510(k) Report

PMA · Approved

Recalled · Class II · 2024

CentriMag Circulatory Support System

P170038PMAApproved Dec 6, 2019742-day reviewClass IIIRecalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Abbott · DSQ Ventricular (Assist) Bypass

Approved on Dec 6, 2019 through premarket approval, which rests on the device’s own clinical evidence after a 742-day review. It has been recalled once.

Decided
Dec 6, 2019
Review
742 days
Class
Class III
Product code
DSQ
Recalls
1

PMA approvals rest on their own clinical evidence and do not claim a predicate, so this family tree starts here.

P170038 has 32 supplements since 2020: changes to the approved device, its labelling or its manufacturing, each approved separately.

PMA supplements approved per year

2020: 42021: 32022: 22023: 22024: 92025: 52026: 7
2020high 9 (2024) · low 2 (2022)2026
View FIG. 1 as a table
YearSupplements approved
20204
20213
20222
20232
20249
20255
20267
FIG. 1 — Supplements to P170038, 2020–2026Source: FDA PMA database

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA PMA database (decision fields and supplements), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DSQ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

Does it have a predicate?

PMA decisions do not claim a predicate device.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 22, 2024Class IIZ-2215-2024Process controlDue to a component change, the console part of a circulatory support system may not conform to the IEC 61000-4-5 standard, so if subjected to a power surge above 1.8kV and up to 2.0 kV, the console may shut down completely without alarm…

No ancestor or descendant has a Class I recall.

How does its category compare?

All DSQ devices

DSQ Ventricular (Assist) Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years. 2019, this device’s year, is highlighted.

FDA decisions per year · DSQ

50in 2025

2007: 72008: 52009: 92010: 122011: 42012: 202013: 362014: 302015: 242016: 372017: 362018: 352019: 342020: 482021: 382022: 332023: 482024: 612025: 502026 (partial year): 46
2007high 61 (2024) · low 4 (2011)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DSQ

Too few DSQ 510(k)s in any year since 2007 to give a median.

Recalls per year · DSQ

4in 2025

2007: 42008: 12009: 42010: 12012: 12016: 2, 2 Class I2018: 32019: 12020: 1, 1 Class I2021: 12022: 112023: 92024: 1, 1 Class I2025: 42026 (partial year): 4
2007high 11 (2022) · low 0 (2011)2026

Class I (4 since 2007) Class II/III not classified

Adverse event reports per year · DSQ

10,419in 2025

2007: 2242008: 2642009: 5292010: 6872011: 1,0822012: 2,1202013: 2,9412014: 3,9032015: 6,7712016: 7,1902017: 9,0552018: 12,6602019: 20,2162020: 14,2392021: 15,7662022: 15,9942023: 13,0552024: 13,0462025: 10,419
2007high 20,216 (2019) · low 224 (2007)2025

MAUDE reports naming product code DSQ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20077—4224
20085—1264
20099—4529
201012—1687
20114—01,082
201220—12,120
201336—02,941
201430—03,903
201524—06,771
201637—2 (2)7,190
201736—09,055
201835—312,660
201934—120,216
202048—1 (1)14,239
202138—115,766
202233—1115,994
202348—913,055
202461—1 (1)13,046
202550—410,419
202646—4—
Cite this page

510(k) Report. “P170038: CentriMag Circulatory Support System.” Lineage record from FDA PMA data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/P170038

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026