510(k) Report

PMA · Approved

Recalled · Class I · 2020

Medtronic Corevalve System

P130021PMAApproved Jan 17, 2014179-day reviewClass IIIRecalls 4 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medtronic, Inc. · NPT Aortic Valve, Prosthesis, Percutaneously Delivered

Approved on Jan 17, 2014 through premarket approval, which rests on the device’s own clinical evidence after a 179-day review. It has a Class I recall. It has been recalled 4 times, including a Class I recall.

Decided
Jan 17, 2014
Review
179 days
Class
Class III
Product code
NPT
Recalls
4 (1 Class I)

PMA approvals rest on their own clinical evidence and do not claim a predicate, so this family tree starts here.

P130021 has 182 supplements since 2014: changes to the approved device, its labelling or its manufacturing, each approved separately.

PMA supplements approved per year

2014: 82015: 82016: 102017: 182018: 102019: 132020: 122021: 222022: 242023: 222024: 162025: 102026: 9
2014high 24 (2022) · low 8 (2014)2026
View FIG. 1 as a table
YearSupplements approved
20148
20158
201610
201718
201810
201913
202012
202122
202224
202322
202416
202510
20269
FIG. 1 — Supplements to P130021, 2014–2026Source: FDA PMA database

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA PMA database (decision fields and supplements), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

Does it have a predicate?

PMA decisions do not claim a predicate device.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 23, 2020Class IIZ-0559-2021Labeling designPotential for Transcatheter Aortic Valve (TAV) leaflet damage when performing a post-implant balloon dilatation (PID).
Oct 23, 2020Class IIZ-0560-2021Labeling designPotential for Transcatheter Aortic Valve (TAV) leaflet damage when performing a post-implant balloon dilatation (PID).
Oct 23, 2020Class IIZ-0561-2021Labeling designPotential for Transcatheter Aortic Valve (TAV) leaflet damage when performing a post-implant balloon dilatation (PID).
Jul 15, 2015Class IZ-2264-2015Process controlMedtronic is recalling the EnVeo R Loading Systems because of the presence of particulate.

No ancestor or descendant has a Class I recall.

How does its category compare?

All NPT devices

NPT Aortic Valve, Prosthesis, Percutaneously Delivered: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · NPT

29in 2025

2011: 12012: 202013: 482014: 562015: 432016: 612017: 652018: 432019: 592020: 482021: 422022: 412023: 492024: 332025: 292026 (partial year): 34
2007high 65 (2017) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NPT

Too few NPT 510(k)s in any year since 2007 to give a median.

Recalls per year · NPT

5in 2024

2014: 4, 3 Class I2017: 1, 1 Class I2019: 12020: 9, 6 Class I2024: 52026 (partial year): 3
2007high 9 (2020) · low 0 (2007)2026

Class I (10 since 2007) Class II/III not classified

Adverse event reports per year · NPT

12,906in 2025

2012: 982013: 1,0962014: 2,1612015: 2,8622016: 2,9962017: 3,1152018: 3,9162019: 4,3612020: 5,8062021: 5,6042022: 5,9942023: 8,3192024: 11,0432025: 12,906
2007high 12,906 (2025) · low 0 (2007)2025

MAUDE reports naming product code NPT, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20111—00
201220—098
201348—01,096
201456—4 (3)2,161
201543—02,862
201661—02,996
201765—1 (1)3,115
201843—03,916
201959—14,361
202048—9 (6)5,806
202142—05,604
202241—05,994
202349—08,319
202433—511,043
202529—012,906
202634—3—
Cite this page

510(k) Report. “P130021: Medtronic Corevalve System.” Lineage record from FDA PMA data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/P130021

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Page last updated Nov 3, 2026