Recalled · Class I · 2025

Concerto/Insync Sentry/Insync Maximo Implantable Cardiovascular Defibrillators with Cardiac Resynchicnization

P010031PMAApproved Jun 26, 2002418-day reviewClass IIIRecalls 11 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medtronic Cardiac Rhythm Disease Management · NIK Defibrillator, Automatic Implantable Cardioverter, With Cardiac Resynchronization (Crt-D)

Approved on Jun 26, 2002 through premarket approval, which rests on the device’s own clinical evidence after a 418-day review. It has a Class I recall. It has been recalled 11 times, including a Class I recall.

Decided
Jun 26, 2002
Review
418 days
Class
Class III
Product code
NIK
Recalls
11 (1 Class I)

PMA approvals rest on their own clinical evidence and do not claim a predicate, so this family tree starts here.

P010031 has 902 supplements since 2002: changes to the approved device, its labelling or its manufacturing, each approved separately.

PMA supplements approved per year

2002: 22003: 82004: 72005: 92006: 252007: 302008: 412009: 412010: 442011: 612012: 522013: 692014: 672015: 412016: 492017: 442018: 412019: 302020: 412021: 402022: 382023: 342024: 322025: 352026: 21
2002high 69 (2013) · low 2 (2002)2026
View FIG. 1 as a table
YearSupplements approved
20022
20038
20047
20059
200625
200730
200841
200941
201044
201161
201252
201369
201467
201541
201649
201744
201841
201930
202041
202140
202238
202334
202432
202535
202621
FIG. 1 — Supplements to P010031, 2002–2026Source: FDA PMA database

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA PMA database (decision fields and supplements), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

Does it have a predicate?

PMA decisions do not claim a predicate device.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 17, 2025Class IIZ-2128-2025Software designA software update for Medtronic CareLink SmartSync Device Manager (SmartSync) is now available. The update eliminates the potential display of an erroneous electrical reset (pop-up) message upon interrogation with a SmartSync Device…
Jun 17, 2025Class IIZ-2127-2025Software designA software update for Medtronic CareLink SmartSync Device Manager (SmartSync) is now available. The update eliminates the potential display of an erroneous electrical reset (pop-up) message upon interrogation with a SmartSync Device…
Apr 6, 2021Class IIZ-1528-2021Software designA longevity estimation error may occur in the early years of device life when a unipolar pacing vector is programmed in the right atrial (RA) lead and/or the right ventricular (RV) lead. No other device features or therapies are impacted.
Apr 6, 2021Class IIZ-1526-2021Software designA longevity estimation error may occur in the early years of device life when a unipolar pacing vector is programmed in the right atrial (RA) lead and/or the right ventricular (RV) lead. No other device features or therapies are impacted.
Apr 6, 2021Class IIZ-1527-2021Software designA longevity estimation error may occur in the early years of device life when a unipolar pacing vector is programmed in the right atrial (RA) lead and/or the right ventricular (RV) lead. No other device features or therapies are impacted.
Apr 6, 2021Class IIZ-1525-2021Software designA longevity estimation error may occur in the early years of device life when a unipolar pacing vector is programmed in the right atrial (RA) lead and/or the right ventricular (RV) lead. No other device features or therapies are impacted.
Feb 3, 2021Class IZ-1306-2021Device DesignA small percentage of implanted cardiac devices, from a well-defined subset, may experience a shortened Recommended Replacement Time (RRT) to End of Service (EOS) interval following an earlier-than-expected RRT observation
Nov 15, 2020Class IIZ-0928-2021Process controlUnder rare circumstances, the component may be susceptible to a failure mechanism that could impact device performance, including potential loss of function.
Nov 15, 2020Class IIZ-0926-2021Process controlUnder rare circumstances, the component may be susceptible to a failure mechanism that could impact device performance, including potential loss of function.
Nov 15, 2020Class IIZ-0927-2021Process controlUnder rare circumstances, the component may be susceptible to a failure mechanism that could impact device performance, including potential loss of function.
Jul 25, 2008Not classifiedZ-0667-2009Process designMaximo II VR ICD, DR ICD and CRT-D inadvertently include the Capture Management algorithm (Atrial/Right Ventricular/Left Ventricular Capture management). The Capture management algorithm monitors daily pacing thresholds and may increase…

No ancestor or descendant has a Class I recall.

How does its category compare?

All NIK devices

NIK Defibrillator, Automatic Implantable Cardioverter, With Cardiac Resynchronization (Crt-D): FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · NIK

56in 2025

2007: 532008: 622009: 432010: 512011: 502012: 832013: 502014: 762015: 492016: 1092017: 1122018: 912019: 622020: 752021: 672022: 622023: 582024: 562025: 562026 (partial year): 44
2007high 112 (2017) · low 43 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NIK

Too few NIK 510(k)s in any year since 2007 to give a median.

Recalls per year · NIK

1in 2022

2007: 92008: 12009: 22010: 72012: 22016: 2, 1 Class I2017: 9, 5 Class I2020: 32022: 1
2007high 9 (2007) · low 0 (2011)2026

Class I (6 since 2007) Class II/III not classified

Adverse event reports per year · NIK

10,855in 2025

2007: 9242008: 3,0212009: 4,1772010: 5,6682011: 5,9442012: 5,5822013: 7,3732014: 10,7312015: 7,8282016: 13,2552017: 16,2652018: 8,3482019: 7,2662020: 9,6092021: 10,1342022: 10,6462023: 11,8972024: 11,6972025: 10,855
2007high 16,265 (2017) · low 924 (2007)2025

MAUDE reports naming product code NIK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200753—9924
200862—13,021
200943—24,177
201051—75,668
201150—05,944
201283—25,582
201350—07,373
201476—010,731
201549—07,828
2016109—2 (1)13,255
2017112—9 (5)16,265
201891—08,348
201962—07,266
202075—39,609
202167—010,134
202262—110,646
202358—011,897
202456—011,697
202556—010,855
202644—0—
Cite this page

510(k) Report. “P010031: Concerto/Insync Sentry/Insync Maximo Implantable Cardiovascular Defibrillators with Cardiac Resynchicnization.” Lineage record from FDA PMA data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/P010031