510(k) Report

PMA · Approved

Recalled · 2009

Therapy Dual 8 Cardiac Ablation System,Theram 8MM Thermister Ablation Catheter Safire TX Ablation Catheter

P040042PMAApproved Nov 18, 2005410-day reviewClass IIIRecalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Irvine Biomedical,Inc.(Ibi) · OAD Catheter, Percutaneous, Cardiac Ablation, For Treatment Of Atrial Flutter

Approved on Nov 18, 2005 through premarket approval, which rests on the device’s own clinical evidence after a 410-day review. It has been recalled 4 times.

Decided
Nov 18, 2005
Review
410 days
Class
Class III
Product code
OAD
Recalls
4

PMA approvals rest on their own clinical evidence and do not claim a predicate, so this family tree starts here.

P040042 has 57 supplements since 2005: changes to the approved device, its labelling or its manufacturing, each approved separately.

PMA supplements approved per year

2005: 12006: 42007: 62008: 22009: 32010: 22011: 42012: 12013: 12014: 22015: 42016: 52017: 12018: 12019: 32020: 42021: 42022: 32023: 12024: 12025: 12026: 3
2005high 6 (2007) · low 1 (2005)2026
View FIG. 1 as a table
YearSupplements approved
20051
20064
20076
20082
20093
20102
20114
20121
20131
20142
20154
20165
20171
20181
20193
20204
20214
20223
20231
20241
20251
20263
FIG. 1 — Supplements to P040042, 2005–2026Source: FDA PMA database

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA PMA database (decision fields and supplements), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in OAD? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

Does it have a predicate?

PMA decisions do not claim a predicate device.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 25, 2009Not classifiedZ-0594-2010Process designThis action is being taken because the product has the potential for the polyimide tubing to detach from the inner core due to the bond failure. If this occurs, treatment may be delayed or interrupted. The firm's initial health hazard…
Nov 25, 2009Not classifiedZ-0593-2010Process designThis action is being taken because the product has the potential for the polyimide tubing to detach from the inner core due to the bond failure. If this occurs, treatment may be delayed or interrupted. The firm's initial health hazard…
Nov 8, 2006Not classifiedZ-0310-2007OtherRecall was initiated after a clinical incident involving separation of the front sheath, preventing deployment of the stent graft. This required the physician to convert the patient to conventional open repair.
Nov 8, 2006Not classifiedZ-0309-2007OtherRecall was initiated after a clinical incident involving separation of the front sheath, preventing deployment of the stent graft. This required the physician to convert the patient to conventional open repair.

No ancestor or descendant has a Class I recall.

How does its category compare?

All OAD devices

OAD Catheter, Percutaneous, Cardiac Ablation, For Treatment Of Atrial Flutter: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · OAD

19in 2025

2007: 142008: 82009: 92010: 122011: 212012: 192013: 302014: 252015: 242016: 472017: 642018: 292019: 392020: 422021: 502022: 202023: 192024: 192025: 192026 (partial year): 14
2007high 64 (2017) · low 8 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OAD

Too few OAD 510(k)s in any year since 2007 to give a median.

Recalls per year · OAD

1in 2022

2009: 22012: 12013: 22015: 42018: 42022: 12026 (partial year): 1
2007high 4 (2015) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OAD

196in 2025

2007: 22008: 432009: 942010: 4072011: 4862012: 4142013: 2562014: 3292015: 2742016: 2392017: 4032018: 4232019: 4752020: 3892021: 4022022: 3442023: 3192024: 2902025: 196
2007high 486 (2011) · low 2 (2007)2025

MAUDE reports naming product code OAD, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200714—02
20088—043
20099—294
201012—0407
201121—0486
201219—1414
201330—2256
201425—0329
201524—4274
201647—0239
201764—0403
201829—4423
201939—0475
202042—0389
202150—0402
202220—1344
202319—0319
202419—0290
202519—0196
202614—1—
Cite this page

510(k) Report. “P040042: Therapy Dual 8 Cardiac Ablation System,Theram 8MM Thermister Ablation Catheter Safire TX Ablation Catheter.” Lineage record from FDA PMA data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/P040042

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Page last updated Nov 3, 2026