510(k) Report

PMA · Approved

Recalled · Class I · 2026

Contak Renewal Models H125 and H120 with the Model 2865 Version 1.8 Application Software

P030005PMAApproved Jan 26, 2004308-day reviewClass IIIRecalls 20 (11 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Guidant Corp. · NKE Pulse Generator, Pacemaker, Implantable, With Cardiac Resynchronization (Crt-P)

Approved on Jan 26, 2004 through premarket approval, which rests on the device’s own clinical evidence after a 308-day review. It has a Class I recall. It has been recalled 20 times, including 11 Class I recalls.

Decided
Jan 26, 2004
Review
308 days
Class
Class III
Product code
NKE
Recalls
20 (11 Class I)

PMA approvals rest on their own clinical evidence and do not claim a predicate, so this family tree starts here.

P030005 has 244 supplements since 2004: changes to the approved device, its labelling or its manufacturing, each approved separately.

PMA supplements approved per year

2004: 52005: 162006: 112007: 92008: 62009: 52010: 102011: 92012: 112013: 122014: 102015: 152016: 162017: 232018: 82019: 142020: 132021: 112022: 112023: 82024: 92025: 62026: 6
2004high 23 (2017) · low 5 (2004)2026
View FIG. 1 as a table
YearSupplements approved
20045
200516
200611
20079
20086
20095
201010
20119
201211
201312
201410
201515
201616
201723
20188
201914
202013
202111
202211
20238
20249
20256
20266
FIG. 1 — Supplements to P030005, 2004–2026Source: FDA PMA database

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA PMA database (decision fields and supplements), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

Does it have a predicate?

PMA decisions do not claim a predicate device.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 19, 2026Class IZ-1771-2026Nonconforming Material/ComponentUpdated software, Brady software maintenance release 6 (SMR6), is now available for the ACCOLADE" family of pacemakers and cardiac resynchronization therapy pacemakers (CRT-Ps)1 and the advisory population is expanding to include all…
Aug 20, 2025Class IZ-0103-2026OtherSoftware to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT…
Aug 20, 2025Class IZ-0101-2026OtherSoftware to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT…
Aug 20, 2025Class IZ-0104-2026OtherSoftware to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT…
Aug 20, 2025Class IZ-0105-2026OtherSoftware to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT…
Aug 20, 2025Class IZ-0102-2026OtherSoftware to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT…
Dec 12, 2024Class IZ-0826-2025Process designA subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery…
Dec 12, 2024Class IZ-0827-2025Process designA subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery…
Jun 3, 2021Class IZ-2019-2021Device DesignThe dual chamber INGENIO family pacemakers or cardiac resynchronization therapy pacemakers (CRT-Ps) may initiate Safety Mode later in device life (i.e., prior to reaching the Explant battery indicator) when the device's battery exhibits…
Jun 3, 2021Class IZ-2018-2021Device DesignThe dual chamber INGENIO family pacemakers or cardiac resynchronization therapy pacemakers (CRT-Ps) may initiate Safety Mode later in device life (i.e., prior to reaching the Explant battery indicator) when the device's battery exhibits…
Jun 3, 2021Class IZ-2020-2021Device DesignThe dual chamber INGENIO family pacemakers or cardiac resynchronization therapy pacemakers (CRT-Ps) may initiate Safety Mode later in device life (i.e., prior to reaching the Explant battery indicator) when the device's battery exhibits…
Jun 3, 2021Class IIZ-2002-2021Nonconforming Material/ComponentLatent release of small amounts of hydrogen within the pacemaker may cause a low voltage capacitor to become electrically compromised over time resulting in accelerated battery depletion of the battery and associated progression of…
Jun 3, 2021Class IIZ-2003-2021Nonconforming Material/ComponentLatent release of small amounts of hydrogen within the pacemaker may cause a low voltage capacitor to become electrically compromised over time resulting in accelerated battery depletion of the battery and associated progression of…
Sep 10, 2018Class IIZ-3259-2018Component design/selectionBoston Scientific has identified a subset of active pacemaker and CRT-Ps systems with an elevated potential for early pacemaker replacement due to hydrogen-induced accelerated battery depletion.
Sep 10, 2018Class IIZ-3258-2018Component design/selectionBoston Scientific has identified a subset of active pacemaker and CRT-Ps systems with an elevated potential for early pacemaker replacement due to hydrogen-induced accelerated battery depletion.
Dec 7, 2017Class IIZ-0365-2018Device DesignBoston Scientific has received reports of intermittent over-sensing of the Minute Ventilation (MV) sensor signal with certain Boston Scientific pacemaker and cardiac resynchronization therapy pacemaker systems (pacemakers).
Dec 7, 2017Class IIZ-0360-2018Device DesignBoston Scientific has received reports of intermittent over-sensing of the Minute Ventilation (MV) sensor signal with certain Boston Scientific pacemaker and cardiac resynchronization therapy pacemaker systems (pacemakers).
Dec 7, 2017Class IIZ-0361-2018Device DesignBoston Scientific has received reports of intermittent over-sensing of the Minute Ventilation (MV) sensor signal with certain Boston Scientific pacemaker and cardiac resynchronization therapy pacemaker systems (pacemakers).
Dec 7, 2017Class IIZ-0366-2018Device DesignBoston Scientific has received reports of intermittent over-sensing of the Minute Ventilation (MV) sensor signal with certain Boston Scientific pacemaker and cardiac resynchronization therapy pacemaker systems (pacemakers).
Jun 24, 2006Not classifiedZ-1298-06OtherGuidant has determined that low-voltage capacitors from a single component supplier may malfunction in a manner that can lead to premature battery depletion without warning in the affected devices.

No ancestor or descendant has a Class I recall.

How does its category compare?

All NKE devices

NKE Pulse Generator, Pacemaker, Implantable, With Cardiac Resynchronization (Crt-P): FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · NKE

22in 2025

2007: 192008: 242009: 222010: 272011: 292012: 192013: 442014: 542015: 582016: 332017: 482018: 292019: 292020: 302021: 262022: 242023: 272024: 232025: 222026 (partial year): 19
2007high 58 (2015) · low 19 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NKE

Too few NKE 510(k)s in any year since 2007 to give a median.

Recalls per year · NKE

5in 2025

2015: 12017: 72024: 2, 2 Class I2025: 5, 5 Class I2026 (partial year): 1, 1 Class I
2007high 7 (2017) · low 0 (2007)2026

Class I (8 since 2007) Class II/III not classified

Adverse event reports per year · NKE

1,939in 2025

2007: 372008: 522009: 812010: 1692011: 1962012: 1642013: 2072014: 1272015: 3562016: 5142017: 6072018: 1,0042019: 7942020: 8062021: 1,1772022: 1,2902023: 1,4692024: 1,5182025: 1,939
2007high 1,939 (2025) · low 37 (2007)2025

MAUDE reports naming product code NKE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200719—037
200824—052
200922—081
201027—0169
201129—0196
201219—0164
201344—0207
201454—0127
201558—1356
201633—0514
201748—7607
201829—01,004
201929—0794
202030—0806
202126—01,177
202224—01,290
202327—01,469
202423—2 (2)1,518
202522—5 (5)1,939
202619—1 (1)—
Cite this page

510(k) Report. “P030005: Contak Renewal Models H125 and H120 with the Model 2865 Version 1.8 Application Software.” Lineage record from FDA PMA data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/P030005

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Page last updated Nov 3, 2026