510(k) Report

PMA · Approved

Recalled · Class II · 2026

Minimed 530G System

P120010PMAApproved Sep 26, 2013478-day reviewClass IIIRecalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medtronic, Inc. · OZO Automated Insulin Dosing , Threshold Suspend

Approved on Sep 26, 2013 through premarket approval, which rests on the device’s own clinical evidence after a 478-day review. It has been recalled 2 times.

Decided
Sep 26, 2013
Review
478 days
Class
Class III
Product code
OZO
Recalls
2

PMA approvals rest on their own clinical evidence and do not claim a predicate, so this family tree starts here.

P120010 has 135 supplements since 2013: changes to the approved device, its labelling or its manufacturing, each approved separately.

PMA supplements approved per year

2013: 22014: 322015: 362016: 222017: 82018: 202019: 92020: 12021: 22022: 22026: 1
2013high 36 (2015) · low 0 (2023)2026
View FIG. 1 as a table
YearSupplements approved
20132
201432
201536
201622
20178
201820
20199
20201
20212
20222
20261
FIG. 1 — Supplements to P120010, 2013–2026Source: FDA PMA database

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA PMA database (decision fields and supplements), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in OZO? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

Does it have a predicate?

PMA decisions do not claim a predicate device.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 13, 2026Class IIZ-1751-2026Device DesignAll Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site…
Mar 13, 2014Class IIZ-2544-2014Device DesignMedtronic is recalling the Medtronic MiniMed Paradigm Insulin Infusion Pumps because they have received reports that users have accidentally programmed the pump to deliver the maximum bolus amount.

No ancestor or descendant has a Class I recall.

How does its category compare?

All OZO devices

OZO Automated Insulin Dosing , Threshold Suspend: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · OZO

1in 2025

2013: 32014: 322015: 352016: 282017: 292018: 532019: 312020: 82021: 72022: 92023: 32024: 22025: 12026 (partial year): 2
2007high 53 (2018) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OZO

Too few OZO 510(k)s in any year since 2007 to give a median.

Recalls per year · OZO

1in 2024

2017: 12018: 12019: 1, 1 Class I2020: 12024: 1, 1 Class I
2007high 1 (2017) · low 0 (2007)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · OZO

3,599in 2025

2014: 15,0802015: 28,7632016: 21,2822017: 30,4352018: 39,9602019: 82,3922020: 77,8122021: 42,5772022: 54,9242023: 29,7392024: 11,8142025: 3,599
2007high 82,392 (2019) · low 0 (2007)2025

MAUDE reports naming product code OZO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20110—00
20120—00
20133—00
201432—015,080
201535—028,763
201628—021,282
201729—130,435
201853—139,960
201931—1 (1)82,392
20208—177,812
20217—042,577
20229—054,924
20233—029,739
20242—1 (1)11,814
20251—03,599
20262—0—
Cite this page

510(k) Report. “P120010: Minimed 530G System.” Lineage record from FDA PMA data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/P120010

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026