Imageready MR Conditional Pacing System and Ingevity Pace/Sense Lead

P150012PMAApproved Apr 25, 2016419-day reviewClass IIIRecalls 28 (14 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Boston Scientific · LWP Implantable Pulse Generator, Pacemaker (Non-Crt)

Approved on Apr 25, 2016 through premarket approval, which rests on the device’s own clinical evidence after a 419-day review. It has a Class I recall. It has been recalled 28 times, including 14 Class I recalls.

Decided
Apr 25, 2016
Review
419 days
Class
Class III
Product code
LWP
Recalls
28 (14 Class I)

PMA approvals rest on their own clinical evidence and do not claim a predicate, so this family tree starts here.

P150012 has 186 supplements since 2016: changes to the approved device, its labelling or its manufacturing, each approved separately.

PMA supplements approved per year

2016: 202017: 292018: 152019: 212020: 172021: 172022: 172023: 152024: 102025: 92026: 16
2016high 29 (2017) · low 9 (2025)2026
View FIG. 1 as a table
YearSupplements approved
201620
201729
201815
201921
202017
202117
202217
202315
202410
20259
202616
FIG. 1 — Supplements to P150012, 2016–2026Source: FDA PMA database

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA PMA database (decision fields and supplements), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

Does it have a predicate?

PMA decisions do not claim a predicate device.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 19, 2026Class IZ-1770-2026Nonconforming Material/ComponentUpdated software, Brady software maintenance release 6 (SMR6), is now available for the ACCOLADE" family of pacemakers and cardiac resynchronization therapy pacemakers (CRT-Ps)1 and the advisory population is expanding to include all…
Aug 20, 2025Class IZ-0084-2026OtherSoftware to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT…
Aug 20, 2025Class IZ-0088-2026OtherSoftware to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT…
Aug 20, 2025Class IZ-0095-2026OtherSoftware to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT…
Aug 20, 2025Class IZ-0081-2026OtherSoftware to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT…
Aug 20, 2025Class IZ-0091-2026OtherSoftware to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT…
Aug 20, 2025Class IZ-0082-2026OtherSoftware to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT…
Aug 20, 2025Class IZ-0089-2026OtherSoftware to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT…
Aug 20, 2025Class IZ-0094-2026OtherSoftware to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT…
Aug 20, 2025Class IZ-0097-2026OtherSoftware to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT…
Dec 12, 2024Class IZ-0822-2025Process designA subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery…
Dec 12, 2024Class IZ-0824-2025Process designA subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery…
Dec 12, 2024Class IZ-0823-2025Process designA subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery…
Jun 3, 2021Class IIZ-1998-2021Nonconforming Material/ComponentLatent release of small amounts of hydrogen within the pacemaker may cause a low voltage capacitor to become electrically compromised over time resulting in accelerated battery depletion of the battery and associated progression of…
Jun 3, 2021Class IZ-2017-2021Device DesignThe dual chamber INGENIO family pacemakers or cardiac resynchronization therapy pacemakers (CRT-Ps) may initiate Safety Mode later in device life (i.e., prior to reaching the Explant battery indicator) when the device's battery exhibits…
Jun 3, 2021Class IIZ-1999-2021Nonconforming Material/ComponentLatent release of small amounts of hydrogen within the pacemaker may cause a low voltage capacitor to become electrically compromised over time resulting in accelerated battery depletion of the battery and associated progression of…
Jun 3, 2021Class IIZ-2000-2021Nonconforming Material/ComponentLatent release of small amounts of hydrogen within the pacemaker may cause a low voltage capacitor to become electrically compromised over time resulting in accelerated battery depletion of the battery and associated progression of…
Sep 10, 2018Class IIZ-3256-2018Component design/selectionBoston Scientific has identified a subset of active pacemaker and CRT-Ps systems with an elevated potential for early pacemaker replacement due to hydrogen-induced accelerated battery depletion.
Sep 10, 2018Class IIZ-3257-2018Component design/selectionBoston Scientific has identified a subset of active pacemaker and CRT-Ps systems with an elevated potential for early pacemaker replacement due to hydrogen-induced accelerated battery depletion.
Sep 10, 2018Class IIZ-3255-2018Component design/selectionBoston Scientific has identified a subset of active pacemaker and CRT-Ps systems with an elevated potential for early pacemaker replacement due to hydrogen-induced accelerated battery depletion.
Dec 7, 2017Class IIZ-0370-2018Device DesignBoston Scientific has received reports of intermittent over-sensing of the Minute Ventilation (MV) sensor signal with certain Boston Scientific pacemaker and cardiac resynchronization therapy pacemaker systems (pacemakers).
Dec 7, 2017Class IIZ-0358-2018Device DesignBoston Scientific has received reports of intermittent over-sensing of the Minute Ventilation (MV) sensor signal with certain Boston Scientific pacemaker and cardiac resynchronization therapy pacemaker systems (pacemakers).
Dec 7, 2017Class IIZ-0359-2018Device DesignBoston Scientific has received reports of intermittent over-sensing of the Minute Ventilation (MV) sensor signal with certain Boston Scientific pacemaker and cardiac resynchronization therapy pacemaker systems (pacemakers).
Dec 7, 2017Class IIZ-0369-2018Device DesignBoston Scientific has received reports of intermittent over-sensing of the Minute Ventilation (MV) sensor signal with certain Boston Scientific pacemaker and cardiac resynchronization therapy pacemaker systems (pacemakers).
Dec 7, 2017Class IIZ-0377-2018Device DesignBoston Scientific has received reports of intermittent over-sensing of the Minute Ventilation (MV) sensor signal with certain Boston Scientific pacemaker and cardiac resynchronization therapy pacemaker systems (pacemakers).
Dec 7, 2017Class IIZ-0357-2018Device DesignBoston Scientific has received reports of intermittent over-sensing of the Minute Ventilation (MV) sensor signal with certain Boston Scientific pacemaker and cardiac resynchronization therapy pacemaker systems (pacemakers).
Nov 30, 2016Class IIZ-0938-2017Process controlBoston Scientific is conducting a recall on Boston Scientific INGEVITY" MRI endocardial pacing leads. There is a possibility that the polyurethane boot at the terminal end of the lead was not securely connected to the lead body.
Jun 15, 2016Class IIZ-2749-2016Process controlFive Accolate MRI and Essentio MRI DR pulse generators with an alternate MICS module configuration were shipped to US customers and implanted. This alternate configuration did not receive FDA approval yet.

