510(k) Report

PMA · Approved

Recalled · Class I · 2025

Pipeline Embolization Device

P100018PMAApproved Apr 6, 2011323-day reviewClass IIIRecalls 12 (10 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Micro Therapeutics, Inc. d/b/a ev3 Neurovascular · OUT Intracranial Aneurysm Flow Diverter

Approved on Apr 6, 2011 through premarket approval, which rests on the device’s own clinical evidence after a 323-day review. It has a Class I recall. It has been recalled 12 times, including 10 Class I recalls.

Decided
Apr 6, 2011
Review
323 days
Class
Class III
Product code
OUT
Recalls
12 (10 Class I)

PMA approvals rest on their own clinical evidence and do not claim a predicate, so this family tree starts here.

P100018 has 51 supplements since 2011: changes to the approved device, its labelling or its manufacturing, each approved separately.

PMA supplements approved per year

2011: 22012: 12013: 42014: 32015: 12016: 12017: 12018: 42019: 42020: 42021: 62022: 22023: 62024: 42025: 42026: 4
2011high 6 (2021) · low 1 (2012)2026
View FIG. 1 as a table
YearSupplements approved
20112
20121
20134
20143
20151
20161
20171
20184
20194
20204
20216
20222
20236
20244
20254
20264
FIG. 1 — Supplements to P100018, 2011–2026Source: FDA PMA database

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA PMA database (decision fields and supplements), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

Does it have a predicate?

PMA decisions do not claim a predicate device.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 30, 2025Class IZ-1245-2025Device DesignUse of embolization device with diameters greater-than-or-equal-to 4mm may lead to incomplete wall apposition and/or braid deformation (fish-mouthing/braid narrowing/braid collapse) noted during procedure and post-procedure, typically…
Jan 30, 2025Class IZ-1246-2025Device DesignUse of embolization device with diameters greater-than-or-equal-to 4mm may lead to incomplete wall apposition and/or braid deformation (fish-mouthing/braid narrowing/braid collapse) noted during procedure and post-procedure, typically…
Jul 13, 2021Class IZ-2402-2021Device DesignDue to potential push wire fractures in the delivery system during use. The issue presents as a fracture in the spiral cut area of the Hypotube.
Jul 13, 2021Class IZ-2403-2021Device DesignDue to potential push wire fractures in the delivery system during use. The issue presents as a fracture in the spiral cut area of the Hypotube.
Feb 14, 2020Class IZ-1527-2020Process change controlEmbolization device delivery system may fracture at the distal section during device implantation. A weakened bond may result in unintended separation, where the distal portion of the device delivery system remains in the patient. If a…
Oct 5, 2016Class IZ-0307-2017Device DesignDelamination and detachment of the PTFE (polytetrafluoroethylene) coating material may lead to PTFE coating in the blood stream.
Oct 5, 2016Class IZ-0308-2017Device DesignDelamination and detachment of the PTFE (polytetrafluoroethylene) coating material may lead to PTFE coating in the blood stream.
Oct 5, 2016Class IZ-0309-2017Device DesignDelamination and detachment of the PTFE (polytetrafluoroethylene) coating material may lead to PTFE coating in the blood stream.
Oct 5, 2016Class IZ-0310-2017Device DesignDelamination and detachment of the PTFE (polytetrafluoroethylene) coating material may lead to PTFE coating in the blood stream.
Sep 4, 2015Class IIZ-0120-2016Process controlThe firm is recalling Pipeline and Pipeline Flex Embolization Devices from U.S. since the devices were shipped with an EU version of the Instructions for Use.
Apr 1, 2014Class IZ-1453-2014Process change controlPotential for the PTFE (polytetrafluoroethylene) coating to delaminate and detach from the delivery wire.
Jul 19, 2013Class IIZ-2045-2013Employee errorCovidien Neurovascular has initiated the recall of several lots of Pipeline Embolization Device (PED) because the Instructions For Use (IFU) do not include the following contraindication: Patients in whom a pre-existing stent is in place…

No ancestor or descendant has a Class I recall.

How does its category compare?

All OUT devices

OUT Intracranial Aneurysm Flow Diverter: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · OUT

10in 2025

2011: 32012: 12013: 42014: 32015: 12016: 12017: 12018: 72019: 52020: 92021: 132022: 72023: 82024: 142025: 102026 (partial year): 6
2007high 14 (2024) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OUT

Too few OUT 510(k)s in any year since 2007 to give a median.

Recalls per year · OUT

3so far in 2026, the most since 2007

2015: 12016: 1, 1 Class I2025: 12026 (partial year): 3
2007high 3 (2026) · low 0 (2007)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · OUT

1,625in 2025

2011: 42012: 3352013: 6182014: 4382015: 5622016: 5402017: 4482018: 4252019: 5282020: 8092021: 8502022: 1,1182023: 1,1862024: 1,3262025: 1,625
2007high 1,625 (2025) · low 0 (2007)2025

MAUDE reports naming product code OUT, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20113—04
20121—0335
20134—0618
20143—0438
20151—1562
20161—1 (1)540
20171—0448
20187—0425
20195—0528
20209—0809
202113—0850
20227—01,118
20238—01,186
202414—01,326
202510—11,625
20266—3—
Cite this page

510(k) Report. “P100018: Pipeline Embolization Device.” Lineage record from FDA PMA data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/P100018

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Page last updated Nov 3, 2026