510(k) Report

510(k) · Cleared

Recalled · Class II · 2013

Becton Dickinson Syringe

K992734510(k) TraditionalCleared Oct 1, 199949-day reviewClass II21 CFR 880.5860Recalls 6

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Bd Becton Dickinson Vacutainer Systems Preanalytic · MEG Syringe, Antistick

Cleared on Oct 1, 1999 after a 49-day review. The FDA summary names its predicate without a number. 10 later devices descend from it. It has been recalled 6 times.

Decided
Oct 1, 1999
Review
49 days
Class
Class II
Product code
MEG
Recalls
6
Predicates
0

10 later devices descend from it, 2 citing it directly.

Descendants, by year cleared 2010: 1 descendant cleared2005: 1 descendant cleared2015: 1 descendant cleared2016: 1 descendant cleared2007: 1 descendant cleared2021: 1 descendant cleared2020: 2 descendants cleared2022: 1 descendant cleared2023: 1 descendant cleared1995200020052010201520202025
  1. This deviceK992734 Becton Dickinson SyringeBd Becton Dickinson Vacutainer Systems Preanalytic · 1999
  2. Descendant · cites it directly · 8 of its ownK050131 Insosafe Bak'snap Retractable Insulin Safety SyringeM.K. Meditech Co., Ltd. · 2005
  3. Descendant · 2 generations on · 5 of its ownK152808 U&U Insulin Syringe with/without Safety Retractable DeviceU&U Medical Technology Co, Ltd. · 2016
  4. Descendant · 3 generations on · 2 of its ownK202570 Insulin Syringe, Insulin Syringe with Safety RetractableJiangsu Micsafe Medical Technology Co., Ltd. · 2021
  5. Descendant · 4 generations on · 1 of its ownK220061 Safety Insulin SyringesBerpu Medical Technology Co., Ltd. · 2022
  6. Descendant · 5 generations on · 0 of its ownK230061 Insulin SyringeAnhui Hongyu Wuzhou Medical Manufacturer Co., Ltd. · 2023

Showing 5 of 10 descendants. See all descendants

Descendants cleared per year

2005: 12007: 12010: 12015: 12016: 12020: 22021: 12022: 12023: 1
2005high 2 (2020) · low 0 (2006)2026

Most influential descendants

Showing 5 of 10 · See all
  • K050131Insosafe Bak'snap Retractable Insulin Safety SyringeM.K. Meditech Co., Ltd. · 2005 · cites it directly8descendants
  • K152808U&U Insulin Syringe with/without Safety Retractable DeviceU&U Medical Technology Co, Ltd. · 2016 · 2 generations on5descendants
  • K202570Insulin Syringe, Insulin Syringe with Safety RetractableJiangsu Micsafe Medical Technology Co., Ltd. · 2021 · 3 generations on2descendants
  • K220061Safety Insulin SyringesBerpu Medical Technology Co., Ltd. · 2022 · 4 generations on1descendant
  • K230061Insulin SyringeAnhui Hongyu Wuzhou Medical Manufacturer Co., Ltd. · 2023 · 5 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K992734Becton Dickinson SyringeBd Becton Dickinson Vacutainer Systems Preanalytic1999—
+1K050131Insosafe Bak'snap Retractable Insulin Safety SyringeM.K. Meditech Co., Ltd.2005—
+2K152808U&U Insulin Syringe with/without Safety Retractable DeviceU&U Medical Technology Co, Ltd.2016—
+3K202570Insulin Syringe, Insulin Syringe with Safety RetractableJiangsu Micsafe Medical Technology Co., Ltd.2021—
+4K220061Safety Insulin SyringesBerpu Medical Technology Co., Ltd.2022—
+5K230061Insulin SyringeAnhui Hongyu Wuzhou Medical Manufacturer Co., Ltd.2023—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200510
200710
201010
201510
201610
202020
202110
202210
202310
FIG. 1 — Lineage of K992734, 1999–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MEG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MEG device

  1. #1 K192551 1ml Luer Slip or Luer Lock Syringe, Syringe…score 19.6 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—Becton Dickinson Syringes
Named without a submission number
—
*—SafetyGlide™ Needle
Named without a submission number
—

