510(k) Report

PMA · Approved

Recalled · Class I · 2023

Heartware Ventricular Assist System

P100047PMAApproved Nov 20, 2012693-day reviewClass IIIRecalls 60 (40 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medtronic · DSQ Ventricular (Assist) Bypass

Approved on Nov 20, 2012 through premarket approval, which rests on the device’s own clinical evidence after a 693-day review. It has a Class I recall. It has been recalled 60 times, including 40 Class I recalls.

Decided
Nov 20, 2012
Review
693 days
Class
Class III
Product code
DSQ
Recalls
60 (40 Class I)

PMA approvals rest on their own clinical evidence and do not claim a predicate, so this family tree starts here.

P100047 has 227 supplements since 2012: changes to the approved device, its labelling or its manufacturing, each approved separately.

PMA supplements approved per year

2012: 62013: 232014: 212015: 142016: 192017: 242018: 182019: 152020: 242021: 182022: 122023: 102024: 72025: 82026: 8
2012high 24 (2017) · low 6 (2012)2026
View FIG. 1 as a table
YearSupplements approved
20126
201323
201421
201514
201619
201724
201818
201915
202024
202118
202212
202310
20247
20258
20268
FIG. 1 — Supplements to P100047, 2012–2026Source: FDA PMA database

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA PMA database (decision fields and supplements), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

Does it have a predicate?

