510(k) Report

PMA · Approved

Recalled · Class II · 2017

Endologix Powerlink System

P040002PMAApproved Oct 29, 2004295-day reviewClass IIIRecalls 16

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Endologix, LLC · MIH System, Endovascular Graft, Aortic Aneurysm Treatment

Approved on Oct 29, 2004 through premarket approval, which rests on the device’s own clinical evidence after a 295-day review. 1 of the 9 devices in its family, counting itself, has a Class I recall. 8 later devices descend from it. It has been recalled 16 times.

Decided
Oct 29, 2004
Review
295 days
Class
Class III
Product code
MIH
Recalls
16

8 later devices descend from it, 1 citing it directly. 1 of the 9 devices in this family, counting this one, has a Class I recall.

Descendants, by year cleared 2011: 2 descendants cleared2012: 1 descendant cleared, 1 with a Class I recall12015: 1 descendant cleared2016: 1 descendant cleared2017: 1 descendant cleared2018: 1 descendant cleared2019: 1 descendant cleared200020052010201520202025
  1. This deviceP040002 Endologix Powerlink SystemEndologix, LLC · 2004
  2. Descendant · cites it directly · 7 of its ownK110090 Afx Introducer SystemEndologix, Inc. · 2011
  3. Descendant · 2 generations on · 6 of its ownK111747 Endologix Afx Introducer SystemEndologix, Inc. · 2011
  4. Descendant · 3 generations on · 5 of its ownK120212 Afx Introducer SystemEndologix, Inc. · 2012 · Recalled (Class I)
  5. Descendant · 4 generations on · 4 of its ownK141275 Endophys Pressure Sensing Sheath KitEndophys, Inc. · 2015
  6. Descendant · 5 generations on · 3 of its ownK160272 Endophys Pressure Sensing Sheath KitEndophys, Inc. · 2016

Showing 5 of 8 descendants. See all descendants

Descendants cleared per year

2011: 22012: 12015: 12016: 12017: 12018: 12019: 1
2011high 2 (2011) · low 0 (2013)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 8 · See all
  • K110090Afx Introducer SystemEndologix, Inc. · 2011 · cites it directly7descendants
  • K111747Endologix Afx Introducer SystemEndologix, Inc. · 2011 · 2 generations on6descendants
  • K120212Afx Introducer SystemEndologix, Inc. · 2012 · 3 generations on5descendants
  • K141275Endophys Pressure Sensing Sheath KitEndophys, Inc. · 2015 · 4 generations on4descendants
  • K160272Endophys Pressure Sensing Sheath KitEndophys, Inc. · 2016 · 5 generations on3descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)P040002Endologix Powerlink SystemEndologix, LLC2004—
+1K110090Afx Introducer SystemEndologix, Inc.2011—
+2K111747Endologix Afx Introducer SystemEndologix, Inc.2011—
+3K120212Afx Introducer SystemEndologix, Inc.2012Class I
+4K141275Endophys Pressure Sensing Sheath KitEndophys, Inc.2015—
+5K160272Endophys Pressure Sensing Sheath KitEndophys, Inc.2016—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201120
201211
201510
201610
201710
201810
201910
FIG. 1 — Lineage of P040002, 2004–2019Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA PMA database (decision fields and supplements), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

Does it have a predicate?

PMA decisions do not claim a predicate device.

What descends from it? (8)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K110090Afx Introducer SystemEndologix, Inc.20117

