Cadence (R) Tiered Therapy Defibrillation System

P910023PMAApproved Apr 30, 1993738-day reviewClass IIIRecalls 40

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Abbott Medical · LWS Implantable Cardioverter Defibrillator (Non-Crt)

Approved on Apr 30, 1993 through premarket approval, which rests on the device’s own clinical evidence after a 738-day review. 2 later devices descend from it. It has been recalled 40 times.

Decided
Apr 30, 1993
Review
738 days
Class
Class III
Product code
LWS
Recalls
40

2 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2010: 2 descendants cleared1990200020102020
  1. This deviceP910023 Cadence (R) Tiered Therapy Defibrillation SystemAbbott Medical · 1993
  2. Descendant · cites it directly · 1 of its ownK093858 Psa Cable, Models 4160 and 4161 and Psa Cable Adapter, Model 4053ASt. Jude Medical, Cardiac Rhythm Management Divisi · 2010
  3. Descendant · 2 generations on · 0 of its ownK101982 Merlin Psa Patient Cable EX 3150, Cable Adapter EX3170, M Adapter…St Jude Medical, Crmd · 2010

Most influential descendants

Showing 2 of 2 · See all
  • K093858Psa Cable, Models 4160 and 4161 and Psa Cable Adapter, Model 4053ASt. Jude Medical, Cardiac Rhythm Management Divisi · 2010 · cites it directly1descendant
  • K101982Merlin Psa Patient Cable EX 3150, Cable Adapter EX3170, M Adapter EX3180St Jude Medical, Crmd · 2010 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)P910023Cadence (R) Tiered Therapy Defibrillation SystemAbbott Medical1993—
+1K093858Psa Cable, Models 4160 and 4161 and Psa Cable Adapter, Model 4053ASt. Jude Medical, Cardiac Rhythm Management Divisi2010—
+2K101982Merlin Psa Patient Cable EX 3150, Cable Adapter EX3170, M Adapter EX3180St Jude Medical, Crmd2010—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201020
FIG. 1 — Lineage of P910023, 1993–2010Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA PMA database (decision fields and supplements), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LWS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

Does it have a predicate?

PMA decisions do not claim a predicate device.

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K093858Psa Cable, Models 4160 and 4161 and Psa Cable Adapter, Model 4053ASt. Jude Medical, Cardiac Rhythm Management Divisi20101

