510(k) Report

510(k) · Cleared

Recalled · Class III · 2018

Autohdl Cholesterol Reagent Set

K992002510(k) TraditionalCleared Aug 5, 199951-day reviewClass I21 CFR 862.1475Recalls 13

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Pointe Scientific, Inc., · LBR Ldl & Vldl Precipitation, Hdl

Cleared on Aug 5, 1999 after a 51-day review. FDA’s file for it is a 510(k) statement, which names no predicate. It has been recalled 13 times.

Decided
Aug 5, 1999
Review
51 days
Class
Class I
Product code
LBR
Recalls
13
Predicates
0

FDA’s file for it is a 510(k) statement, which names no predicate.

FIG. 1 — Lineage of K992002Source: FDA 510(k) database · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LBR? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against?

FDA’s file for it is a 510(k) statement, which names no predicate.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 10, 2018Class IIIZ-0389-2019Error in labelingVials of the autoHDL/LDL Calibrator were shipped without a vial label.
Mar 13, 2015Class IIIZ-2764-2015Nonconforming Material/ComponentCrystals may be observed in the R1 reagent. Crystals may be a result of the reagent being frozen during transport. The crystals do not impact product performance.
Jan 29, 2009Not classifiedZ-1200-2009Nonconforming Material/ComponentThe R1 reagent has microbial contamination.
Jan 29, 2009Not classifiedZ-1201-2009Nonconforming Material/ComponentThe R1 reagent has microbial contamination.
Sep 18, 2006Not classifiedZ-0074-2007OtherUnexpected changes in QC and proficiency results due to QC and proficiency material matrix.
Sep 18, 2006Not classifiedZ-0075-2007OtherUnexpected changes in QC and proficiency results due to QC and proficiency material matrix.
Sep 18, 2006Not classifiedZ-0076-2007OtherUnexpected changes in QC and proficiency results due to QC and proficiency material matrix.
Jun 12, 2006Not classifiedZ-1454-06OtherUnexpected changes in QC and proficiency results due to QC and proficiency material matrix.
Jun 12, 2006Not classifiedZ-1457-06OtherUnexpected changes in QC and proficiency results due to QC and proficiency material matrix.
Jun 12, 2006Not classifiedZ-1455-06OtherUnexpected changes in QC and proficiency results due to QC and proficiency material matrix.
Jun 12, 2006Not classifiedZ-1458-06OtherUnexpected changes in QC and proficiency results due to QC and proficiency material matrix.
Jun 12, 2006Not classifiedZ-1459-06OtherUnexpected changes in QC and proficiency results due to QC and proficiency material matrix.
Jun 12, 2006Not classifiedZ-1456-06OtherUnexpected changes in QC and proficiency results due to QC and proficiency material matrix.

No ancestor or descendant has a Class I recall.

How does its category compare?

All LBR devices

LBR Ldl & Vldl Precipitation, Hdl: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LBR

1in 2015

2008: 12015: 1
2007high 1 (2008) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LBR

Too few LBR 510(k)s in any year since 2007 to give a median.

Recalls per year · LBR

1in 2018

2009: 42016: 12018: 1
2007high 4 (2009) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LBR

8in 2025

2007: 12008: 12011: 12012: 12016: 12019: 22021: 12022: 22023: 22024: 12025: 8
2007high 8 (2025) · low 0 (2009)2025

MAUDE reports naming product code LBR, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—01
20081—01
20090—40
20100—00
20110—01
20120—01
20130—00
20140—00
20151—00
20160—11
20170—00
20180—10
20190—02
20200—00
20210—01
20220—02
20230—02
20240—01
20250—08
20260—0—

What is it indicated for?

This product is to be used in a diagnostic laboratory setting, by qualified laboratory technologists, for the quantitative determination of high density lipoprotein (HDL) cholesterol in serum or plasma. HDL cholesterol is recognized as a useful tool in identifying patients who are at a higher risk for coronary heart disease. Low HDL cholesterol levels are associated with an increased risk. This reagent set is intended for in vitro diagnostic use only.

Indications for use read by OCR from a scanned FDA 510(k) statement, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K992002: Autohdl Cholesterol Reagent Set.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K992002

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Page last updated Nov 3, 2026