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 4, 2017Class IIZ-1049-2017Device DesignEndologix updated information on the rates of Type III endoleaks and suggestions for patient surveillance and treatment.
Jan 4, 2017Class IIZ-1048-2017Device DesignEndologix updated information on the rates of Type III endoleaks and suggestions for patient surveillance and treatment.
Dec 27, 2016Class IIZ-1038-2017OtherEndologix updated the Instructions for Use (IFU) and implemented modifications to the AFX Endovascular AAA System (AFX System).
Dec 27, 2016Class IIZ-1047-2017Nonconforming Material/ComponentEndologix updated the Instructions for Use (IFU) and implemented modifications to the AFX Endovascular AAA System (AFX System).
Dec 27, 2016Class IIZ-1037-2017Device DesignEndologix updated the Instructions for Use (IFU) and implemented modifications to the AFX Endovascular AAA System (AFX System).
Dec 27, 2016Class IIZ-1035-2017OtherEndologix initiated a correction by updating reports for the cumulative clinical data and information on the Powerlink System and the AFX Endovascular AAA System (AFX System).
Dec 27, 2016Class IIZ-1039-2017Device DesignEndologix updated the Instructions for Use (IFU) and implemented modifications to the AFX Endovascular AAA System (AFX System).
Dec 27, 2016Class IIZ-1054-2017Under Investigation by firmEndologix updated the Instructions for Use (IFU) and implemented modifications to the AFX Endovascular AAA System (AFX System).
Dec 14, 2005Not classifiedZ-0590-06OtherThe tip may separate from the catheter sheath inner core during insertion of the graft. The delivery catheter is being recalled. The implanted devices are not affected.
Dec 14, 2005Not classifiedZ-0591-06OtherThe tip may separate from the catheter sheath inner core during insertion of the graft. The delivery catheter is being recalled. The implanted devices are not affected.
Dec 14, 2005Not classifiedZ-0592-06OtherThe tip may separate from the catheter sheath inner core during insertion of the graft. The delivery catheter is being recalled. The implanted devices are not affected.
Dec 14, 2005Not classifiedZ-0594-06OtherThe tip may separate from the catheter sheath inner core during insertion of the graft. The delivery catheter is being recalled. The implanted devices are not affected.
Dec 14, 2005Not classifiedZ-0595-06OtherThe tip may separate from the catheter sheath inner core during insertion of the graft. The delivery catheter is being recalled. The implanted devices are not affected.
Dec 14, 2005Not classifiedZ-0589-06OtherThe tip may separate from the catheter sheath inner core during insertion of the graft. The delivery catheter is being recalled. The implanted devices are not affected.
Dec 14, 2005Not classifiedZ-0596-06OtherThe tip may separate from the catheter sheath inner core during insertion of the graft. The delivery catheter is being recalled. The implanted devices are not affected.
Dec 14, 2005Not classifiedZ-0593-06OtherThe tip may separate from the catheter sheath inner core during insertion of the graft. The delivery catheter is being recalled. The implanted devices are not affected.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K120212Afx Introducer SystemEndologix, Inc. · 2012 · 3 generations on · ◆ in the 2012 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • Of its descendants, K110090 (Endologix, Inc., 2011) has gone furthest, with 7 descendants of its own.

How does its category compare?

All MIH devices

MIH System, Endovascular Graft, Aortic Aneurysm Treatment: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MIH

50in 2025

2007: 332008: 452009: 492010: 462011: 642012: 572013: 482014: 802015: 532016: 472017: 412018: 392019: 342020: 442021: 502022: 492023: 502024: 622025: 502026 (partial year): 35
2007high 80 (2014) · low 33 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MIH

Too few MIH 510(k)s in any year since 2007 to give a median.

Recalls per year · MIH

1in 2025

2007: 12011: 292012: 32014: 1, 1 Class I2015: 12016: 62017: 32018: 4, 4 Class I2020: 42021: 42022: 22025: 1, 1 Class I2026 (partial year): 2, 1 Class I
2007high 29 (2011) · low 0 (2008)2026

Class I (7 since 2007) Class II/III not classified

Adverse event reports per year · MIH

3,302in 2025

2007: 1,3232008: 1,7452009: 2,6002010: 2,4932011: 3,2412012: 3,7872013: 3,8832014: 4,2492015: 4,4712016: 4,2962017: 4,1052018: 4,1782019: 3,5922020: 3,0572021: 3,6862022: 3,4552023: 3,5622024: 3,3202025: 3,302
2007high 4,471 (2015) · low 1,323 (2007)2025

MAUDE reports naming product code MIH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200733—11,323
200845—01,745
200949—02,600
201046—02,493
201164—293,241
201257—33,787
201348—03,883
201480—1 (1)4,249
201553—14,471
201647—64,296
201741—34,105
201839—4 (4)4,178
201934—03,592
202044—43,057
202150—43,686
202249—23,455
202350—03,562
202462—03,320
202550—1 (1)3,302
202635—2 (1)—
Cite this page

510(k) Report. “P040002: Endologix Powerlink System.” Lineage record from FDA PMA data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/P040002

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026