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 16, 2018Class IIZ-2320-2018Software designThe firm is releasing new firmware for high voltage implantable cardiac devices (ICD) and cardiac resynchronization therapy defibrillators (CRT-D) intended to strengthen security by providing an additional layer of protection against…
Apr 16, 2018Class IIZ-2325-2018Software designThe firm is releasing new firmware for high voltage implantable cardiac devices (ICD) and cardiac resynchronization therapy defibrillators (CRT-D) intended to strengthen security by providing an additional layer of protection against…
Apr 16, 2018Class IIZ-2321-2018Software designThe firm is releasing new firmware for high voltage implantable cardiac devices (ICD) and cardiac resynchronization therapy defibrillators (CRT-D) intended to strengthen security by providing an additional layer of protection against…
Apr 16, 2018Class IIZ-2318-2018Software designThe firm is releasing new firmware for high voltage implantable cardiac devices (ICD) and cardiac resynchronization therapy defibrillators (CRT-D) intended to strengthen security by providing an additional layer of protection against…
Apr 16, 2018Class IIZ-2319-2018Software designThe firm is releasing new firmware for high voltage implantable cardiac devices (ICD) and cardiac resynchronization therapy defibrillators (CRT-D) intended to strengthen security by providing an additional layer of protection against…
Mar 31, 2016Class IIZ-1632-2016OtherSt. Jude Medical is recalling the Ellipse ICD (Implantable Cardioverter Defibrillator) due to the potential inability to deliver high voltage therapy.
Aug 19, 2014Class IIZ-2572-2014Component design/selectionAn anomaly may result in delayed, partial, or no delivery of high voltage therapy. Internal damage to the high voltage capacitors, which may result in extended charge time, occurring during capacitor maintenance or charging for high…
Jan 16, 2008Not classifiedZ-1327-2008Software designA condition that could lead to a ventricular sensing anomaly in Epic and Atlas family of lmplantable Cardioverter Defibrillators (ICDs) has been identified. A loss of ventricular sensing would prevent the ICD from being able to detect an…
Jan 16, 2008Not classifiedZ-1341-2008Software designA condition that could lead to a ventricular sensing anomaly in Epic and Atlas family of lmplantable Cardioverter Defibrillators (ICDs) has been identified. A loss of ventricular sensing would prevent the ICD from being able to detect an…
Jan 16, 2008Not classifiedZ-1336-2008Software designA condition that could lead to a ventricular sensing anomaly in Epic and Atlas family of lmplantable Cardioverter Defibrillators (ICDs) has been identified. A loss of ventricular sensing would prevent the ICD from being able to detect an…
Jan 16, 2008Not classifiedZ-1347-2008Software designA condition that could lead to a ventricular sensing anomaly in Epic and Atlas family of lmplantable Cardioverter Defibrillators (ICDs) has been identified. A loss of ventricular sensing would prevent the ICD from being able to detect an…
Jan 16, 2008Not classifiedZ-1340-2008Software designA condition that could lead to a ventricular sensing anomaly in Epic and Atlas family of lmplantable Cardioverter Defibrillators (ICDs) has been identified. A loss of ventricular sensing would prevent the ICD from being able to detect an…
Jan 16, 2008Not classifiedZ-1335-2008Software designA condition that could lead to a ventricular sensing anomaly in Epic and Atlas family of lmplantable Cardioverter Defibrillators (ICDs) has been identified. A loss of ventricular sensing would prevent the ICD from being able to detect an…
Jan 16, 2008Not classifiedZ-1333-2008Software designA condition that could lead to a ventricular sensing anomaly in Epic and Atlas family of lmplantable Cardioverter Defibrillators (ICDs) has been identified. A loss of ventricular sensing would prevent the ICD from being able to detect an…
Jan 16, 2008Not classifiedZ-1323-2008Software designA condition that could lead to a ventricular sensing anomaly in Epic and Atlas family of lmplantable Cardioverter Defibrillators (ICDs) has been identified. A loss of ventricular sensing would prevent the ICD from being able to detect an…
Jan 16, 2008Not classifiedZ-1325-2008Software designA condition that could lead to a ventricular sensing anomaly in Epic and Atlas family of lmplantable Cardioverter Defibrillators (ICDs) has been identified. A loss of ventricular sensing would prevent the ICD from being able to detect an…
Jan 16, 2008Not classifiedZ-1339-2008Software designA condition that could lead to a ventricular sensing anomaly in Epic and Atlas family of lmplantable Cardioverter Defibrillators (ICDs) has been identified. A loss of ventricular sensing would prevent the ICD from being able to detect an…
Jan 16, 2008Not classifiedZ-1346-2008Software designA condition that could lead to a ventricular sensing anomaly in Epic and Atlas family of lmplantable Cardioverter Defibrillators (ICDs) has been identified. A loss of ventricular sensing would prevent the ICD from being able to detect an…
Jan 16, 2008Not classifiedZ-1324-2008Software designA condition that could lead to a ventricular sensing anomaly in Epic and Atlas family of lmplantable Cardioverter Defibrillators (ICDs) has been identified. A loss of ventricular sensing would prevent the ICD from being able to detect an…
Jan 16, 2008Not classifiedZ-1344-2008Software designA condition that could lead to a ventricular sensing anomaly in Epic and Atlas family of lmplantable Cardioverter Defibrillators (ICDs) has been identified. A loss of ventricular sensing would prevent the ICD from being able to detect an…
Jan 16, 2008Not classifiedZ-1330-2008Software designA condition that could lead to a ventricular sensing anomaly in Epic and Atlas family of lmplantable Cardioverter Defibrillators (ICDs) has been identified. A loss of ventricular sensing would prevent the ICD from being able to detect an…
Jan 16, 2008Not classifiedZ-1326-2008Software designA condition that could lead to a ventricular sensing anomaly in Epic and Atlas family of lmplantable Cardioverter Defibrillators (ICDs) has been identified. A loss of ventricular sensing would prevent the ICD from being able to detect an…
Jan 16, 2008Not classifiedZ-1345-2008Software designA condition that could lead to a ventricular sensing anomaly in Epic and Atlas family of lmplantable Cardioverter Defibrillators (ICDs) has been identified. A loss of ventricular sensing would prevent the ICD from being able to detect an…
Jan 16, 2008Not classifiedZ-1331-2008Software designA condition that could lead to a ventricular sensing anomaly in Epic and Atlas family of lmplantable Cardioverter Defibrillators (ICDs) has been identified. A loss of ventricular sensing would prevent the ICD from being able to detect an…
Jan 16, 2008Not classifiedZ-1328-2008Software designA condition that could lead to a ventricular sensing anomaly in Epic and Atlas family of lmplantable Cardioverter Defibrillators (ICDs) has been identified. A loss of ventricular sensing would prevent the ICD from being able to detect an…
Jan 16, 2008Not classifiedZ-1337-2008Software designA condition that could lead to a ventricular sensing anomaly in Epic and Atlas family of lmplantable Cardioverter Defibrillators (ICDs) has been identified. A loss of ventricular sensing would prevent the ICD from being able to detect an…
Jan 16, 2008Not classifiedZ-1329-2008Software designA condition that could lead to a ventricular sensing anomaly in Epic and Atlas family of lmplantable Cardioverter Defibrillators (ICDs) has been identified. A loss of ventricular sensing would prevent the ICD from being able to detect an…
Jan 16, 2008Not classifiedZ-1343-2008Software designA condition that could lead to a ventricular sensing anomaly in Epic and Atlas family of lmplantable Cardioverter Defibrillators (ICDs) has been identified. A loss of ventricular sensing would prevent the ICD from being able to detect an…
Jan 16, 2008Not classifiedZ-1321-2008Software designA condition that could lead to a ventricular sensing anomaly in Epic and Atlas family of lmplantable Cardioverter Defibrillators (ICDs) has been identified. A loss of ventricular sensing would prevent the ICD from being able to detect an…
Jan 16, 2008Not classifiedZ-1338-2008Software designA condition that could lead to a ventricular sensing anomaly in Epic and Atlas family of lmplantable Cardioverter Defibrillators (ICDs) has been identified. A loss of ventricular sensing would prevent the ICD from being able to detect an…
Jan 16, 2008Not classifiedZ-1334-2008Software designA condition that could lead to a ventricular sensing anomaly in Epic and Atlas family of lmplantable Cardioverter Defibrillators (ICDs) has been identified. A loss of ventricular sensing would prevent the ICD from being able to detect an…
Jan 16, 2008Not classifiedZ-1320-2008Software designA condition that could lead to a ventricular sensing anomaly in Epic and Atlas family of lmplantable Cardioverter Defibrillators (ICDs) has been identified. A loss of ventricular sensing would prevent the ICD from being able to detect an…
Jan 16, 2008Not classifiedZ-1322-2008Software designA condition that could lead to a ventricular sensing anomaly in Epic and Atlas family of lmplantable Cardioverter Defibrillators (ICDs) has been identified. A loss of ventricular sensing would prevent the ICD from being able to detect an…
Jan 16, 2008Not classifiedZ-1332-2008Software designA condition that could lead to a ventricular sensing anomaly in Epic and Atlas family of lmplantable Cardioverter Defibrillators (ICDs) has been identified. A loss of ventricular sensing would prevent the ICD from being able to detect an…
Jan 16, 2008Not classifiedZ-1342-2008Software designA condition that could lead to a ventricular sensing anomaly in Epic and Atlas family of lmplantable Cardioverter Defibrillators (ICDs) has been identified. A loss of ventricular sensing would prevent the ICD from being able to detect an…
Oct 6, 2005Not classifiedZ-0149-06OtherVendor-supplied memory chip in a limited number of St. Jude Medical ICDs can trigger a temporary loss of pacing function and permanent los of defibrillation support.
Oct 6, 2005Not classifiedZ-0146-06OtherVendor-supplied memory chip in a limited number of St. Jude Medical ICDs can trigger a temporary loss of pacing function and permanent los of defibrillation support.
Oct 6, 2005Not classifiedZ-0150-06OtherVendor-supplied memory chip in a limited number of St. Jude Medical ICDs can trigger a temporary loss of pacing function and permanent los of defibrillation support.
Oct 6, 2005Not classifiedZ-0147-06OtherVendor-supplied memory chip in a limited number of St. Jude Medical ICDs can trigger a temporary loss of pacing function and permanent los of defibrillation support.
Mar 10, 2005Not classifiedZ-0709-05OtherSensitivity of a magnetic switch was not being set correctly. This could result in delivery of an unneeded shock during surgery with use of electrocautery.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K093858 (St. Jude Medical, Cardiac Rhythm Management Divisi, 2010) has gone furthest, with 1 descendant of its own.

