JKA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new JKA device
- #1 K200027 Blood Collection Needle with/without Holder…score 15.6 · 3 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K931367 | Vacutainer Brand Luer AdapterSummary page 3…The predicate devices, K number, and clearance date are also identified in the table below. Manufacturer Predicate Device K-Number Cléarance Date Be_cton VACUTAINER® | K931367 5/18/93 Dickinson Brand Multiple VACUTAINER | Sample Luer Systems Adapter Terumo Medical TERUMO® K 983490 11/30/98 Corporation Br… | 1993 |
| * | K983490 | Terumo Venoject,Luer AdapterSummary page 3…r Predicate Device K-Number Cléarance Date Be_cton VACUTAINER® | K931367 5/18/93 Dickinson Brand Multiple VACUTAINER | Sample Luer Systems Adapter Terumo Medical TERUMO® K 983490 11/30/98 Corporation Brand - VENOJECT™ Luer Adapter Cfim Aoz, Dhme Lh i 3i, 49 Andrea Hroncich Date Regulatery Affairs Associate… | 1998 |
What descends from it? (14)
See all 146 later devices cite it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K200027 | Blood Collection Needle with/without Holder, Safety Blood Collection Needle with/without… | 2020 | 6 |
| K190485 | Kawasumi Multiple Sample Adapter with Pre-Attached Holder (MBCH) | 2019 | 1 |
| K160532 | Greiner Holdex | 2016 | 1 |
| K243649 | BD Vacutainer® Multiple Sample Luer Adapter | 2025 | 0 |
| K181743 | Blood Transfer Accessory (BTA) | 2019 | 0 |
| K181754 | Blood Collection Accessory (BUCA) | 2019 | 0 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Feb 24, 2022 | Class III | Z-0837-2022 | Material/Component Contamination | Recalled devices were packaged in convenience kits. |
| Jan 6, 2022 | Class III | Z-0665-2022 | Under Investigation by firm | May shed contaminant particles into the urine specimen that may be misidentified as Red Blood Cells when used in conjunction with the Sysmex UF-1000i or UX-2000 Analyzers and lead to false elevations in RBC counts |
| Sep 8, 2006 | Not classified | Z-0253-2007 | Other | Reports of failure of the Luer Lok Access Device to lock securely to certain catheter devices because the male luer taper surface is not within the specification. |
| Jun 21, 2006 | Not classified | Z-1462-06 | Other | Firm received complaints indicating failure of the Non-Patient (NP) sleeve to function properly. This sleeve covers the cannula and prevents leakage during blood collection. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 6 years and passes through 3 companies; a new generation was cleared every 5.6 years on median.
- None of the 2 ancestors has a recall.
- Of its descendants, K200027 (Jiangsu Caina Medical Co.,Ltd, 2020) has gone furthest, with 6 descendants of its own.
- 3 other devices also cite its predicate K931367.
How does its category compare?
All JKA devicesJKA Tubes, Vials, Systems, Serum Separators, Blood Collection: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · JKA
7in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · JKA
115in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · JKA
13in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · JKA
6,089in 2025
MAUDE reports naming product code JKA, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 4 | 281 | 3 | 73 |
| 2008 | 2 | — | 3 | 78 |
| 2009 | 2 | — | 3 | 42 |
| 2010 | 4 | 28 | 4 | 37 |
| 2011 | 0 | — | 14 | 67 |
| 2012 | 1 | — | 2 | 54 |
| 2013 | 3 | 244 | 1 | 33 |
| 2014 | 1 | — | 1 | 68 |
| 2015 | 1 | — | 6 | 41 |
| 2016 | 4 | 116 | 1 | 65 |
| 2017 | 2 | — | 2 | 2,039 |
| 2018 | 1 | — | 2 | 9,731 |
| 2019 | 6 | 181 | 9 | 6,469 |
| 2020 | 6 | 181 | 0 | 3,774 |
| 2021 | 4 | 217 | 3 | 3,610 |
| 2022 | 9 | 164 | 2 | 3,375 |
| 2023 | 17 | 189 | 2 | 4,400 |
| 2024 | 6 | 240 | 0 | 4,609 |
| 2025 | 7 | 115 | 13 | 6,089 |
| 2026 | 8 | 250 | 1 | — |
What is it indicated for?
The VACUTAINER® Brand Luer Adapter is a sterile, non-invasive device used to connect venous access devices such as needles, blood collection sets, and infusion sets to blood collection tubes. They are also used in connection with non-needle devices for collection of blood from catheters. The VACUTAINER® Brand Luer Adapter is sold by itself and as a component of other VACUTAINER Brand devices.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K991088 · printed Nov 5, 2026
Cite this page
510(k) Report. “K991088: Vacutainer Brand Multiple Sample Luer Adapter, Blood Collection Sets and Safety-Lok Blood Collection, Safety Blood Colle.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K991088
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026