510(k) Report

510(k) · Cleared

Cleared

Syva Emit II Plus Cocaine Metabolite Assay, Model 9H029UL

K993988510(k) TraditionalCleared Jan 27, 200064-day reviewClass II21 CFR 862.3250

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Syva Co. · DIO Enzyme Immunoassay, Cocaine And Cocaine Metabolites

Cleared on Jan 27, 2000 after a 64-day review. The FDA summary names its predicate without a number. 6 later devices descend from it.

Decided
Jan 27, 2000
Review
64 days
Class
Class II
Product code
DIO
Recalls
None
Predicates
0

6 later devices descend from it, 2 citing it directly.

Descendants, by year cleared 2003: 1 descendant cleared2017: 1 descendant cleared2006: 2 descendants cleared2004: 1 descendant cleared2011: 1 descendant cleared200020052010201520202025
  1. This deviceK993988 Syva Emit II Plus Cocaine Metabolite Assay, Model 9H029ULSyva Co. · 2000
  2. Descendant · cites it directly · 4 of its ownK031512 Emit II Plus Cocaine Metabolite AssayDade Behring, Inc. · 2003
  3. Descendant · 2 generations on · 2 of its ownK033713 Urine Cocaine Metabolite Screen Flex Reagent Cartridge, Catalog# DF92ADade Behring, Inc. · 2004
  4. Descendant · 3 generations on · 1 of its ownK062128 The Dimension Vista Flex Reagent Cartridges / Kit AssaysDade Behring, Inc. · 2006
  5. Descendant · cites it directly · 0 of its ownK170293 Emit II Plus Cocaine Metabolite AssaySiemens Healthcare Diagnostics, Inc. · 2017
  6. Descendant · 4 generations on · 0 of its ownK102045 Dimension Visa HBA1C, Model K3105ASiemens Heatlhcare Diagnostics, Inc. · 2011 · Recalled

Showing 5 of 6 descendants. See all descendants

Descendants cleared per year

2003: 12004: 12006: 22011: 12017: 1
2003high 2 (2006) · low 0 (2005)2026

Most influential descendants

Showing 5 of 6 · See all
  • K031512Emit II Plus Cocaine Metabolite AssayDade Behring, Inc. · 2003 · cites it directly4descendants
  • K033713Urine Cocaine Metabolite Screen Flex Reagent Cartridge, Catalog# DF92ADade Behring, Inc. · 2004 · 2 generations on2descendants
  • K062128The Dimension Vista Flex Reagent Cartridges / Kit AssaysDade Behring, Inc. · 2006 · 3 generations on1descendant
  • K170293Emit II Plus Cocaine Metabolite AssaySiemens Healthcare Diagnostics, Inc. · 2017 · cites it directly0descendants
  • K102045Dimension Visa HBA1C, Model K3105ASiemens Heatlhcare Diagnostics, Inc. · 2011 · 4 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K993988Syva Emit II Plus Cocaine Metabolite Assay, Model 9H029ULSyva Co.2000—
+1K031512Emit II Plus Cocaine Metabolite AssayDade Behring, Inc.2003—
+2K033713Urine Cocaine Metabolite Screen Flex Reagent Cartridge, Catalog# DF92ADade Behring, Inc.2004—
+3K062128The Dimension Vista Flex Reagent Cartridges / Kit AssaysDade Behring, Inc.2006—
+1K170293Emit II Plus Cocaine Metabolite AssaySiemens Healthcare Diagnostics, Inc.2017—
+4K102045Dimension Visa HBA1C, Model K3105ASiemens Heatlhcare Diagnostics, Inc.20113

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200310
200410
200620
201110
201710
FIG. 1 — Lineage of K993988, 2000–2017Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DIO? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DIO device

  1. #1 K213211 Carolina Liquid Chemistries Cocaine and…score 3.0 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—Emit® II Cocaine Metabolite Assay
Named without a submission number
—

What descends from it? (6)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K031512Emit II Plus Cocaine Metabolite AssayDade Behring, Inc.20034
K170293Emit II Plus Cocaine Metabolite AssaySiemens Healthcare Diagnostics, Inc.20170

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K031512 (Dade Behring, Inc., 2003) has gone furthest, with 4 descendants of its own.

How does its category compare?

All DIO devices

DIO Enzyme Immunoassay, Cocaine And Cocaine Metabolites: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DIO

1in 2022

2009: 12011: 52012: 52013: 22014: 22015: 22016: 32017: 32018: 12020: 12021: 12022: 1
2007high 5 (2011) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DIO

267in 2017

2011: 65 days median2012: 139 days median2016: 88 days median2017: 267 days median
2007high 267 (2017) · low 65 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DIO

1in 2013

2013: 1
2007high 1 (2013) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DIO

1in 2023

2011: 42014: 12015: 72016: 92017: 162018: 122019: 222022: 12023: 1
2007high 22 (2019) · low 0 (2007)2025

MAUDE reports naming product code DIO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20091—00
20100—00
201156504
2012513900
20132—10
20142—01
20152—07
201638809
20173267016
20181—012
20190—022
20201—00
20211—00
20221—01
20230—01
20240—00
20250—00
20260—0—

What is it indicated for?

The Emit® II Plus Cocaine Metabolite Assay is a homogeneous drugs-of-abuse enzyme immunoassay with a 150 ng/mL or 300 ng/mL cutoff (SAMSHA initial test cutoff level). The assay is intended for use in the qualitative and semiquantitative analyses of benzoylecgonine (cocaine metabolite) in human urine. Emit® II Plus assays are designed for use with a number of chemistry analyzers. The Emit® II Plus Cocaine Metabolite Assay provides only a preliminary…

Read the full statement (123 words)

The Emit® II Plus Cocaine Metabolite Assay is a homogeneous drugs-of-abuse enzyme immunoassay with a 150 ng/mL or 300 ng/mL cutoff (SAMSHA initial test cutoff level). The assay is intended for use in the qualitative and semiquantitative analyses of benzoylecgonine (cocaine metabolite) in human urine. Emit® II Plus assays are designed for use with a number of chemistry analyzers.

The Emit® II Plus Cocaine Metabolite Assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgement should be applied to any drug-of-abuse test result, particularly when preliminary positive results are used.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K993988: Syva Emit II Plus Cocaine Metabolite Assay, Model 9H029UL.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K993988

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026