510(k) Report

PMA · Approved

Recalled · Class II · 2023

Tacticath Quartz Set

P130026PMAApproved Oct 24, 2014329-day reviewClass IIIRecalls 6

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

St Jude Medical · OAE Catheter, Percutaneous, Cardiac Ablation, For Treatment Of Atrial Fibrillation

Approved on Oct 24, 2014 through premarket approval, which rests on the device’s own clinical evidence after a 329-day review. It has been recalled 6 times.

Decided
Oct 24, 2014
Review
329 days
Class
Class III
Product code
OAE
Recalls
6

PMA approvals rest on their own clinical evidence and do not claim a predicate, so this family tree starts here.

P130026 has 99 supplements since 2015: changes to the approved device, its labelling or its manufacturing, each approved separately.

PMA supplements approved per year

2015: 102016: 112017: 32018: 92019: 182020: 112021: 102022: 72023: 62024: 42025: 42026: 6
2015high 18 (2019) · low 3 (2017)2026
View FIG. 1 as a table
YearSupplements approved
201510
201611
20173
20189
201918
202011
202110
20227
20236
20244
20254
20266
FIG. 1 — Supplements to P130026, 2015–2026Source: FDA PMA database

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA PMA database (decision fields and supplements), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in OAE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

Does it have a predicate?

PMA decisions do not claim a predicate device.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 18, 2023Class IIZ-0814-2024Under Investigation by firmWhen the catheters are used with the EnSiteX EP System, the system does not correctly recognize the catheter as the appropriate product model and instead reads and displays it as a different model. This results in some features specific…
Mar 24, 2022Class IIZ-0956-2022Process controlDevices were manufactured with a "First Use Date" preprogrammed in the EEPROM which disables the use of the single-use device.
Mar 24, 2022Class IIZ-0957-2022Process controlDevices were manufactured with a "First Use Date" preprogrammed in the EEPROM which disables the use of the single-use device.
May 22, 2020Class IIZ-2541-2020Process controlHigh rate of reported complaints for the loss of contact force information while using the TactiCath Quartz Contact Force Ablation Catheter
Apr 5, 2019Class IIZ-1493-2019Software designIn reported cases, the device log on the TactiSys Quartz Equipment operating on Software Version 1.7.0 fills the allocated disk space, which prevents the storage of new log data. This may lead to intermittent contact force data to be…
Nov 25, 2014Class IIZ-1284-2015Software designA configuration update needs to be done on TactiSys to appropriately recognize all TactiCath catheters.

No ancestor or descendant has a Class I recall.

How does its category compare?

All OAE devices

OAE Catheter, Percutaneous, Cardiac Ablation, For Treatment Of Atrial Fibrillation: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · OAE

20in 2025

2009: 12010: 12011: 142012: 102013: 102014: 152015: 152016: 192017: 152018: 282019: 342020: 272021: 312022: 262023: 432024: 322025: 202026 (partial year): 25
2007high 43 (2023) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OAE

Too few OAE 510(k)s in any year since 2007 to give a median.

Recalls per year · OAE

5in 2025

2010: 52011: 12019: 12022: 22023: 12024: 7, 7 Class I2025: 5, 5 Class I
2007high 7 (2024) · low 0 (2007)2026

Class I (12 since 2007) Class II/III not classified

Adverse event reports per year · OAE

1,412in 2025

2009: 22011: 692012: 702013: 1112014: 1312015: 2742016: 3112017: 3872018: 8842019: 9082020: 7442021: 9142022: 1,1612023: 1,3112024: 2,0982025: 1,412
2007high 2,098 (2024) · low 0 (2007)2025

MAUDE reports naming product code OAE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20091—02
20101—50
201114—169
201210—070
201310—0111
201415—0131
201515—0274
201619—0311
201715—0387
201828—0884
201934—1908
202027—0744
202131—0914
202226—21,161
202343—11,311
202432—7 (7)2,098
202520—5 (5)1,412
202625—0—
Cite this page

510(k) Report. “P130026: Tacticath Quartz Set.” Lineage record from FDA PMA data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/P130026

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026