P110042PMAApproved Sep 28, 2012280-day reviewClass IIIRecalls 11

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Boston Scientific Corporation · LWS Implantable Cardioverter Defibrillator (Non-Crt)

Approved on Sep 28, 2012 through premarket approval, which rests on the device’s own clinical evidence after a 280-day review. It has been recalled 11 times.

Decided
Sep 28, 2012
Review
280 days
Class
Class III
Product code
LWS
Recalls
11

PMA approvals rest on their own clinical evidence and do not claim a predicate, so this family tree starts here.

P110042 has 208 supplements since 2012: changes to the approved device, its labelling or its manufacturing, each approved separately.

PMA supplements approved per year

2012: 12013: 202014: 182015: 102016: 122017: 302018: 202019: 132020: 192021: 172022: 102023: 152024: 92025: 22026: 12
2012high 30 (2017) · low 1 (2012)2026
View FIG. 1 as a table
YearSupplements approved
20121
201320
201418
201510
201612
201730
201820
201913
202019
202117
202210
202315
20249
20252
202612
FIG. 1 — Supplements to P110042, 2012–2026Source: FDA PMA database

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA PMA database (decision fields and supplements), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LWS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

Does it have a predicate?

PMA decisions do not claim a predicate device.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 11, 2023Class IIZ-2441-2023Software designThere is a potential for a rare interaction between the EMBLEM S-ICD and LATITUDE communicator, which may cause S-ICD sensing disablement for a 24-hour interval.
Jul 11, 2023Class IIZ-2442-2023Software designThere is a potential for a rare interaction between the EMBLEM S-ICD and LATITUDE communicator, which may cause S-ICD sensing disablement for a 24-hour interval.
Sep 30, 2022Class IIZ-0152-2023Under Investigation by firmThere is an incorrect manufacturing date/timestamp within the software which causes an inaccurate display of battery capacity.
Sep 30, 2022Class IIZ-0151-2023Under Investigation by firmThere is an incorrect manufacturing date/timestamp within the software which causes an inaccurate display of battery capacity.
Nov 1, 2018Class IIZ-0474-2019Device DesignExperienced accelerated battery depletion and a shortened replacement interval due to latent internal battery malfunctions.
Jun 29, 2017Class IIZ-3037-2017Software designThe device can deliver an atypical amount of energy due to memory corruption inside the device.
Jun 29, 2017Class IIZ-3039-2017Software designThe device can deliver an atypical amount of energy due to memory corruption inside the device.
Jun 29, 2017Class IIZ-3038-2017Software designThe device can deliver an atypical amount of energy due to memory corruption inside the device.
Jan 12, 2017Class IIZ-1178-2017Software designThere is a potential for radio frequency (RF) interference to alter wireless communication from a programmer, which in rare instances may cause an S-ICD to perform an unintended command. This behavior can only occur during an active…
Sep 17, 2015Class IIZ-0232-2016Process designDevices shipped to US in dual-channel rather than US approved single-channel RF Telemetry communication mode. While this does not impact therapy delivery of device; this can impact likelihood of successfully establishing telemetry during…
Mar 1, 2013Class IIZ-0965-2013Software designAn internal protective fuse can be unintentionally activated while the device is charging its capacitors for shock delivery or induction. Should this occur, the defibrillator would not be able to deliver therapy or communicate with the…

No ancestor or descendant has a Class I recall.

How does its category compare?

All LWS devices

LWS Implantable Cardioverter Defibrillator (Non-Crt): FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · LWS

97in 2025

2007: 702008: 942009: 802010: 1192011: 1362012: 1292013: 1482014: 1682015: 1002016: 1452017: 1622018: 1422019: 982020: 1252021: 1112022: 912023: 1012024: 1082025: 972026 (partial year): 93
2007high 168 (2014) · low 70 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LWS

Too few LWS 510(k)s in any year since 2007 to give a median.

Recalls per year · LWS

2in 2025

2007: 42008: 32009: 82010: 62011: 82012: 52013: 12015: 2, 1 Class I2016: 2, 2 Class I2017: 7, 5 Class I2018: 12020: 6, 3 Class I2022: 32023: 22025: 2, 1 Class I
2007high 8 (2009) · low 0 (2014)2026

Class I (12 since 2007) Class II/III not classified

Adverse event reports per year · LWS

24,303in 2025

2007: 10,3302008: 11,3272009: 13,7682010: 17,3212011: 15,0972012: 12,8202013: 17,2422014: 18,5942015: 14,0402016: 16,6382017: 18,4542018: 17,2282019: 16,1842020: 16,9092021: 19,6542022: 23,4542023: 28,7012024: 25,2022025: 24,303
2007high 28,701 (2023) · low 10,330 (2007)2025

MAUDE reports naming product code LWS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200770—410,330
200894—311,327
200980—813,768
2010119—617,321
2011136—815,097
2012129—512,820
2013148—117,242
2014168—018,594
2015100—2 (1)14,040
2016145—2 (2)16,638
2017162—7 (5)18,454
2018142—117,228
201998—016,184
2020125—6 (3)16,909
2021111—019,654
202291—323,454
2023101—228,701
2024108—025,202
202597—2 (1)24,303
202693—0—
Cite this page

510(k) Report. “P110042: Subcutaneous Implantable Defibrillator (S-Icd) System.” Lineage record from FDA PMA data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/P110042

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Page last updated Nov 3, 2026