510(k) Report

PMA · Approved

Recalled · Class II · 2024

Itrel (R) Totally Implantable Spinal Cord Stim. Sys

P840001PMAApproved Nov 30, 1984322-day reviewClass IIIRecalls 9

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medtronic Neuromodulation · LGW Stimulator, Spinal-Cord, Totally Implanted For Pain Relief

Approved on Nov 30, 1984 through premarket approval, which rests on the device’s own clinical evidence after a 322-day review. It has been recalled 9 times.

Decided
Nov 30, 1984
Review
322 days
Class
Class III
Product code
LGW
Recalls
9

PMA approvals rest on their own clinical evidence and do not claim a predicate, so this family tree starts here.

P840001 has 612 supplements since 1985: changes to the approved device, its labelling or its manufacturing, each approved separately.

PMA supplements approved per year

1985: 41986: 41987: 31988: 31989: 31990: 41991: 21992: 21993: 31994: 11995: 71997: 21998: 11999: 22000: 42001: 92002: 72003: 42004: 32005: 82006: 82007: 52008: 192009: 202010: 282011: 292012: 292013: 272014: 282015: 302016: 322017: 362018: 312019: 342020: 262021: 232022: 302023: 242024: 282025: 292026: 20
1985high 36 (2017) · low 0 (1996)2026
View FIG. 1 as a table
YearSupplements approved
19854
19864
19873
19883
19893
19904
19912
19922
19933
19941
19957
19972
19981
19992
20004
20019
20027
20034
20043
20058
20068
20075
200819
200920
201028
201129
201229
201327
201428
201530
201632
201736
201831
201934
202026
202123
202230
202324
202428
202529
202620
FIG. 1 — Supplements to P840001, 1985–2026Source: FDA PMA database

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA PMA database (decision fields and supplements), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LGW? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

Does it have a predicate?

PMA decisions do not claim a predicate device.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 18, 2024Class IIZ-0372-2025Under Investigation by firmThe Wireless Rechargers in the kits may be unable to enter a recharging session upon first use.
Mar 28, 2024Class IIZ-1691-2024Under Investigation by firminability to reprogram one device
Dec 19, 2023Class IIZ-0780-2024Under Investigation by firmMedtronic has discovered an instance in which the Intellis A710 Clinician Programmer Application (CP App) could no longer establish a session with the Implantable Neurostimulator (INS) which resulted in the inability for the clinician to…
Jul 9, 2021Class IIZ-2159-2021Error in labelingSpecific PINs of the Vectris SureScan MRI lead kits contain the incorrect lead electrode spacing information printed on the shelf box and the sterile pack labeling.
May 3, 2016Class IIZ-1926-2016Under Investigation by firmLoss of therapy during recharging of device. Reportedly, a charging session was terminated prior to obtaining recharge threshold voltage of 3.615, which triggered rapid battery depletion state. As a result, the device depleted to 1.925…
Jan 9, 2015Class IIZ-1147-2015OtherCurrent labeling indicates that cycling improves device longevity in every programming scenario. However, enabling cycling at certain parameter settings may decrease the device battery longevity. With cycling enabled, longevity estimate…
Feb 27, 2014Class IIZ-1524-2014Labeling False and MisleadingMedtronic neurostimulators (used with pain and deep brain stimulation therapy) have a cycling feature that, when enabled, allows stimulation output to be toggled on and off based on the programmed settings. The use of this feature is…
Sep 30, 2013Class IIZ-0278-2014Software designMedtronic notified customers regarding Loss of Stimulation of specific Implantable Neuro-Stimulators under the following specific conditions: Switching between groups with Multiple programs or creating program groups. Therapy can be…
Aug 2, 2006Not classifiedZ-1536-06OtherModel 3777 Standard Octad Lead (Lot V009545) and Model 3778 Compact Octad Lead (Lot V009546) were mispackaged. The result of this error is that lead kits labeled 3777 lot V009545 may contain a 3778 compact-spaced lead, while those labeled…

No ancestor or descendant has a Class I recall.

How does its category compare?

All LGW devices

LGW Stimulator, Spinal-Cord, Totally Implanted For Pain Relief: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LGW

65in 2025

2007: 202008: 532009: 382010: 562011: 472012: 682013: 712014: 682015: 722016: 922017: 942018: 642019: 692020: 522021: 562022: 602023: 632024: 942025: 652026 (partial year): 56
2007high 94 (2017) · low 20 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LGW

Too few LGW 510(k)s in any year since 2007 to give a median.

Recalls per year · LGW

3in 2025

2007: 42008: 42009: 22010: 22011: 12012: 12013: 32016: 22017: 22020: 12021: 12022: 52023: 22024: 22025: 32026 (partial year): 2, 1 Class I
2007high 5 (2022) · low 0 (2014)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · LGW

31,054in 2025

2007: 1,0942008: 3,0182009: 5,2512010: 9,2652011: 11,1322012: 17,9152013: 22,4322014: 11,2502015: 11,6902016: 15,6432017: 18,6122018: 23,6262019: 30,0382020: 28,1062021: 24,3332022: 23,2692023: 33,5282024: 35,0742025: 31,054
2007high 35,074 (2024) · low 1,094 (2007)2025

MAUDE reports naming product code LGW, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200720—41,094
200853—43,018
200938—25,251
201056—29,265
201147—111,132
201268—117,915
201371—322,432
201468—011,250
201572—011,690
201692—215,643
201794—218,612
201864—023,626
201969—030,038
202052—128,106
202156—124,333
202260—523,269
202363—233,528
202494—235,074
202565—331,054
202656—2 (1)—
Cite this page

510(k) Report. “P840001: Itrel (R) Totally Implantable Spinal Cord Stim. Sys.” Lineage record from FDA PMA data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/P840001

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Page last updated Nov 3, 2026