510(k) Report

PMA · Approved

Recalled · Class I · 2026

HeartMate 3™ Left Ventricular Assist System

P160054PMAApproved Aug 23, 2017261-day reviewClass IIIRecalls 18 (9 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Abbott Medical · DSQ Ventricular (Assist) Bypass

Approved on Aug 23, 2017 through premarket approval, which rests on the device’s own clinical evidence after a 261-day review. It has a Class I recall. It has been recalled 18 times, including 9 Class I recalls.

Decided
Aug 23, 2017
Review
261 days
Class
Class III
Product code
DSQ
Recalls
18 (9 Class I)

PMA approvals rest on their own clinical evidence and do not claim a predicate, so this family tree starts here.

P160054 has 132 supplements since 2017: changes to the approved device, its labelling or its manufacturing, each approved separately.

PMA supplements approved per year

2017: 22018: 122019: 92020: 92021: 82022: 72023: 182024: 262025: 212026: 20
2017high 26 (2024) · low 2 (2017)2026
View FIG. 1 as a table
YearSupplements approved
20172
201812
20199
20209
20218
20227
202318
202426
202521
202620
FIG. 1 — Supplements to P160054, 2017–2026Source: FDA PMA database

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA PMA database (decision fields and supplements), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

Does it have a predicate?

PMA decisions do not claim a predicate device.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 13, 2026Class IIZ-2520-2026Under Investigation by firmDue to 11 volt Backup Battery failures.
May 13, 2026Class IIZ-2519-2026Under Investigation by firmDue to 11 volt Backup Battery failures.
May 13, 2026Class IIZ-2518-2026Under Investigation by firmDue to 11 volt Backup Battery failures.
Oct 9, 2025Class IIZ-0505-2026Device DesignDue to potential corrosion issues associated with the Backup Battery cable connector that may cause a Backup Battery Fault Alarm.
Oct 9, 2025Class IIZ-0506-2026Device DesignDue to potential corrosion issues associated with the Backup Battery cable connector that may cause a Backup Battery Fault Alarm.
Jun 23, 2025Class IZ-2290-2025Vendor change controlDue to misaligned V-Lock feature on the Mobile Power Unit power cords, which prevents the locking mechanism from properly engaging with the Mobile Power Unit and may cause unintended disconnect.
Jun 23, 2025Class IZ-2291-2025Vendor change controlDue to misaligned V-Lock feature on the Mobile Power Unit power cords, which prevents the locking mechanism from properly engaging with the Mobile Power Unit and may cause unintended disconnect.
Mar 13, 2025Class IZ-1570-2025Process controlLeft Ventricular Assist System Mobile Power Unit (MPU) may have electrical component that may cause MPU not to turn on, unprompted shut down, or turning off/restarting, may cause system controller to indicate Yellow Wrench or No External…
Jun 25, 2024Class IIZ-2583-2024Process designLeft ventricular assist system controller UI membrane/screen may lift along the edge of the controller housing, and if controller is exposed to fluid, ingress may occur, which may damage printed circuit board assemblies, which could…
May 8, 2024Class IZ-1921-2024Device DesignSystem monitor screen may display atypical behavior: Overlapping screens/buttons, Frozen screen, Distorted text, blanks, or zeroes in place of values, Unresponsive buttons where the user is unable to initiate a command, which may lead to…
Mar 1, 2024Class IZ-1692-2024Device DesignPotential for a leak path at the seal interface between the left ventricle assist system inflow cannula and the titanium apical cuff.
Mar 1, 2024Class IZ-1693-2024Device DesignPotential for a leak path at the seal interface between the left ventricle assist system inflow cannula and the titanium apical cuff.
Feb 19, 2024Class IZ-1301-2024Material/Component ContaminationObserved outflow graft deformation known as Extrinsic Outflow Graft Obstruction (EOGO) associated with Left Ventricular Assist System.
Jan 3, 2024Class IZ-0842-2024Software designDue to software and controller systems that results in unexpected pump start or pump stop.
Dec 2, 2019Class IIZ-0746-2020Device DesignExcessive static electricity can potentially cause unrecoverable power loss and damage to the mobile power unit that can be used to power the left ventricular assist system, which could lead to no or reduced blood flow from the left…
Mar 30, 2019Class IIZ-2349-2019Process designOutflow Graft leaking at the pump connection during the implantation process. If the C-Ring is insufficiently seated, it renders the device nonfunctional because the Outflow Graft connection will leak during the implant procedure.
Mar 30, 2019Class IIZ-2348-2019Process designOutflow Graft leaking at the pump connection during the implantation process. If the C-Ring is insufficiently seated, it renders the device nonfunctional because the Outflow Graft connection will leak during the implant procedure.
Apr 5, 2018Class IZ-1774-2018Device DesignReports of outflow graft twist occlusions. Patients whose devices experience outflow graft occlusions will experience a persistent low flow alarm. Outflow graft occlusions can result in serious adverse events such as hemodynamic…

No ancestor or descendant has a Class I recall.

How does its category compare?

All DSQ devices

DSQ Ventricular (Assist) Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · DSQ

50in 2025

2007: 72008: 52009: 92010: 122011: 42012: 202013: 362014: 302015: 242016: 372017: 362018: 352019: 342020: 482021: 382022: 332023: 482024: 612025: 502026 (partial year): 46
2007high 61 (2024) · low 4 (2011)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DSQ

Too few DSQ 510(k)s in any year since 2007 to give a median.

Recalls per year · DSQ

4in 2025

2007: 42008: 12009: 42010: 12012: 12016: 2, 2 Class I2018: 32019: 12020: 1, 1 Class I2021: 12022: 112023: 92024: 1, 1 Class I2025: 42026 (partial year): 4
2007high 11 (2022) · low 0 (2011)2026

Class I (4 since 2007) Class II/III not classified

Adverse event reports per year · DSQ

10,419in 2025

2007: 2242008: 2642009: 5292010: 6872011: 1,0822012: 2,1202013: 2,9412014: 3,9032015: 6,7712016: 7,1902017: 9,0552018: 12,6602019: 20,2162020: 14,2392021: 15,7662022: 15,9942023: 13,0552024: 13,0462025: 10,419
2007high 20,216 (2019) · low 224 (2007)2025

MAUDE reports naming product code DSQ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20077—4224
20085—1264
20099—4529
201012—1687
20114—01,082
201220—12,120
201336—02,941
201430—03,903
201524—06,771
201637—2 (2)7,190
201736—09,055
201835—312,660
201934—120,216
202048—1 (1)14,239
202138—115,766
202233—1115,994
202348—913,055
202461—1 (1)13,046
202550—410,419
202646—4—
Cite this page

510(k) Report. “P160054: HeartMate 3™ Left Ventricular Assist System.” Lineage record from FDA PMA data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/P160054

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Page last updated Nov 3, 2026