510(k) Report

PMA · Approved

Recalled · Class I · 2012

Model 203 Life Pulse High Frequency Ventilator

P850064PMAApproved Jun 30, 19881,030-day reviewClass IIIRecalls 4 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Bunnell, Inc. · LSZ Ventilator, High Frequency

Approved on Jun 30, 1988 through premarket approval, which rests on the device’s own clinical evidence after a 1,030-day review. It has a Class I recall. It has been recalled 4 times, including a Class I recall.

Decided
Jun 30, 1988
Review
1,030 days
Class
Class III
Product code
LSZ
Recalls
4 (1 Class I)

PMA approvals rest on their own clinical evidence and do not claim a predicate, so this family tree starts here.

P850064 has 48 supplements since 1989: changes to the approved device, its labelling or its manufacturing, each approved separately.

PMA supplements approved per year

1989: 21990: 11992: 11993: 11994: 21995: 11997: 11998: 11999: 22002: 12010: 12013: 22014: 12015: 42016: 32017: 32018: 32019: 32020: 22021: 32022: 12023: 42024: 22025: 12026: 2
1989high 4 (2015) · low 0 (1991)2026
View FIG. 1 as a table
YearSupplements approved
19892
19901
19921
19931
19942
19951
19971
19981
19992
20021
20101
20132
20141
20154
20163
20173
20183
20193
20202
20213
20221
20234
20242
20251
20262
FIG. 1 — Supplements to P850064, 1989–2026Source: FDA PMA database

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA PMA database (decision fields and supplements), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

Does it have a predicate?

PMA decisions do not claim a predicate device.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 19, 2012Class IZ-0514-2013Component design/selectionCustomer complaints received indicate the heater wire insulation can melt, causing a short which creates sparking and smoke, in the circuit close to the humidifier cartridge. Bunnell is recalling all lots of circuits distributed between…
Oct 25, 2010Not classifiedZ-0902-2011Software designProduct software coding error may compromise patient safety.
Oct 25, 2010Not classifiedZ-0901-2011Software designProduct software coding error may compromise patient safety.
Apr 4, 2005Not classifiedZ-0893-06OtherPotential loss of air flow or change in cycling of the ventilator. Valve Drive Regulator Board may not function properly at elevated temperatures. The increase in temperature could cause a reduction of power, causing the solenoid in the…

No ancestor or descendant has a Class I recall.

How does its category compare?

All LSZ devices

LSZ Ventilator, High Frequency: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LSZ

2in 2025

2007: 12010: 12012: 12013: 32014: 22015: 42016: 32017: 32018: 32019: 32020: 22021: 32022: 22023: 42024: 42025: 22026 (partial year): 2
2007high 4 (2015) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LSZ

Too few LSZ 510(k)s in any year since 2007 to give a median.

Recalls per year · LSZ

2in 2023

2010: 22012: 1, 1 Class I2022: 12023: 2
2007high 2 (2010) · low 0 (2007)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · LSZ

72in 2025

2007: 352008: 582009: 352010: 352011: 472012: 472013: 342014: 292015: 1952016: 3942017: 1402018: 882019: 882020: 1542021: 2242022: 2202023: 1862024: 1102025: 72
2007high 394 (2016) · low 29 (2014)2025

MAUDE reports naming product code LSZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—035
20080—058
20090—035
20101—235
20110—047
20121—1 (1)47
20133—034
20142—029
20154—0195
20163—0394
20173—0140
20183—088
20193—088
20202—0154
20213—0224
20222—1220
20234—2186
20244—0110
20252—072
20262—0—
Cite this page

510(k) Report. “P850064: Model 203 Life Pulse High Frequency Ventilator.” Lineage record from FDA PMA data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/P850064

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Page last updated Nov 3, 2026