510(k) Report

PMA · Approved

Recalled · Class I · 2026

Medtronic (R) Synchromed (TM) Pump & Infusion System

P860004PMAApproved Mar 14, 1988770-day reviewClass IIIRecalls 40 (2 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medtronic, Inc. · LKK Pump, Infusion, Implanted, Programmable

Approved on Mar 14, 1988 through premarket approval, which rests on the device’s own clinical evidence after a 770-day review. It has a Class I recall. It has been recalled 40 times, including 2 Class I recalls.

Decided
Mar 14, 1988
Review
770 days
Class
Class III
Product code
LKK
Recalls
40 (2 Class I)

PMA approvals rest on their own clinical evidence and do not claim a predicate, so this family tree starts here.

P860004 has 442 supplements since 1988: changes to the approved device, its labelling or its manufacturing, each approved separately.

PMA supplements approved per year

1988: 21989: 61990: 31991: 51992: 21993: 11994: 71995: 21996: 21997: 21998: 31999: 32000: 12001: 32002: 62003: 22004: 52005: 132006: 92007: 42008: 92009: 152010: 202011: 232012: 162013: 192014: 152015: 282016: 202017: 262018: 292019: 252020: 212021: 162022: 212023: 182024: 142025: 192026: 7
1988high 29 (2018) · low 1 (1993)2026
View FIG. 1 as a table
YearSupplements approved
19882
19896
19903
19915
19922
19931
19947
19952
19962
19972
19983
19993
20001
20013
20026
20032
20045
200513
20069
20074
20089
200915
201020
201123
201216
201319
201415
201528
201620
201726
201829
201925
202021
202116
202221
202318
202414
202519
20267
FIG. 1 — Supplements to P860004, 1988–2026Source: FDA PMA database

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA PMA database (decision fields and supplements), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

Does it have a predicate?

