510(k) Report

PMA · Approved

Recalled · Class I · 2026

IMPELLA 2.5 System

P140003PMAApproved Mar 23, 2015370-day reviewClass IIIRecalls 25 (15 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Abiomed, Inc. · OZD Temporary Non-Roller Type Left Heart Support Blood Pump

Approved on Mar 23, 2015 through premarket approval, which rests on the device’s own clinical evidence after a 370-day review. It has a Class I recall. It has been recalled 25 times, including 15 Class I recalls.

Decided
Mar 23, 2015
Review
370 days
Class
Class III
Product code
OZD
Recalls
25 (15 Class I)

PMA approvals rest on their own clinical evidence and do not claim a predicate, so this family tree starts here.

P140003 has 127 supplements since 2015: changes to the approved device, its labelling or its manufacturing, each approved separately.

PMA supplements approved per year

2015: 42016: 82017: 122018: 172019: 192020: 122021: 132022: 152023: 82024: 52025: 112026: 3
2015high 19 (2019) · low 3 (2026)2026
View FIG. 1 as a table
YearSupplements approved
20154
20168
201712
201817
201919
202012
202113
202215
20238
20245
202511
20263
FIG. 1 — Supplements to P140003, 2015–2026Source: FDA PMA database

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA PMA database (decision fields and supplements), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

Does it have a predicate?

PMA decisions do not claim a predicate device.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 18, 2026Not classifiedZ-0208-2027Under Investigation by firmRetrospective review of servicing records identified AIC units requiring various hardware, firmware, and/or software update(s): 1. RFID firmware issue that may result in failure of the AIC to recognize the purge cassette during case setup…
Sep 18, 2026Not classifiedZ-0209-2027Under Investigation by firmRetrospective review of servicing records identified AIC units requiring various hardware, firmware, and/or software update(s): 1. RFID firmware issue that may result in failure of the AIC to recognize the purge cassette during case setup…
Sep 18, 2026Not classifiedZ-0210-2027Under Investigation by firmRetrospective review of servicing records identified AIC units requiring various hardware, firmware, and/or software update(s): 1. RFID firmware issue that may result in failure of the AIC to recognize the purge cassette during case setup…
Aug 12, 2026Class IZ-3149-2026Under Investigation by firmThe potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.
Aug 12, 2026Class IZ-3148-2026Under Investigation by firmThe potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.
Aug 12, 2026Class IZ-3150-2026Under Investigation by firmThe potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.
May 18, 2026Class IZ-2457-2026Under Investigation by firmExposure to the low purge pressure occurrence may result in persistent low purge pressure alarms and, in some cases, interruption or loss of mechanical circulatory support.
May 14, 2026Class IZ-2360-2026Under Investigation by firmPotential software error in the Automated Impella Controller ("AIC" ) when used in conjunction with left ventricular Impella devices.
Apr 20, 2026Class IZ-2211-2026Under Investigation by firmA retrospective review identified that there are AIC units requiring specific hardware updates to address potential safety concerns. These hardware updates have been implemented through Abiomed s servicing process; therefore, AICs that…
Feb 18, 2026Class IZ-1589-2026Device DesignIncreased risk of purge leaks with Generation 1 purge cassettes.
Feb 16, 2026Class IZ-1591-2026Software designPotential that the "Purge System Blocked" alarm display on the Automated Impella Controller (AIC) could be delayed when using first generation Impella 5.5 (0550-0008*) pumps.
Dec 4, 2025Class IIZ-1030-2026Under Investigation by firmDevice packaged in incorrect outer box carton.
Oct 1, 2025Class IZ-0069-2026Software designPotential cybersecurity vulnerabilities related to the operating system in the Automated Impella Controller (AIC).
Sep 16, 2025Class IZ-0136-2026Device DesignPotential for Automated Impella Controller (AIC) purge pressure issues due to purge retainer failures.
Jun 23, 2025Class IZ-2164-2025Device DesignA potential issue with the Automated Impella Controller (AIC) not detecting an Impella pump when it is connected
May 31, 2024Class IZ-2319-2024Release of Material/Component prior to receiving test resultsNine (9) Impella CP pumps failed inspection and were inadvertently released.
Dec 27, 2023Class IZ-0980-2024Labeling Change ControlIFU has been updated to include warnings about the risk of the inlet perforating through the myocardial wall of the left ventricle due to operator handling.
Dec 6, 2023Class IIZ-0636-2024No Marketing ApplicationThe following features have been disabled from the web-based portal because the FDA has not evaluated these features for safety and effectiveness: Notifications via email feature; AIC alarm color on case tile feature; Sort case tiles by…
Oct 10, 2023Class IIZ-0440-2024Software designPump not detected as connected to controller due to software issue.
Oct 10, 2023Class IIZ-0441-2024Software designPump not detected as connected to controller due to software issue.
Apr 17, 2023Class IZ-1590-2023Device DesignThe Impella 5.5 pump experienced a heightened complaint rate for purge leaks.
Mar 1, 2021Class IIZ-1175-2024Software designRetrospective reporting for the release of Technical Bulletin IMP-2643 AIC Version 8.5 Software Update Available. The update resolved an issue in version 8.4 where a pump was not recognized by the AIC.
Sep 4, 2019Class IIZ-0345-2020Under Investigation by firmThis one pump shipped with incorrect pump programming, the initial production working version of the program. The pump would have immediately run at P9 for all selected P-levels, rather than a 10 second ramp up period. The pump may have…
Feb 19, 2019Class IIZ-3251-2026Under Investigation by firmRetrospective report of issues remedied in 2019 with firmware and software patches: 1. Defects in optical bench (OPB) firmware V2.19, which affected the accuracy and stability of the pressure signal the AIC system performance depends on.…
Jan 17, 2011Class IZ-2238-2026Under Investigation by firmRetrospective submission for following issues identified: 1. Alarm failures, power-path faults, electrical-short risks, and inaccurate motor-current sensing were identified in certain AIC consoles. 2. There is improper alignment between…

No ancestor or descendant has a Class I recall.

How does its category compare?

All OZD devices

OZD Temporary Non-Roller Type Left Heart Support Blood Pump: FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.

FDA decisions per year · OZD

11in 2025

2015: 52016: 82017: 122018: 172019: 192020: 122021: 132022: 152023: 82024: 52025: 112026 (partial year): 3
2007high 19 (2019) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OZD

Too few OZD 510(k)s in any year since 2007 to give a median.

Recalls per year · OZD

9so far in 2026, the most since 2007

2011: 1, 1 Class I2019: 22023: 3, 1 Class I2025: 3, 2 Class I2026 (partial year): 9, 6 Class I
2007high 9 (2026) · low 0 (2007)2026

Class I (10 since 2007) Class II/III not classified

Adverse event reports per year · OZD

9,726in 2025

2016: 182017: 1092018: 1492019: 3002020: 3412021: 5672022: 5422023: 3,8492024: 18,5902025: 9,726
2007high 18,590 (2024) · low 0 (2007)2025

MAUDE reports naming product code OZD, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20110—1 (1)0
20120—00
20130—00
20140—00
20155—00
20168—018
201712—0109
201817—0149
201919—2300
202012—0341
202113—0567
202215—0542
20238—3 (1)3,849
20245—018,590
202511—3 (2)9,726
20263—9 (6)—
Cite this page

510(k) Report. “P140003: IMPELLA 2.5 System.” Lineage record from FDA PMA data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/P140003

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.