510(k) Report

PMA · Approved

Recalled · Class II · 2024

Ams Sphincter 800 Urinary Control System

P000053PMAApproved Jun 14, 2001177-day reviewClass III21 CFR 876.5280Recalls 6

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Boston Scientific Corp · EZY Device, Incontinence, Mechanical/Hydraulic

Approved on Jun 14, 2001 through premarket approval, which rests on the device’s own clinical evidence after a 177-day review. It has been recalled 6 times.

Decided
Jun 14, 2001
Review
177 days
Class
Class III
Product code
EZY
Recalls
6

PMA approvals rest on their own clinical evidence and do not claim a predicate, so this family tree starts here.

P000053 has 135 supplements since 2003: changes to the approved device, its labelling or its manufacturing, each approved separately.

PMA supplements approved per year

2003: 12004: 32006: 22007: 62008: 62009: 72010: 62011: 42012: 82013: 22014: 42015: 82016: 132017: 122018: 122019: 102020: 102021: 22022: 62023: 42024: 32025: 42026: 2
2003high 13 (2016) · low 0 (2005)2026
View FIG. 1 as a table
YearSupplements approved
20031
20043
20062
20076
20086
20097
20106
20114
20128
20132
20144
20158
201613
201712
201812
201910
202010
20212
20226
20234
20243
20254
20262
FIG. 1 — Supplements to P000053, 2003–2026Source: FDA PMA database

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA PMA database (decision fields and supplements), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in EZY? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

Does it have a predicate?

PMA decisions do not claim a predicate device.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 28, 2024Class IIZ-0531-2025Error in labelingPotential for mislabeled packaging in which a device labeled as a Pressure Regulating Balloon (UPN 72400024) which instead contained a Control Pump (UPN 72400098)
Oct 28, 2024Class IIZ-0530-2025Error in labelingPotential for mislabeled packaging in which a device labeled as a Pressure Regulating Balloon (UPN 72400024) which instead contained a Control Pump (UPN 72400098)
Jul 19, 2019Class IIZ-2442-2019Error in labelingMislabled: AMS 800 occlusive cuff labeled as 4.5 cm was a length of 4.0 cm and therefore..
Mar 1, 2013Class IIZ-1280-2013Release of Material/Component prior to receiving test resultsAmerican Medical Systems, Inc. is initiating a recall on two components of the AMS 800¿ Urinary Control System, Control Pump & Pressure-Regulating Balloon. These products were released despite not meeting all final product release criteria.
Mar 1, 2013Class IIZ-1279-2013Release of Material/Component prior to receiving test resultsAmerican Medical Systems, Inc. is initiating a recall on two components of the AMS 800¿ Urinary Control System, Control Pump & Pressure-Regulating Balloon. These products were released despite not meeting all final product release criteria.
May 9, 2011Not classifiedZ-2669-2011Nonconforming Material/ComponentAmerican Medical Systems is recalling all unexpired AMS 800 and AMS Acticon Neosphincter Control Pumps. Based upon a review of our product test procedures, we are unable to confirm that all Control Pumps have met our requirements. Test…

No ancestor or descendant has a Class I recall.

How does its category compare?

All EZY devices

EZY Device, Incontinence, Mechanical/Hydraulic: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · EZY

6in 2025

2007: 62008: 62009: 72010: 62011: 42012: 82013: 22014: 42015: 92016: 142017: 122018: 122019: 102020: 102021: 32022: 62023: 42024: 32025: 62026 (partial year): 4
2007high 14 (2016) · low 2 (2013)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · EZY

Too few EZY 510(k)s in any year since 2007 to give a median.

Recalls per year · EZY

2in 2024

2011: 12021: 12024: 2
2007high 2 (2024) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · EZY

2,066in 2025

2007: 72008: 22009: 142010: 252011: 262012: 232013: 222014: 92015: 132016: 62017: 92018: 8732019: 2,7532020: 1,7432021: 1,6972022: 1,5532023: 1,7782024: 2,1102025: 2,066
2007high 2,753 (2019) · low 2 (2008)2025

MAUDE reports naming product code EZY, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20076—07
20086—02
20097—014
20106—025
20114—126
20128—023
20132—022
20144—09
20159—013
201614—06
201712—09
201812—0873
201910—02,753
202010—01,743
20213—11,697
20226—01,553
20234—01,778
20243—22,110
20256—02,066
20264—0—
Cite this page

510(k) Report. “P000053: Ams Sphincter 800 Urinary Control System.” Lineage record from FDA PMA data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/P000053

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Page last updated Nov 3, 2026