PMA decisions do not claim a predicate device.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 31, 2023Class IIZ-0475-2024Labeling designUpdates to the HVAD system instructions for use (IFU) and patient manual (PM) will be provided. These updates will clarify (1) the conditions under which an HVAD [Controller Fault] alarm may sound and the recommended troubleshooting…
May 16, 2023Class IIZ-1993-2023Software change controlMedtronic has identified that the logfiles downloaded from the recently updated Model 1521 monitors (Serial Numbers: MON4xxxxxx for OUS and MON5xxxxxx for US), are unable to be processed by the Autologs web portal.
Jun 28, 2022Class IZ-1512-2022Device DesignBattery performance issues. Battery electrical faults render it unable to power the controller, unable to accept charge from the battery charger, and/or result in the battery to appear to remain charged while discharging.
May 5, 2022Class IZ-1266-2022Process controlBatteries for the HVAD system have a weld defect affecting internal components within the HVAD battery, resulting in the battery to malfunction and no longer provide power or prevent the battery from holding a complete charge or properly…
Apr 14, 2022Class IZ-1207-2022Under Investigation by firmThe pump may have a welding defect that can lead the pump to malfunction. The issue is under investigation by the firm. It is suspected that a weld defect on the center post cap may allow moisture into the center post and corrode the…
Mar 30, 2022Class IIZ-1106-2022Labeling designFinnish and Turkish translation errors in HVAD controller and monitor displays as will as Instructions for Use, Patient Manual, and Emergency Responder Guide for the HeartWare Ventricular Assist Device (HVAD) system.
Mar 30, 2022Class IIZ-1103-2022Labeling designFinnish and Turkish translation errors in HVAD controller and monitor displays as will as Instructions for Use, Patient Manual, and Emergency Responder Guide for the HeartWare Ventricular Assist Device (HVAD) system.
Mar 30, 2022Class IIZ-1105-2022Labeling designFinnish and Turkish translation errors in HVAD controller and monitor displays as will as Instructions for Use, Patient Manual, and Emergency Responder Guide for the HeartWare Ventricular Assist Device (HVAD) system.
Mar 30, 2022Class IIZ-1104-2022Labeling designFinnish and Turkish translation errors in HVAD controller and monitor displays as will as Instructions for Use, Patient Manual, and Emergency Responder Guide for the HeartWare Ventricular Assist Device (HVAD) system.
Mar 30, 2022Class IIZ-1110-2022Labeling designErrors and inconsistencies that were identified in the Instruction for Use (IFU), Emergency Responder Guide (ERG), and Patient Manual (PM) for the HeartWare Ventricular Assist Device (HVAD) system.
Mar 30, 2022Class IIZ-1107-2022Labeling designFinnish and Turkish translation errors in HVAD controller and monitor displays as will as Instructions for Use, Patient Manual, and Emergency Responder Guide for the HeartWare Ventricular Assist Device (HVAD) system.
Mar 30, 2022Class IIZ-1102-2022Labeling designFinnish and Turkish translation errors in HVAD controller and monitor displays as will as Instructions for Use, Patient Manual, and Emergency Responder Guide for the HeartWare Ventricular Assist Device (HVAD) system.
Jan 27, 2022Class IIZ-0726-2022Labeling designUpdated cleaning instructions are being provided for the Controller AC Adapter, DC Adapter, and Battery to provide clarity to avoid cleaning the power source connector pins, as this could remove the lubricant that is applied to the pins…
Aug 6, 2021Class IIZ-2478-2021Device DesignThere is more force required to slide the driveline cover (boot cover) away from the connector after completion of a Driveline Strain Relief Repair than over an unrepaired strain relief. This may lead to difficulty in pulling back the…
Jun 3, 2021Class IZ-2197-2021Unknown/Undetermined by firmMedtronic is stopping the distribution and sale of the HVAD System due to a growing body of observational clinical comparisons that shows a higher frequency of neurological adverse events and mortality associated with the system when…
Jun 3, 2021Class IZ-2189-2021Unknown/Undetermined by firmMedtronic is stopping the distribution and sale of the HVAD System due to a growing body of observational clinical comparisons that shows a higher frequency of neurological adverse events and mortality associated with the system when…
Jun 3, 2021Class IZ-2188-2021Unknown/Undetermined by firmMedtronic is stopping the distribution and sale of the HVAD System due to a growing body of observational clinical comparisons that shows a higher frequency of neurological adverse events and mortality associated with the system when…
Jun 3, 2021Class IZ-2192-2021Unknown/Undetermined by firmMedtronic is stopping the distribution and sale of the HVAD System due to a growing body of observational clinical comparisons that shows a higher frequency of neurological adverse events and mortality associated with the system when…
Jun 3, 2021Class IZ-2191-2021Unknown/Undetermined by firmMedtronic is stopping the distribution and sale of the HVAD System due to a growing body of observational clinical comparisons that shows a higher frequency of neurological adverse events and mortality associated with the system when…
Jun 3, 2021Class IZ-2196-2021Unknown/Undetermined by firmMedtronic is stopping the distribution and sale of the HVAD System due to a growing body of observational clinical comparisons that shows a higher frequency of neurological adverse events and mortality associated with the system when…
Jun 3, 2021Class IZ-2194-2021Unknown/Undetermined by firmMedtronic is stopping the distribution and sale of the HVAD System due to a growing body of observational clinical comparisons that shows a higher frequency of neurological adverse events and mortality associated with the system when…
Jun 3, 2021Class IZ-2195-2021Unknown/Undetermined by firmMedtronic is stopping the distribution and sale of the HVAD System due to a growing body of observational clinical comparisons that shows a higher frequency of neurological adverse events and mortality associated with the system when…
Jun 3, 2021Class IZ-2190-2021Unknown/Undetermined by firmMedtronic is stopping the distribution and sale of the HVAD System due to a growing body of observational clinical comparisons that shows a higher frequency of neurological adverse events and mortality associated with the system when…
Jun 3, 2021Class IZ-2193-2021Unknown/Undetermined by firmMedtronic is stopping the distribution and sale of the HVAD System due to a growing body of observational clinical comparisons that shows a higher frequency of neurological adverse events and mortality associated with the system when…
May 28, 2021Class IIZ-2072-2021Nonconforming Material/ComponentThe pump has an impeller with a shroud height that did not meet the lower control limit.
Feb 26, 2021Class IZ-1341-2021Device DesignThere is the potential for the Controller power cables (AC Adapter, DC Adapter, and Battery cables), HVAD Alarm Adapter, and HVAD Monitor data cables to cause damage to the HVAD Controller ports when inserting the cables into the…
Feb 26, 2021Class IZ-1336-2021Device DesignThere is the potential for the Controller power cables (AC Adapter, DC Adapter, and Battery cables), HVAD Alarm Adapter, and HVAD Monitor data cables to cause damage to the HVAD Controller ports when inserting the cables into the…
Feb 26, 2021Class IZ-1339-2021Device DesignThere is the potential for the Controller power cables (AC Adapter, DC Adapter, and Battery cables), HVAD Alarm Adapter, and HVAD Monitor data cables to cause damage to the HVAD Controller ports when inserting the cables into the…
Feb 26, 2021Class IIZ-1438-2021Device DesignIFU and Patient Manual updated to describe HVAD Controller Connection Verification. The labeling is being corrected to reflect the actual verified behavior of a successful connector to the controller, which is a "beep" and not "audible…
Feb 26, 2021Class IZ-1340-2021Device DesignThere is the potential for the Controller power cables (AC Adapter, DC Adapter, and Battery cables), HVAD Alarm Adapter, and HVAD Monitor data cables to cause damage to the HVAD Controller ports when inserting the cables into the…
Feb 26, 2021Class IZ-1338-2021Device DesignThere is the potential for the Controller power cables (AC Adapter, DC Adapter, and Battery cables), HVAD Alarm Adapter, and HVAD Monitor data cables to cause damage to the HVAD Controller ports when inserting the cables into the…
Feb 26, 2021Class IIZ-1469-2021Device DesignIFU and Patient Manual updated to describe Expected Alarm Duration. The Instructions for Use (IFU) is being updated to advise clinicians that alarms that resolve after a very short period may mean there is an intermittent problem and the…
Feb 26, 2021Class IZ-1426-2021Device DesignMedtronic HeartWare HVAD System Instructions for Use (IFU) and Patient Manual (PM) updates regarding: Carrying Cases, Driveline Cover Orientation, and Controller Power Up Sequence issues. The PM is being updated to clarify use of the…
Feb 26, 2021Class IZ-1337-2021Device DesignThere is the potential for the Controller power cables (AC Adapter, DC Adapter, and Battery cables), HVAD Alarm Adapter, and HVAD Monitor data cables to cause damage to the HVAD Controller ports when inserting the cables into the…
Nov 19, 2020Class IZ-0946-2021Under Investigation by firmPumps manufactured with impellers from a subset of lots from a single supplier are failing to initially start, restart, or experiencing a delay in restart within the maximum number of restart attempts at a rate substantially higher than…
Apr 3, 2020Class IZ-2057-2020Device DesignThe Outflow Graft may be subject to tears and the Strain Relief screw may break during the pre-implant pump assembly and attachment to the HVAD Pump.
Apr 3, 2020Class IZ-2059-2020Device DesignThe Outflow Graft may be subject to tears and the Strain Relief screw may break during the pre-implant pump assembly and attachment to the HVAD Pump.
Apr 3, 2020Class IZ-2058-2020Device DesignThe Outflow Graft may be subject to tears and the Strain Relief screw may break during the pre-implant pump assembly and attachment to the HVAD Pump.
Apr 3, 2020Class IZ-2056-2020Device DesignThe Outflow Graft may be subject to tears and the Strain Relief screw may break during the pre-implant pump assembly and attachment to the HVAD Pump.
Apr 3, 2020Class IZ-2055-2020Device DesignThe Outflow Graft may be subject to tears and the Strain Relief screw may break during the pre-implant pump assembly and attachment to the HVAD Pump.
Jan 23, 2020Class IZ-1428-2020Device DesignPotential for an HVAD System user to mistakenly insert the Battery Charger AC Adapter into an HVAD Controller power port.
Jan 23, 2020Class IZ-1427-2020Device DesignPotential for an HVAD System user to mistakenly insert the Battery Charger AC Adapter into an HVAD Controller power port.
Nov 19, 2018Class IIZ-0582-2019OtherHVAD Battery Charger units manufactured with wrong inductors.
Aug 28, 2018Class IIZ-0067-2019Device DesignCertain HVAD System Controller units may not meet the labeled standard for protection against water or fluid ingress, specifically they may have the potential to develop hairline cracks at the power ports of the controller housing.
May 2, 2018Class IZ-1903-2018Device DesignPossible transient electrical connection interruption between an HVAD System power source (Battery, AC Adapter, or DC Adapter) and the Controller, while the power source is connected, that may cause unintended switching to the secondary…
Mar 10, 2017Class IZ-1832-2017Device DesignIn April 2015, HeartWare, now a part of Medtronic, notified users of a safety issue related to the HVAD Pump Driveline Splice Kit. Medtronic now intends to send a follow-up communication to identified clinicians informing them that a new…
Feb 3, 2017Class IZ-1538-2017Software designProduct recall of all HVAD Controllers currently on the market and to exchange them for the next generation HVAD Controller (Controller2.0)
Aug 17, 2016Class IZ-0006-2017Component design/selectionForeign material found within the driveline connector of the HVAD Pump that could cause increased impedance resulting in electrical faults.
Jun 8, 2016Class IZ-0005-2017Device DesignLoose connector ports in controller body.
Jan 7, 2016Class IZ-1887-2016Software designHeartware Ventricular Assist System Battery Cell experiencing premature depletion.