PMA decisions do not claim a predicate device.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 8, 2026Class IIZ-2201-2026Under Investigation by firmSoftware issue: When programming the SynchroMed Pump with the A810 CP App in Flex Infusion Mode, the infusion steps should automatically be arranged in consecutive order, from earliest to latest. However, in rare cases, the steps may be…
Jan 9, 2025Class IIZ-1506-2025Under Investigation by firmProduct complaints were received describing the A820 myPTM app taking longer than expected for patients to interact with their implantable pump.
May 10, 2024Class IIZ-2352-2024Device DesignDesign update to the Ascenda Intrathecal Catheter Models 8780, 8781, and 8784. The intent of the design update is to reduce the potential for tissue growth into the Ascenda catheter connector which may potentially lead to catheter…
Nov 13, 2023Class IIZ-0615-2024Under Investigation by firmMedtronic recently identified that if the SynchroMed II pump switches into telemetry mode due to electromagnetic interference (EMI) from an MRI scan, while the pump is sounding an alarm, the pump will not resume drug delivery after…
Nov 13, 2023Class IIZ-0614-2024Under Investigation by firmMedtronic recently identified that if the SynchroMed II pump switches into telemetry mode due to electromagnetic interference (EMI) from an MRI scan, while the pump is sounding an alarm, the pump will not resume drug delivery after…
Jul 13, 2020Class IIZ-0522-2024Software Design ChangeThe previous software application version (1.1.300) is missing a decimal separator (a comma) for parameter range guidance values displayed on some of the programming screens: Catheter, Reservoir, Infusion, Bolus, myPTM, and Alarm.
Oct 11, 2019Class IZ-0508-2020Process controlThere is a potential for the presence of a foreign particle inside the pump motor assembly which could interfere with the motor gear rotation and lead to a permanent motor stall.
Sep 21, 2017Class IIIZ-0537-2018OtherThere is a possibility that specific lots of catheters that have recently been distributed may be at risk for an increased potential for kinking at the proximal end where the catheter connects to the drug infusion pump.
Aug 9, 2017Class IIZ-0896-2018Device DesignMedtronic implemented a final design change intended to address the motor stall due to corrosion and is recalling devices with the previous design that are not implanted.
Oct 3, 2016Class IZ-0788-2017Under Investigation by firmMedtronic is following up to a May 2013 communication regarding the Priming Bolus function for the SynchroMed Infusion System. Medtronic is updating the Model 8870 software application card (to version AAU01) and the SynchroMed pump…
Feb 9, 2016Class IIZ-1271-2016Under Investigation by firmMedtronic Neuromodulation is initiating a voluntary removal of a single Model 8781 Ascenda" Intrathecal Catheter kit. We have identified that this kit may have been produced with incorrect outer package labels. As a result, it may not…
Apr 10, 2015Class IIZ-1681-2015Process controlMedtronic is conducting a recall of a specific subset of Model 8637-20 and 8637-40 SynchroMed II implantable drug pumps because the audible alarm could cease to function.
Jan 14, 2011Not classifiedZ-1061-2011OtherThis letter provides important reminders concerning the potential for a pocket fill during a SynchroMed II or SynchroMed EL implantable drug pump refill procedure, and important patient management recommendations that will be added to our…
Jan 14, 2011Not classifiedZ-1062-2011OtherThis letter provides important reminders concerning the potential for a pocket fill during a SynchroMed II or SynchroMed EL implantable drug pump refill procedure, and important patient management recommendations that will be added to our…
Jan 14, 2011Not classifiedZ-1060-2011OtherThis letter provides important reminders concerning the potential for a pocket fill during a SynchroMed II or SynchroMed EL implantable drug pump refill procedure, and important patient management recommendations that will be added to our…
Jul 1, 2009Not classifiedZ-2276-2009Nonconforming Material/ComponentThis recall is related to the potential for reduced battery performance in a small percentage of Medtronic Model 8637 SynchroMed II pumps with batteries manufactured during two distinct time periods prior to April 2005. Nature of the…
Mar 23, 2009Not classifiedZ-1214-2010OtherRoutine monitoring of Medtronic's vigilance program identified user related programming errors. Medtronic has received reports of both overdose and underdose that were the result of programming use of SynchroMed implantable infusion…
Apr 14, 2008Not classifiedZ-2174-2008Device DesignThe SC catheters cannot completely engage with the Model 700-04M portal connector. The result of this condition is a possible leakage or disconnect of the SC catheter from the portal during the continuous therapy trial period.
Apr 14, 2008Not classifiedZ-2171-2008Device DesignThe SC catheters cannot completely engage with the Model 700-04M portal connector. The result of this condition is a possible leakage or disconnect of the SC catheter from the portal during the continuous therapy trial period.
Apr 14, 2008Not classifiedZ-2173-2008Device DesignThe SC catheters cannot completely engage with the Model 700-04M portal connector. The result of this condition is a possible leakage or disconnect of the SC catheter from the portal during the continuous therapy trial period.
Apr 14, 2008Not classifiedZ-2172-2008Device DesignThe SC catheters cannot completely engage with the Model 700-04M portal connector. The result of this condition is a possible leakage or disconnect of the SC catheter from the portal during the continuous therapy trial period.
Jan 16, 2008Not classifiedZ-1153-2008Device DesignDevice/Drug Interaction - The company updated the labeling for the devices to include current patient management and treatment recommendations. The company received reports of inflammatory mass formations at or near the distal tip of…
Jan 16, 2008Not classifiedZ-1146-2008Device DesignDevice/Drug Interaction - The company updated the labeling for the devices to include current patient management and treatment recommendations. The company received reports of inflammatory mass formations at or near the distal tip of…
Jan 16, 2008Not classifiedZ-1149-2008Device DesignDevice/Drug Interaction - The company updated the labeling for the devices to include current patient management and treatment recommendations. The company received reports of inflammatory mass formations at or near the distal tip of…
Jan 16, 2008Not classifiedZ-1147-2008Device DesignDevice/Drug Interaction - The company updated the labeling for the devices to include current patient management and treatment recommendations. The company received reports of inflammatory mass formations at or near the distal tip of…
Jan 16, 2008Not classifiedZ-1144-2008Device DesignDevice/Drug Interaction - The company updated the labeling for the devices to include current patient management and treatment recommendations. The company received reports of inflammatory mass formations at or near the distal tip of…
Jan 16, 2008Not classifiedZ-1148-2008Device DesignDevice/Drug Interaction - The company updated the labeling for the devices to include current patient management and treatment recommendations. The company received reports of inflammatory mass formations at or near the distal tip of…
Jan 16, 2008Not classifiedZ-1151-2008Device DesignDevice/Drug Interaction - The company updated the labeling for the devices to include current patient management and treatment recommendations. The company received reports of inflammatory mass formations at or near the distal tip of…
Jan 16, 2008Not classifiedZ-1150-2008Device DesignDevice/Drug Interaction - The company updated the labeling for the devices to include current patient management and treatment recommendations. The company received reports of inflammatory mass formations at or near the distal tip of…
Jan 16, 2008Not classifiedZ-1142-2008Device DesignDevice/Drug Interaction - The company updated the labeling for the devices to include current patient management and treatment recommendations. The company received reports of inflammatory mass formations at or near the distal tip of…
Jan 16, 2008Not classifiedZ-1152-2008Device DesignDevice/Drug Interaction - The company updated the labeling for the devices to include current patient management and treatment recommendations. The company received reports of inflammatory mass formations at or near the distal tip of…
Jan 16, 2008Not classifiedZ-1145-2008Device DesignDevice/Drug Interaction - The company updated the labeling for the devices to include current patient management and treatment recommendations. The company received reports of inflammatory mass formations at or near the distal tip of…
Jan 16, 2008Not classifiedZ-1143-2008Device DesignDevice/Drug Interaction - The company updated the labeling for the devices to include current patient management and treatment recommendations. The company received reports of inflammatory mass formations at or near the distal tip of…
Jan 16, 2008Not classifiedZ-1154-2008Device DesignDevice/Drug Interaction - The company updated the labeling for the devices to include current patient management and treatment recommendations. The company received reports of inflammatory mass formations at or near the distal tip of…
Aug 3, 2007Not classifiedZ-0739-2008Device DesignPump motor stall due to gear shaft wear
Jul 21, 2006Not classifiedZ-1415-06OtherTip dislodgement during implantation-Medtronic is recalling Model 8731 Intrathecal Catheter and Model 8598 Intrathecal Catheter Distal Revision Kit because the platinum-iridium tip may be dislodged by the guide wire during implantation.
Jul 21, 2006Not classifiedZ-1414-06OtherTip dislodgement during implantation-Medtronic is recalling Model 8731 Intrathecal Catheter and Model 8598 Intrathecal Catheter Distal Revision Kit because the platinum-iridium tip may be dislodged by the guide wire during implantation.
Nov 19, 2004Not classifiedZ-0391-05OtherOuter package of Lioresal Intrethecal Refill Kit, model 8561, identifies the kit to contain 1-20 ml ampule of baclofen injection for intrathecal administration, 500 mcg/ml (10 mg/20 ml). Lot N0016487 incorrectly contains 1-20 ml ampule of…
Aug 24, 2004Not classifiedZ-1334-04OtherWith the Model 8870 software application card versions AAA 02, BBB 04, BBC 02, and BBD 01, which is used in conjunction with the Model 8840 N'Vision Clinician Programmer, users may have mistakenly entered a periodic bolus interval into…
Jul 23, 2004Not classifiedZ-1384-04OtherThe sterile tray label for the Model 8540 Catheter Access Port kit may be incorrectly labeled with a Model 8551 Refill kit label. The outer box of the affected Model 8540 Catheter Access Port kit is labeled correctly and the component…

