# P060030: Cobas Ampliprep/Cobas Taqman HCV Test

- Pathway: PMA
- Applicant: Roche Molecular Systems, Inc.
- Decided: Oct 30, 2008 after 734 days
- Product code: MZP (Assay, Hybridization And/Or Nucleic Acid Amplification For Detection Of Hepatitis C Rna,Hepatitis C Virus)
- Device class: 2
- Regulation: 21 CFR 866.3170
- Recalls of this device: 2 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
No predicates in its family tree.
No later device cites it as a predicate yet.
0 of 1 device in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- None. PMA decisions do not claim a predicate device.

## Recalls of this device
- Apr 14, 2011 · Not classified · Z-2587-2011: The COBAS Ampliprep/COBAS TaqMan HCV Test has been shown to under-quantitate a subset of genotype 4 patient specimens by approximately 1.0-1.5 log 10 in the absence of any sequence mismatches.
- Apr 14, 2011 · Not classified · Z-2586-2011: The COBAS Ampliprep/COBAS TaqMan HCV Test has been shown to under-quantitate a subset of genotype 4 patient specimens by approximately 1.0-1.5 log 10 in the absence of any sequence mismatches.

Source page: https://510kreport.com/device/P060030
Methodology: https://510kreport.com/methodology