510(k) Report

PMA · Approved

Recalled · Class II · 2025

Abre Venous Self-expanding Stent System

P200026PMAApproved Oct 21, 2020177-day reviewClass IIIRecalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medtronic Vascular, Inc. · QAN Stent, Iliac Vein

Approved on Oct 21, 2020 through premarket approval, which rests on the device’s own clinical evidence after a 177-day review. It has been recalled 3 times.

Decided
Oct 21, 2020
Review
177 days
Class
Class III
Product code
QAN
Recalls
3

PMA approvals rest on their own clinical evidence and do not claim a predicate, so this family tree starts here.

P200026 has 19 supplements since 2021: changes to the approved device, its labelling or its manufacturing, each approved separately.

PMA supplements approved per year

2021: 32022: 22023: 12024: 62025: 32026: 4
2021high 6 (2024) · low 1 (2023)2026
View FIG. 1 as a table
YearSupplements approved
20213
20222
20231
20246
20253
20264
FIG. 1 — Supplements to P200026, 2021–2026Source: FDA PMA database

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA PMA database (decision fields and supplements), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in QAN? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

Does it have a predicate?

PMA decisions do not claim a predicate device.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 4, 2025Class IIZ-1098-2026Under Investigation by firmStents may be impacted by a Nitinol anomaly which causes a reduced stent diameter when deployed and corresponding reduced radial force. Issue may result in stent migration.
Nov 19, 2021Class IIZ-0464-2022OtherThere have been reports of stent migration.
Mar 16, 2021Class IIZ-1532-2021Process controlIncorrect size printed on the device; packaging is labeled correctly.

No ancestor or descendant has a Class I recall.

How does its category compare?

All QAN devices

QAN Stent, Iliac Vein: FDA decisions, review times, recalls and adverse event reports over 20 years. 2020, this device’s year, is highlighted.

FDA decisions per year · QAN

5in 2025

2019: 42020: 102021: 112022: 52023: 72024: 92025: 52026 (partial year): 8
2007high 11 (2021) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · QAN

Too few QAN 510(k)s in any year since 2007 to give a median.

Recalls per year · QAN

1in 2021

2021: 1
2007high 1 (2021) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · QAN

177in 2025

2019: 242020: 1172021: 4252022: 1152023: 2082024: 1282025: 177
2007high 425 (2021) · low 0 (2007)2025

MAUDE reports naming product code QAN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20110—00
20120—00
20130—00
20140—00
20150—00
20160—00
20170—00
20180—00
20194—024
202010—0117
202111—1425
20225—0115
20237—0208
20249—0128
20255—0177
20268—0—
Cite this page

510(k) Report. “P200026: Abre Venous Self-expanding Stent System.” Lineage record from FDA PMA data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/P200026

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Page last updated Nov 3, 2026