No ancestor or descendant has a Class I recall.

How does its category compare?

All LWP devices

LWP Implantable Pulse Generator, Pacemaker (Non-Crt): FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · LWP

46in 2025

2007: 582008: 872009: 782010: 662011: 912012: 1232013: 1162014: 1262015: 972016: 972017: 1162018: 972019: 882020: 902021: 702022: 652023: 542024: 582025: 462026 (partial year): 40
2007high 126 (2014) · low 46 (2025)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LWP

Too few LWP 510(k)s in any year since 2007 to give a median.

Recalls per year · LWP

21in 2025

2007: 72008: 22009: 82010: 72013: 22014: 22015: 22016: 12017: 212024: 4, 4 Class I2025: 21, 21 Class I2026 (partial year): 1, 1 Class I
2007high 21 (2017) · low 0 (2011)2026

Class I (26 since 2007) Class II/III not classified

Adverse event reports per year · LWP

11,931in 2025

2007: 492008: 4962009: 8052010: 6782011: 6842012: 1,0872013: 5,1642014: 9,4002015: 9,4222016: 12,3542017: 11,6432018: 14,9362019: 9,6272020: 8,9892021: 9,3742022: 14,2862023: 12,6332024: 10,4952025: 11,931
2007high 14,936 (2018) · low 49 (2007)2025

MAUDE reports naming product code LWP, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200758—749
200887—2496
200978—8805
201066—7678
201191—0684
2012123—01,087
2013116—25,164
2014126—29,400
201597—29,422
201697—112,354
2017116—2111,643
201897—014,936
201988—09,627
202090—08,989
202170—09,374
202265—014,286
202354—012,633
202458—4 (4)10,495
202546—21 (21)11,931
202640—1 (1)—
Cite this page

510(k) Report. “P150012: Imageready MR Conditional Pacing System and Ingevity Pace/Sense Lead.” Lineage record from FDA PMA data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/P150012

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Page last updated Nov 3, 2026