No ancestor or descendant has a Class I recall.

How does its category compare?

All DSQ devices

DSQ Ventricular (Assist) Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · DSQ

50in 2025

2007: 72008: 52009: 92010: 122011: 42012: 202013: 362014: 302015: 242016: 372017: 362018: 352019: 342020: 482021: 382022: 332023: 482024: 612025: 502026 (partial year): 46
2007high 61 (2024) · low 4 (2011)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DSQ

Too few DSQ 510(k)s in any year since 2007 to give a median.

Recalls per year · DSQ

4in 2025

2007: 42008: 12009: 42010: 12012: 12016: 2, 2 Class I2018: 32019: 12020: 1, 1 Class I2021: 12022: 112023: 92024: 1, 1 Class I2025: 42026 (partial year): 4
2007high 11 (2022) · low 0 (2011)2026

Class I (4 since 2007) Class II/III not classified

Adverse event reports per year · DSQ

10,419in 2025

2007: 2242008: 2642009: 5292010: 6872011: 1,0822012: 2,1202013: 2,9412014: 3,9032015: 6,7712016: 7,1902017: 9,0552018: 12,6602019: 20,2162020: 14,2392021: 15,7662022: 15,9942023: 13,0552024: 13,0462025: 10,419
2007high 20,216 (2019) · low 224 (2007)2025

MAUDE reports naming product code DSQ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20077—4224
20085—1264
20099—4529
201012—1687
20114—01,082
201220—12,120
201336—02,941
201430—03,903
201524—06,771
201637—2 (2)7,190
201736—09,055
201835—312,660
201934—120,216
202048—1 (1)14,239
202138—115,766
202233—1115,994
202348—913,055
202461—1 (1)13,046
202550—410,419
202646—4—
Cite this page

510(k) Report. “P100047: Heartware Ventricular Assist System.” Lineage record from FDA PMA data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/P100047

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Page last updated Nov 3, 2026