Thoratec Heartmate II Left Ventricular Assist System

P060040PMAApproved Apr 21, 2008486-day reviewClass IIIRecalls 18 (12 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Abbott Medical · DSQ Ventricular (Assist) Bypass

Approved on Apr 21, 2008 through premarket approval, which rests on the device’s own clinical evidence after a 486-day review. It has a Class I recall. It has been recalled 18 times, including 12 Class I recalls.

Decided
Apr 21, 2008
Review
486 days
Class
Class III
Product code
DSQ
Recalls
18 (12 Class I)

PMA approvals rest on their own clinical evidence and do not claim a predicate, so this family tree starts here.

P060040 has 125 supplements since 2008: changes to the approved device, its labelling or its manufacturing, each approved separately.

PMA supplements approved per year

2008: 12009: 32010: 92011: 12012: 82013: 62014: 82015: 92016: 162017: 72018: 22019: 32020: 42021: 32022: 62023: 102024: 132025: 102026: 6
2008high 16 (2016) · low 1 (2008)2026
View FIG. 1 as a table
YearSupplements approved
20081
20093
20109
20111
20128
20136
20148
20159
201616
20177
20182
20193
20204
20213
20226
202310
202413
202510
20266
FIG. 1 — Supplements to P060040, 2008–2026Source: FDA PMA database

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA PMA database (decision fields and supplements), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

Does it have a predicate?

PMA decisions do not claim a predicate device.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 9, 2025Class IIZ-0507-2026Device DesignDue to potential corrosion issues associated with the Backup Battery cable connector that may cause a Backup Battery Fault Alarm.
Oct 9, 2025Class IIZ-0508-2026Device DesignDue to potential corrosion issues associated with the Backup Battery cable connector that may cause a Backup Battery Fault Alarm.
Jun 23, 2025Class IZ-2290-2025Vendor change controlDue to misaligned V-Lock feature on the Mobile Power Unit power cords, which prevents the locking mechanism from properly engaging with the Mobile Power Unit and may cause unintended disconnect.
Jun 23, 2025Class IZ-2291-2025Vendor change controlDue to misaligned V-Lock feature on the Mobile Power Unit power cords, which prevents the locking mechanism from properly engaging with the Mobile Power Unit and may cause unintended disconnect.
Mar 13, 2025Class IZ-1570-2025Process controlLeft Ventricular Assist System Mobile Power Unit (MPU) may have electrical component that may cause MPU not to turn on, unprompted shut down, or turning off/restarting, may cause system controller to indicate Yellow Wrench or No External…
May 8, 2024Class IZ-1921-2024Device DesignSystem monitor screen may display atypical behavior: Overlapping screens/buttons, Frozen screen, Distorted text, blanks, or zeroes in place of values, Unresponsive buttons where the user is unable to initiate a command, which may lead to…
Feb 19, 2024Class IZ-1300-2024Material/Component ContaminationObserved outflow graft deformation known as Extrinsic Outflow Graft Obstruction (EOGO) associated with Left Ventricular Assist System.
Mar 30, 2017Class IZ-2053-2017Device DesignSt. Jude Medical is providing all users of their HeartMate II LVAS with Pocket Controller with new software and updates to hardware to make the exchange to a backup controller easier. This is related to recall Z-1227/1230-2014.
Mar 30, 2017Class IZ-2052-2017Device DesignSt. Jude Medical is providing all users of their HeartMate II LVAS with Pocket Controller with new software and updates to hardware to make the exchange to a backup controller easier. This is related to recall Z-1227/1230-2014.
Mar 30, 2017Class IZ-2051-2017Device DesignSt. Jude Medical is providing all users of their HeartMate II LVAS with Pocket Controller with new software and updates to hardware to make the exchange to a backup controller easier. This is related to recall Z-1227/1230-2014.
Mar 4, 2014Class IZ-1230-2014Device DesignDiscovery of serious injuries and deaths associated with the process of changing from a primary System controller to their back-up System controller in patients using the Pocket System controller model.
Mar 4, 2014Class IZ-1228-2014Device DesignDiscovery of serious injuries and deaths associated with the process of changing from a primary System controller to their back-up System controller in patients using the Pocket System controller model.
Mar 4, 2014Class IZ-1227-2014Device DesignDiscovery of serious injuries and deaths associated with the process of changing from a primary System controller to their back-up System controller in patients using the Pocket System controller model.
Mar 4, 2014Class IZ-229-2014Device DesignDiscovery of serious injuries and deaths associated with the process of changing from a primary System controller to their back-up System controller in patients using the Pocket System controller model.
Feb 23, 2012Not classifiedZ-1228-2012Device DesignA trend of disconnected bend reliefs on the sealed outflow graft used with the HeartMate II LVAS has been identified
Sep 1, 2010Not classifiedZ-0058-2011Device DesignConnection failure-- product may malfunction because of bent pins or broken wires in power cord connection
Oct 14, 2009Not classifiedZ-0422-2010OtherPatient reports alerted Thoratec that during the process of disconnecting power leads from the battery clips, the threaded connector of the battery clip housing was observed to be loose, and in some cases dislodged from the battery clip…
Oct 24, 2008Not classifiedZ-0496-2009Device DesignThe firm has become aware that, over time, wear and fatigue of the percutaneous lead connecting the HeartMate II Left Ventricular Assist System (HM II LVAS) blood pump with external system controller may result in damage that has the…

No ancestor or descendant has a Class I recall.

How does its category compare?

All DSQ devices

DSQ Ventricular (Assist) Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.

FDA decisions per year · DSQ

50in 2025

2007: 72008: 52009: 92010: 122011: 42012: 202013: 362014: 302015: 242016: 372017: 362018: 352019: 342020: 482021: 382022: 332023: 482024: 612025: 502026 (partial year): 46
2007high 61 (2024) · low 4 (2011)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DSQ

Too few DSQ 510(k)s in any year since 2007 to give a median.

Recalls per year · DSQ

4in 2025

2007: 42008: 12009: 42010: 12012: 12016: 2, 2 Class I2018: 32019: 12020: 1, 1 Class I2021: 12022: 112023: 92024: 1, 1 Class I2025: 42026 (partial year): 4
2007high 11 (2022) · low 0 (2011)2026

Class I (4 since 2007) Class II/III not classified

Adverse event reports per year · DSQ

10,419in 2025

2007: 2242008: 2642009: 5292010: 6872011: 1,0822012: 2,1202013: 2,9412014: 3,9032015: 6,7712016: 7,1902017: 9,0552018: 12,6602019: 20,2162020: 14,2392021: 15,7662022: 15,9942023: 13,0552024: 13,0462025: 10,419
2007high 20,216 (2019) · low 224 (2007)2025

MAUDE reports naming product code DSQ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20077—4224
20085—1264
20099—4529
201012—1687
20114—01,082
201220—12,120
201336—02,941
201430—03,903
201524—06,771
201637—2 (2)7,190
201736—09,055
201835—312,660
201934—120,216
202048—1 (1)14,239
202138—115,766
202233—1115,994
202348—913,055
202461—1 (1)13,046
202550—410,419
202646—4—
Cite this page

510(k) Report. “P060040: Thoratec Heartmate II Left Ventricular Assist System.” Lineage record from FDA PMA data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/P060040

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Page last updated Nov 3, 2026