510(k) Report

510(k) · Cleared

Recalled · Class II · 2013

Elecsys Serum B-Crosslaps/Serum Immunoassay

K993706510(k) TraditionalCleared Jul 24, 2000265-day reviewClass I21 CFR 862.1400Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Roche Diagnostics Corp. · JMM Column Chromatography & Color Development, Hydroxyproline

Cleared on Jul 24, 2000 as substantially equivalent to K990843 (Osteometer Biotech A/S, 1999) after a 265-day review. Its family tree reaches back 3 generations to 1996. 5 later devices descend from it. It has been recalled once.

Decided
Jul 24, 2000
Review
265 days
Class
Class I
Product code
JMM
Recalls
1
Predicates
1

Its longest line of descent runs back 3 generations to K960171 (1996). 5 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2005: 1 descendant cleared2011: 1 descendant cleared2013: 1 descendant cleared2018: 1 descendant cleared2026: 1 descendant clearedK960171 · Active Crosslaps Elisa · Diagnostic Systems Laboratories, Inc. · 1996K972788 · Crosslaps Elisa · Devices and Diagnostics Consulting Group · 1998K990843 · Serum Crosslaps One Step Elisa · Osteometer Biotech A/S · 19991995200020052010201520202025
Descendants, by generation 1 generation forward: 1 descendant2 generations forward: 1 descendant3 generations forward: 1 descendant4 generations forward: 2 descendants K990843 · Serum Crosslaps One Step Elisa · Osteometer Biotech A/S · 1999K972788 · Crosslaps Elisa · Devices and Diagnostics Consulting Group · 1998K960171 · Active Crosslaps Elisa · Diagnostic Systems Laboratories, Inc. · 19960
  1. Root · 3 generations backK960171 Active Crosslaps ElisaDiagnostic Systems Laboratories, Inc. · 1996
  2. 2 generations backK972788 Crosslaps ElisaDevices and Diagnostics Consulting Group · 1998
  3. 1 generation backK990843 Serum Crosslaps One Step ElisaOsteometer Biotech A/S · 1999
  4. This deviceK993706 Elecsys Serum B-Crosslaps/Serum ImmunoassayRoche Diagnostics Corp. · 2000
  5. Descendant · cites it directly · 4 of its ownK051543 Elecsys Precicontrol BoneRoche Diagnostics Corp. · 2005
  6. Descendant · 2 generations on · 3 of its ownK111506 Elecsys Precicontrol Varia 3Roche Diagnostics · 2011 · Recalled
  7. Descendant · 3 generations on · 2 of its ownK132828 Elecsys Calcitonin; Immunoassay, Calset, Calcheck 5; Elecsys…Roche Diagnostics · 2013
  8. Descendant · 4 generations on · 0 of its ownK252431 Elecsys CalcitoninRoche Diagnostics GmbH · 2026
  9. Descendant · 4 generations on · 0 of its ownK182012 ADVIA Centaur Calcitonin (CALCT) assayAxis-Shield Diagnostics Limited · 2018

Most influential descendants

Showing 5 of 5 · See all
  • K051543Elecsys Precicontrol BoneRoche Diagnostics Corp. · 2005 · cites it directly4descendants
  • K111506Elecsys Precicontrol Varia 3Roche Diagnostics · 2011 · 2 generations on3descendants
  • K132828Elecsys Calcitonin; Immunoassay, Calset, Calcheck 5; Elecsys Precicontrol Varia 3Roche Diagnostics · 2013 · 3 generations on2descendants
  • K252431Elecsys CalcitoninRoche Diagnostics GmbH · 2026 · 4 generations on0descendants
  • K182012ADVIA Centaur Calcitonin (CALCT) assayAxis-Shield Diagnostics Limited · 2018 · 4 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K960171Active Crosslaps ElisaDiagnostic Systems Laboratories, Inc.1996—
−2K972788Crosslaps ElisaDevices and Diagnostics Consulting Group1998—
−1K990843Serum Crosslaps One Step ElisaOsteometer Biotech A/S1999—
0 (this device)K993706Elecsys Serum B-Crosslaps/Serum ImmunoassayRoche Diagnostics Corp.2000—
+1K051543Elecsys Precicontrol BoneRoche Diagnostics Corp.2005—
+2K111506Elecsys Precicontrol Varia 3Roche Diagnostics20111
+3K132828Elecsys Calcitonin; Immunoassay, Calset, Calcheck 5; Elecsys Precicontrol…Roche Diagnostics2013—
+4K252431Elecsys CalcitoninRoche Diagnostics GmbH2026—
+4K182012ADVIA Centaur Calcitonin (CALCT) assayAxis-Shield Diagnostics Limited2018—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200510
201110
201310
201810
202610
FIG. 1 — Lineage of K993706, 1996–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JMM? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K990843Serum Crosslaps One Step Elisa
Summary page 1
…/serum Immunoassay is substantially _equivalent to other devices marketed in the United States. We claim equivalence to the Osteometer Serum CrossLaps™ One Step ELISA (K990843). Device name Proprietary name ELECSYS® B-CrossLaps/serum Immunoassay Common name Electrochemiluminescence Immunoassay, B-Cros…
Osteometer Biotech A/S1999—

What descends from it? (5)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K051543Elecsys Precicontrol BoneRoche Diagnostics Corp.20054

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 25, 2013Class IIZ-1727-2013Device DesignRoche Diagnostics internal investigations in R&D have shown that the claims for biotin interference and the sample stability given in the Elecsys ¿-CrossLaps (¿-CL) assay package insert version 5, could not be reproduced. The ¿-CrossLaps…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 4 years and passes through 4 companies; a new generation was cleared every 1.0 years on median.
  • None of the 3 ancestors has a recall.
  • Of its descendants, K051543 (Roche Diagnostics Corp., 2005) has gone furthest, with 4 descendants of its own.

How does its category compare?

All JMM devices

JMM Column Chromatography & Color Development, Hydroxyproline: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JMM

FDA has decided no JMM device since 2007.

Median 510(k) review, days · JMM

Too few JMM 510(k)s in any year since 2007 to give a median.

Recalls per year · JMM

1in 2012

2012: 1
2007high 1 (2012) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JMM

1in 2025

2019: 12021: 12023: 12024: 12025: 1
2007high 1 (2019) · low 0 (2007)2025

MAUDE reports naming product code JMM, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20110—00
20120—10
20130—00
20140—00
20150—00
20160—00
20170—00
20180—00
20190—01
20200—00
20210—01
20220—00
20230—01
20240—01
20250—01
20260—0—

What is it indicated for?

For the in vitro quantitative determination of degradation products of type I collagen in human serum and plasma, for assessing individual bone resorption. The test may be used as an aid in monitoring anti-resorptive therapies (e.g. bisphosphonates, hormone replacement therapy — HRT) in postmenopausal women and individuals diagnosed with osteopenia. The elecrochemiluminescence immunoassay “ECLIA” is intended for use on the Roche ELECSYS 1010 and 2010 immunoassay analyzers.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K993706: Elecsys Serum B-Crosslaps/Serum Immunoassay.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K993706

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026