510(k) Report

PMA · Approved

Recalled · Class II · 2025

Cardiomems HF Pressure Measurement System

P100045PMAApproved May 28, 20141,260-day reviewClass IIIRecalls 13

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Abbott Medical · MOM System, Hemodynamic, Implantable

Approved on May 28, 2014 through premarket approval, which rests on the device’s own clinical evidence after a 1,260-day review. It has been recalled 13 times.

Decided
May 28, 2014
Review
1,260 days
Class
Class III
Product code
MOM
Recalls
13

PMA approvals rest on their own clinical evidence and do not claim a predicate, so this family tree starts here.

P100045 has 90 supplements since 2014: changes to the approved device, its labelling or its manufacturing, each approved separately.

PMA supplements approved per year

2014: 12015: 32016: 82017: 82018: 92019: 82020: 62021: 112022: 62023: 72024: 82025: 72026: 8
2014high 11 (2021) · low 1 (2014)2026
View FIG. 1 as a table
YearSupplements approved
20141
20153
20168
20178
20189
20198
20206
202111
20226
20237
20248
20257
20268
FIG. 1 — Supplements to P100045, 2014–2026Source: FDA PMA database

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA PMA database (decision fields and supplements), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MOM? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

Does it have a predicate?

PMA decisions do not claim a predicate device.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 3, 2025Class IIZ-1427-2025Software designDue to issues associated with data migration of patient information and results to a Cloud service, the firm has received complaint associated with some patient profile information being duplicated while other patient information is…
Nov 6, 2024Class IIZ-0987-2025Under Investigation by firmDue to a data migration from on-premises server to the cloud environment, some customers reported that they were not able to transmit readings.
Oct 5, 2024Class IIZ-0988-2025Under Investigation by firmAs a result of a cloud data migration of Patient and Hospital Electronic Systems (PES and HES) reverted to an outdated configuration causing incorrect/inaccurate pulmonary artery (PA) pressure readings.
Oct 5, 2024Class IIZ-0989-2025Under Investigation by firmAs a result of a cloud data migration of Patient and Hospital Electronic Systems (PES and HES) reverted to an outdated configuration causing incorrect/inaccurate pulmonary artery (PA) pressure readings.
Oct 4, 2023Class IIZ-0218-2024Device DesignPotential for damaged and frayed power connector plug with repeated bending or manipulation beyond 90 degrees.
Feb 7, 2023Class IIZ-1225-2023Device DesignSelect CardioMEMS PA Sensors (Model CM2000) operate outside of the intended frequency range at higher elevations when readings are taken over approximately 2,000 ft. (610 meters) above sea level; operation outside of the intended…
Feb 7, 2023Class IIZ-1223-2023Device DesignSelect CardioMEMS PA Sensors (Model CM2000) operate outside of the intended frequency range at higher elevations when readings are taken over approximately 2,000 ft. (610 meters) above sea level; operation outside of the intended…
Feb 7, 2023Class IIZ-1224-2023Device DesignSelect CardioMEMS PA Sensors (Model CM2000) operate outside of the intended frequency range at higher elevations when readings are taken over approximately 2,000 ft. (610 meters) above sea level; operation outside of the intended…
Feb 7, 2023Class IIZ-1218-2023Device DesignRadiated Radiofrequency (RF) emissions of the CardioMEMS Patient Electronics Systems (PES) Models CM1000, CM1010 and CM1100 and CardioMEMS Hospital Electronics Systems (HES) Model CM3000 exceed certain emissions limits and have the…
Feb 7, 2023Class IIZ-1221-2023Device DesignRadiated Radiofrequency (RF) emissions of the CardioMEMS Patient Electronics Systems (PES) Models CM1000, CM1010 and CM1100 and CardioMEMS Hospital Electronics Systems (HES) Model CM3000 exceed certain emissions limits and have the…
Feb 7, 2023Class IIZ-1219-2023Device DesignRadiated Radiofrequency (RF) emissions of the CardioMEMS Patient Electronics Systems (PES) Models CM1000, CM1010 and CM1100 and CardioMEMS Hospital Electronics Systems (HES) Model CM3000 exceed certain emissions limits and have the…
Feb 7, 2023Class IIZ-1220-2023Device DesignRadiated Radiofrequency (RF) emissions of the CardioMEMS Patient Electronics Systems (PES) Models CM1000, CM1010 and CM1100 and CardioMEMS Hospital Electronics Systems (HES) Model CM3000 exceed certain emissions limits and have the…
Jun 14, 2018Class IIZ-2522-2018Process controlAbbott is advising customers that a small number of CardioMEMS(R) Hospital Electronics Systems (Model CM3000) and Patient Electronics Systems (Model CM1100) may deliver a system error, known as Error 5. While this error message is…

No ancestor or descendant has a Class I recall.

How does its category compare?

All MOM devices

MOM System, Hemodynamic, Implantable: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · MOM

24in 2025

2014: 22015: 32016: 82017: 82018: 92019: 82020: 62021: 112022: 62023: 72024: 162025: 242026 (partial year): 18
2007high 24 (2025) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MOM

Too few MOM 510(k)s in any year since 2007 to give a median.

Recalls per year · MOM

3in 2024

2023: 12024: 3
2007high 3 (2024) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MOM

2,189in 2025

2015: 342016: 612017: 2912018: 1,2122019: 6722020: 6952021: 7132022: 7242023: 1,8862024: 1,9162025: 2,189
2007high 2,189 (2025) · low 0 (2007)2025

MAUDE reports naming product code MOM, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20110—00
20120—00
20130—00
20142—00
20153—034
20168—061
20178—0291
20189—01,212
20198—0672
20206—0695
202111—0713
20226—0724
20237—11,886
202416—31,916
202524—02,189
202618—0—
Cite this page

510(k) Report. “P100045: Cardiomems HF Pressure Measurement System.” Lineage record from FDA PMA data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/P100045

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Page last updated Nov 3, 2026