510(k) Report

510(k) · Cleared

Recalled · Class II · 2018

Autostat II Rheumatoid Factor Igm Elisa

K994338510(k) TraditionalCleared Feb 3, 200042-day reviewClass II21 CFR 866.5775Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Cogent Diagnotics, Ltd. · DHR System, Test, Rheumatoid Factor

Cleared on Feb 3, 2000 after a 42-day review. FDA’s file for it is a 510(k) statement, which names no predicate. It has been recalled 2 times.

Decided
Feb 3, 2000
Review
42 days
Class
Class II
Product code
DHR
Recalls
2
Predicates
0

FDA’s file for it is a 510(k) statement, which names no predicate.

FIG. 1 — Lineage of K994338Source: FDA 510(k) database · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DHR? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against?

FDA’s file for it is a 510(k) statement, which names no predicate.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 19, 2018Class IIZ-2200-2019Labeling designReports when using Rheumatoid Factor test kits, control and negative samples may be elevated following a sample containing high levels of IgG . The negative control may report out of range; a negative sample may recover as a low positive…
Feb 21, 2018Class IIZ-2134-2019Nonconforming Material/ComponentReports of higher than anticipated positivity rate for patients tested using IgM Kit. Some negative samples may be reported as positive for IgM antibodies to Rheumatoid Factor.

No ancestor or descendant has a Class I recall.

How does its category compare?

All DHR devices

DHR System, Test, Rheumatoid Factor: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DHR

1in 2020

2007: 12009: 12010: 12011: 12015: 12016: 12017: 12018: 12019: 12020: 1
2007high 1 (2007) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DHR

Too few DHR 510(k)s in any year since 2007 to give a median.

Recalls per year · DHR

1in 2019

2010: 12011: 32012: 22018: 22019: 1
2007high 3 (2011) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DHR

2in 2025

2010: 192011: 4002012: 912018: 12024: 32025: 2
2007high 400 (2011) · low 0 (2007)2025

MAUDE reports naming product code DHR, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—00
20080—00
20091—00
20101—119
20111—3400
20120—291
20130—00
20140—00
20151—00
20161—00
20171—00
20181—21
20191—10
20201—00
20210—00
20220—00
20230—00
20240—03
20250—02
20260—0—

What is it indicated for?

Enzyme Linked Immunosorbent Test for the qualitative and quantitative determination of specific IgM rheumatoid factor (RF) antibodies in human serum. The results of an IgM assay are used in conjunction with other diagnostic tests and clinical investigations. The assay may be used on the [unclear] automated EIA instrument. In vitro diagnostic use only.

Indications for use read by OCR from a scanned FDA 510(k) statement, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K994338: Autostat II Rheumatoid Factor Igm Elisa.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K994338

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Page last updated Nov 3, 2026