510(k) Report

PMA · Approved

Recalled · Class II · 2020

Oncomine Dx Target Test

P160045PMAApproved Jun 22, 2017248-day reviewClass IIIRecalls 6

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Life Technologies Corporation · PQP Next Generation Sequencing Oncology Panel, Somatic Or Germline Variant Detection System

Approved on Jun 22, 2017 through premarket approval, which rests on the device’s own clinical evidence after a 248-day review. It has been recalled 6 times.

Decided
Jun 22, 2017
Review
248 days
Class
Class III
Product code
PQP
Recalls
6

PMA approvals rest on their own clinical evidence and do not claim a predicate, so this family tree starts here.

P160045 has 44 supplements since 2017: changes to the approved device, its labelling or its manufacturing, each approved separately.

PMA supplements approved per year

2017: 32018: 62019: 52020: 62021: 42022: 52023: 72024: 42025: 32026: 1
2017high 7 (2023) · low 1 (2026)2026
View FIG. 1 as a table
YearSupplements approved
20173
20186
20195
20206
20214
20225
20237
20244
20253
20261
FIG. 1 — Supplements to P160045, 2017–2026Source: FDA PMA database

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA PMA database (decision fields and supplements), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in PQP? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

Does it have a predicate?

PMA decisions do not claim a predicate device.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 16, 2020Class IIZ-1705-2020Incorrect or no expiration dateExtended Expiration Date prior to clearance
Mar 16, 2020Class IIZ-1706-2020Incorrect or no expiration dateExtended Expiration Date prior to clearance
Mar 16, 2020Class IIZ-1707-2020Incorrect or no expiration dateExtended Expiration Date prior to clearance
Mar 16, 2020Class IIZ-1704-2020Incorrect or no expiration dateExtended Expiration Date prior to clearance
Feb 8, 2018Class IIZ-1480-2018Process controlThe lower concentration of the RNA panel could cause a NO CALL or FALSE NEGATIVE for the reporting of ROS1 fusion. If a physician relies on a false negative result, crizotinib could be improperly excluded from the patient's treatment…
Dec 4, 2017Class IIZ-1279-2018Device DesignThe Oncomine Dx Target Test may report erroneous results when used off-label for samples other than non-small cell lung cancer (NSCLC) tumor specimens.

No ancestor or descendant has a Class I recall.

How does its category compare?

All PQP devices

PQP Next Generation Sequencing Oncology Panel, Somatic Or Germline Variant Detection System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · PQP

37so far in 2026, the most since 2007

2016: 12017: 62018: 152019: 212020: 252021: 172022: 332023: 342024: 342025: 212026 (partial year): 37
2007high 37 (2026) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · PQP

Too few PQP 510(k)s in any year since 2007 to give a median.

Recalls per year · PQP

3so far in 2026, the most since 2007

2018: 12019: 12022: 22026 (partial year): 3
2007high 3 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · PQP

2in 2023

2017: 12018: 12020: 12021: 12022: 12023: 2
2007high 2 (2023) · low 0 (2007)2025

MAUDE reports naming product code PQP, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20110—00
20120—00
20130—00
20140—00
20150—00
20161—00
20176—01
201815—11
201921—10
202025—01
202117—01
202233—21
202334—02
202434—00
202521—00
202637—3—
Cite this page

510(k) Report. “P160045: Oncomine Dx Target Test.” Lineage record from FDA PMA data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/P160045

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Page last updated Nov 3, 2026