No ancestor or descendant has a Class I recall.

How does its category compare?

All LKK devices

LKK Pump, Infusion, Implanted, Programmable: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LKK

22in 2025

2007: 82008: 82009: 102010: 292011: 342012: 332013: 332014: 282015: 402016: 302017: 422018: 552019: 572020: 492021: 322022: 322023: 192024: 222025: 222026 (partial year): 9
2007high 57 (2019) · low 8 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LKK

Too few LKK 510(k)s in any year since 2007 to give a median.

Recalls per year · LKK

5in 2024

2007: 162008: 352009: 62010: 22011: 42012: 3, 2 Class I2016: 1, 1 Class I2017: 22019: 1, 1 Class I2020: 12023: 22024: 52026 (partial year): 2
2007high 35 (2008) · low 0 (2013)2026

Class I (4 since 2007) Class II/III not classified

Adverse event reports per year · LKK

5,732in 2025

2007: 2,6762008: 3,3982009: 3,2252010: 4,0672011: 3,9132012: 5,5572013: 8,0552014: 7,4582015: 8,4512016: 8,4612017: 7,7512018: 6,9362019: 5,6522020: 5,1152021: 4,7132022: 4,3102023: 4,0412024: 4,6182025: 5,732
2007high 8,461 (2016) · low 2,676 (2007)2025

MAUDE reports naming product code LKK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20078—162,676
20088—353,398
200910—63,225
201029—24,067
201134—43,913
201233—3 (2)5,557
201333—08,055
201428—07,458
201540—08,451
201630—1 (1)8,461
201742—27,751
201855—06,936
201957—1 (1)5,652
202049—15,115
202132—04,713
202232—04,310
202319—24,041
202422—54,618
202522—05,732
20269—2—
Cite this page

510(k) Report. “P860004: Medtronic (R) Synchromed (TM) Pump & Infusion System.” Lineage record from FDA PMA data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/P860004

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Page last updated Nov 3, 2026