What descends from it? (10)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K050131Insosafe Bak'snap Retractable Insulin Safety SyringeM.K. Meditech Co., Ltd.20058
K092746Micronjet 600Nanopass Technologies, Ltd.20100

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 1, 2013Class IIZ-1577-2013Process controlThere has been an increase in complaints regarding the needle/safety mechanism disengaging from the syringe.
Jul 24, 2006Not classifiedZ-1525-06OtherBD received reports of needle assembly disengagement from the syringe.
Oct 25, 2005Not classifiedZ-0403-06OtherSafety mechanisms on the syringes are rotating too freely. This can result in the safety mechanism not holding its rotational position and increasing risk of an accidental needle-stick.
Oct 25, 2005Not classifiedZ-0402-06OtherSafety mechanisms on the syringes are rotating too freely. This can result in the safety mechanism not holding its rotational position and increasing risk of an accidental needle-stick.
Oct 25, 2005Not classifiedZ-0404-06OtherSafety mechanisms on the syringes are rotating too freely. This can result in the safety mechanism not holding its rotational position and increasing risk of an accidental needle-stick.
Oct 25, 2005Not classifiedZ-0405-06OtherSafety mechanisms on the syringes are rotating too freely. This can result in the safety mechanism not holding its rotational position and increasing risk of an accidental needle-stick.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K050131 (M.K. Meditech Co., Ltd., 2005) has gone furthest, with 8 descendants of its own.

How does its category compare?

All MEG devices

MEG Syringe, Antistick: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MEG

1in 2025

2007: 82008: 42009: 22010: 22011: 52012: 52013: 12014: 52015: 72016: 32017: 32018: 12019: 42020: 42021: 52022: 132023: 52025: 12026 (partial year): 1
2007high 13 (2022) · low 0 (2024)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MEG

229in 2023

2007: 100 days median2008: 145 days median2011: 113 days median2012: 109 days median2014: 153 days median2015: 253 days median2016: 179 days median2017: 130 days median2019: 249 days median2020: 296 days median2021: 152 days median2022: 216 days median2023: 229 days median
2007high 296 (2020) · low 100 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MEG

1in 2024

2008: 12010: 12013: 52015: 12024: 1, 1 Class I
2007high 5 (2013) · low 0 (2007)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · MEG

148in 2025

2007: 212008: 202009: 262010: 342011: 342012: 222013: 272014: 362015: 512016: 942017: 1812018: 2692019: 4972020: 3532021: 5412022: 1702023: 1202024: 1122025: 148
2007high 541 (2021) · low 20 (2008)2025

MAUDE reports naming product code MEG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20078100021
20084145120
20092—026
20102—134
20115113034
20125109022
20131—527
20145153036
20157253151
20163179094
201731300181
20181—0269
201942490497
202042960353
202151520541
2022132160170
202352290120
20240—1 (1)112
20251—0148
20261—0—

What is it indicated for?

The Becton Dickinson SafetyGlide™ Syringes are used for a variety of injections and aspirations of fluid from vials, ampules and parts of the body below the surface of the skin. The insulin syringe has scale lines in insulin units and is typically used for insulin injections. The allergy syringes come in test and treatment versions. The allergy test syringe has an intra-dermal bevel for intra-dermal injections. The allergy treatment syringe has a regular…

Read the full statement (140 words)

The Becton Dickinson SafetyGlide™ Syringes are used for a variety of injections and aspirations of fluid from vials, ampules and parts of the body below the surface of the skin. The insulin syringe has scale lines in insulin units and is typically used for insulin injections. The allergy syringes come in test and treatment versions. The allergy test syringe has an intra-dermal bevel for intra-dermal injections. The allergy treatment syringe has a regular bevel which is typically used for subcutaneous injections. The tuberculin syringe has a regular bevel which can be used for any of the 3 types of common injections (intra-dermal, intra-muscular or subcutaneous).

The SafetyGlide™ Syringe contains a mechanism that covers the needle point after use. In the activated position the needle cover guards against accidental needle sticks during normal handling and disposal of the used needle/syringe combination.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K992734: Becton Dickinson Syringe.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K992734

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026