How does its category compare?

All LWS devices

LWS Implantable Cardioverter Defibrillator (Non-Crt): FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LWS

97in 2025

2007: 702008: 942009: 802010: 1192011: 1362012: 1292013: 1482014: 1682015: 1002016: 1452017: 1622018: 1422019: 982020: 1252021: 1112022: 912023: 1012024: 1082025: 972026 (partial year): 93
2007high 168 (2014) · low 70 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LWS

Too few LWS 510(k)s in any year since 2007 to give a median.

Recalls per year · LWS

2in 2025

2007: 42008: 32009: 82010: 62011: 82012: 52013: 12015: 2, 1 Class I2016: 2, 2 Class I2017: 7, 5 Class I2018: 12020: 6, 3 Class I2022: 32023: 22025: 2, 1 Class I
2007high 8 (2009) · low 0 (2014)2026

Class I (12 since 2007) Class II/III not classified

Adverse event reports per year · LWS

24,303in 2025

2007: 10,3302008: 11,3272009: 13,7682010: 17,3212011: 15,0972012: 12,8202013: 17,2422014: 18,5942015: 14,0402016: 16,6382017: 18,4542018: 17,2282019: 16,1842020: 16,9092021: 19,6542022: 23,4542023: 28,7012024: 25,2022025: 24,303
2007high 28,701 (2023) · low 10,330 (2007)2025

MAUDE reports naming product code LWS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200770—410,330
200894—311,327
200980—813,768
2010119—617,321
2011136—815,097
2012129—512,820
2013148—117,242
2014168—018,594
2015100—2 (1)14,040
2016145—2 (2)16,638
2017162—7 (5)18,454
2018142—117,228
201998—016,184
2020125—6 (3)16,909
2021111—019,654
202291—323,454
2023101—228,701
2024108—025,202
202597—2 (1)24,303
202693—0—
Cite this page

510(k) Report. “P910023: Cadence (R) Tiered Therapy Defibrillation System.” Lineage record from FDA PMA data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/P910023

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026