510(k) Report

PMA · Approved

Recalled · Class II · 2021

Prometra Programmable Infusion Pump System

P080012PMAApproved Feb 7, 20121,397-day reviewClass IIIRecalls 13

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Infusyn Therapeutics, LLC · LKK Pump, Infusion, Implanted, Programmable

Approved on Feb 7, 2012 through premarket approval, which rests on the device’s own clinical evidence after a 1,397-day review. It has been recalled 13 times.

Decided
Feb 7, 2012
Review
1,397 days
Class
Class III
Product code
LKK
Recalls
13

PMA approvals rest on their own clinical evidence and do not claim a predicate, so this family tree starts here.

P080012 has 63 supplements since 2012: changes to the approved device, its labelling or its manufacturing, each approved separately.

PMA supplements approved per year

2012: 62013: 92014: 32015: 82016: 42017: 92018: 22019: 92020: 52021: 22022: 32024: 22025: 1
2012high 9 (2013) · low 0 (2023)2025
View FIG. 1 as a table
YearSupplements approved
20126
20139
20143
20158
20164
20179
20182
20199
20205
20212
20223
20242
20251
FIG. 1 — Supplements to P080012, 2012–2025Source: FDA PMA database

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA PMA database (decision fields and supplements), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LKK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

Does it have a predicate?

PMA decisions do not claim a predicate device.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 26, 2021Class IIZ-1996-2021Error in labelingPhysician Order Form (PL-15400-00 Nov 2018) distributed is discontinued due to the text 'Refill the pump and program refill with the same medication removed prior to MRI' may be misinterpreted as 'reuse the same drug solution extracted…
Dec 24, 2019Class IIZ-1125-2020Software change controlA pump alarm function anomaly in the pump firmware code may result in the pump not sending the expected error code notifications to the Programmer, or not emit an audible alarm tone in certain "out of tolerance" conditions associated with…
Dec 24, 2019Class IIZ-1123-2020Software change controlA pump alarm function anomaly in the pump firmware code may result in the pump not sending the expected error code notifications to the Programmer, or not emit an audible alarm tone in certain "out of tolerance" conditions associated with…
Dec 24, 2019Class IIZ-1121-2020Software change controlA pump alarm function anomaly in the pump firmware code may result in the pump not sending the expected error code notifications to the Programmer, or not emit an audible alarm tone in certain "out of tolerance" conditions associated with…
Dec 24, 2019Class IIZ-1120-2020Software change controlA pump alarm function anomaly in the pump firmware code may result in the pump not sending the expected error code notifications to the Programmer, or not emit an audible alarm tone in certain "out of tolerance" conditions associated with…
Dec 24, 2019Class IIZ-1124-2020Software change controlA pump alarm function anomaly in the pump firmware code may result in the pump not sending the expected error code notifications to the Programmer, or not emit an audible alarm tone in certain "out of tolerance" conditions associated with…
Dec 24, 2019Class IIZ-1122-2020Software change controlA pump alarm function anomaly in the pump firmware code may result in the pump not sending the expected error code notifications to the Programmer, or not emit an audible alarm tone in certain "out of tolerance" conditions associated with…
Mar 21, 2019Class IIZ-1713-2019Software change controlThe pump shuts down unexpectedly if Error 115, an alarm associated with a watchdog timer timeout, is triggered. There has been an increase of Error 115 occurrences due to a software defect caused by a Healthcare Provider using a Clinician…
Mar 21, 2019Class IIZ-1712-2019Software change controlThe pump shuts down unexpectedly if Error 115, an alarm associated with a watchdog timer timeout, is triggered. There has been an increase of Error 115 occurrences due to a software defect caused by a Healthcare Provider using a Clinician…
Mar 21, 2019Class IIZ-1714-2019Software change controlThe pump shuts down unexpectedly if Error 115, an alarm associated with a watchdog timer timeout, is triggered. There has been an increase of Error 115 occurrences due to a software defect caused by a Healthcare Provider using a Clinician…
Mar 21, 2019Class IIZ-1716-2019Software change controlThe pump shuts down unexpectedly if Error 115, an alarm associated with a watchdog timer timeout, is triggered. There has been an increase of Error 115 occurrences due to a software defect caused by a Healthcare Provider using a Clinician…
Mar 21, 2019Class IIZ-1715-2019Software change controlThe pump shuts down unexpectedly if Error 115, an alarm associated with a watchdog timer timeout, is triggered. There has been an increase of Error 115 occurrences due to a software defect caused by a Healthcare Provider using a Clinician…
May 22, 2017Class IIZ-2740-2017Under Investigation by firmFlowonix Medical received a report of a patient implanted with the Prometra II Programmable Pump who may have received a fatal drug overdose during an MRI procedure. The Prometra II pump has an FDA-approved design feature intended to…

No ancestor or descendant has a Class I recall.

How does its category compare?

All LKK devices

LKK Pump, Infusion, Implanted, Programmable: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · LKK

22in 2025

2007: 82008: 82009: 102010: 292011: 342012: 332013: 332014: 282015: 402016: 302017: 422018: 552019: 572020: 492021: 322022: 322023: 192024: 222025: 222026 (partial year): 9
2007high 57 (2019) · low 8 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LKK

Too few LKK 510(k)s in any year since 2007 to give a median.

Recalls per year · LKK

5in 2024

2007: 162008: 352009: 62010: 22011: 42012: 3, 2 Class I2016: 1, 1 Class I2017: 22019: 1, 1 Class I2020: 12023: 22024: 52026 (partial year): 2
2007high 35 (2008) · low 0 (2013)2026

Class I (4 since 2007) Class II/III not classified

Adverse event reports per year · LKK

5,732in 2025

2007: 2,6762008: 3,3982009: 3,2252010: 4,0672011: 3,9132012: 5,5572013: 8,0552014: 7,4582015: 8,4512016: 8,4612017: 7,7512018: 6,9362019: 5,6522020: 5,1152021: 4,7132022: 4,3102023: 4,0412024: 4,6182025: 5,732
2007high 8,461 (2016) · low 2,676 (2007)2025

MAUDE reports naming product code LKK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20078—162,676
20088—353,398
200910—63,225
201029—24,067
201134—43,913
201233—3 (2)5,557
201333—08,055
201428—07,458
201540—08,451
201630—1 (1)8,461
201742—27,751
201855—06,936
201957—1 (1)5,652
202049—15,115
202132—04,713
202232—04,310
202319—24,041
202422—54,618
202522—05,732
20269—2—
Cite this page

510(k) Report. “P080012: Prometra Programmable Infusion Pump System.” Lineage record from FDA PMA data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/P080012

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.