# P050051: Abbott Architect Ausab

- Pathway: PMA
- Applicant: Abbott Laboratories, Inc.
- Decided: Jun 1, 2006 after 174 days
- Product code: SEI (Qualitative And Quantitative Hepatitis B Virus Antibody Assays)
- Device class: 2
- Regulation: 21 CFR 866.3173
- Recalls of this device: 5 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
No predicates in its family tree.
No later device cites it as a predicate yet.
0 of 1 device in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- None. PMA decisions do not claim a predicate device.

## Recalls of this device
- Apr 24, 2025 · Class II · Z-1842-2025: transit delays of certain cold chain products due to severe thunderstorms which may caused the products to be stored outside the recommended labeled storage conditions. This failure could increase the
- May 14, 2007 · Not classified · Z-1097-2007: During an internal study, the Architect AUSAB assay was found to read preparations of the WHO standard up to 53% higher than the WHO assigned value. This has the potential to cause elevated results an
- May 14, 2007 · Not classified · Z-1096-2007: During an internal study, the Architect AUSAB assay was found to read preparations of the WHO standard up to 53% higher than the WHO assigned value. This has the potential to cause elevated results an
- May 14, 2007 · Not classified · Z-1098-2007: During an internal study, the Architect AUSAB assay was found to read preparations of the WHO standard up to 53% higher than the WHO assigned value. This has the potential to cause elevated results an
- Jan 2, 2007 · Not classified · Z-0413-2007: The ARCHITECT AUSAB and ARCHITECT Anti-HCV assay parameter default interpretation screens when using ARCHITECT Assay CD-ROM US Version 06E58-21 does not align with the result interpretation options in