510(k) Report

PMA · Approved

Recalled · Class I · 2015

TVL (TM) Lead System

P950022PMAApproved May 10, 1996315-day reviewClass IIIRecalls 3 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Abbott Medical · LWS Implantable Cardioverter Defibrillator (Non-Crt)

Approved on May 10, 1996 through premarket approval, which rests on the device’s own clinical evidence after a 315-day review. 1 of the 6 devices in its family, counting itself, has a Class I recall. 5 later devices descend from it. It has been recalled 3 times, including a Class I recall.

Decided
May 10, 1996
Review
315 days
Class
Class III
Product code
LWS
Recalls
3 (1 Class I)

5 later devices descend from it, 3 citing it directly. 1 of the 6 devices in this family, counting this one, has a Class I recall.

Descendants, by year cleared 2010: 2 descendants cleared2008: 1 descendant cleared2009: 1 descendant cleared2019: 1 descendant cleared1995200020052010201520202025
  1. This deviceP950022 TVL (TM) Lead SystemAbbott Medical · 1996
  2. Descendant · cites it directly · 1 of its ownK093858 Psa Cable, Models 4160 and 4161 and Psa Cable Adapter, Model 4053ASt. Jude Medical, Cardiac Rhythm Management Divisi · 2010
  3. Descendant · cites it directly · 1 of its ownK090163 Stylet, Models 4060, 4062, 4064, 4090, 4091 and 4078St Jude Medical · 2009
  4. Descendant · 2 generations on · 0 of its ownK191469 Biotronik StyletsBiotronik, Inc. · 2019
  5. Descendant · 2 generations on · 0 of its ownK101982 Merlin Psa Patient Cable EX 3150, Cable Adapter EX3170, M Adapter…St Jude Medical, Crmd · 2010
  6. Descendant · cites it directly · 0 of its ownK080924 Mond Rvot Stylet Models 4140, 4141, 4150, 4151St. Jude Medical, Cardiac Rhythm Management Divisi · 2008

Most influential descendants

Showing 5 of 5 · See all
  • K093858Psa Cable, Models 4160 and 4161 and Psa Cable Adapter, Model 4053ASt. Jude Medical, Cardiac Rhythm Management Divisi · 2010 · cites it directly1descendant
  • K090163Stylet, Models 4060, 4062, 4064, 4090, 4091 and 4078St Jude Medical · 2009 · cites it directly1descendant
  • K191469Biotronik StyletsBiotronik, Inc. · 2019 · 2 generations on0descendants
  • K101982Merlin Psa Patient Cable EX 3150, Cable Adapter EX3170, M Adapter EX3180St Jude Medical, Crmd · 2010 · 2 generations on0descendants
  • K080924Mond Rvot Stylet Models 4140, 4141, 4150, 4151St. Jude Medical, Cardiac Rhythm Management Divisi · 2008 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)P950022TVL (TM) Lead SystemAbbott Medical1996—
+1K093858Psa Cable, Models 4160 and 4161 and Psa Cable Adapter, Model 4053ASt. Jude Medical, Cardiac Rhythm Management Divisi2010—
+1K090163Stylet, Models 4060, 4062, 4064, 4090, 4091 and 4078St Jude Medical2009—
+2K191469Biotronik StyletsBiotronik, Inc.2019—
+2K101982Merlin Psa Patient Cable EX 3150, Cable Adapter EX3170, M Adapter EX3180St Jude Medical, Crmd2010—
+1K080924Mond Rvot Stylet Models 4140, 4141, 4150, 4151St. Jude Medical, Cardiac Rhythm Management Divisi2008—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200810
200910
201020
201910
FIG. 1 — Lineage of P950022, 1996–2019Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA PMA database (decision fields and supplements), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

Does it have a predicate?

PMA decisions do not claim a predicate device.

What descends from it? (5)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K093858Psa Cable, Models 4160 and 4161 and Psa Cable Adapter, Model 4053ASt. Jude Medical, Cardiac Rhythm Management Divisi20101
K090163Stylet, Models 4060, 4062, 4064, 4090, 4091 and 4078St Jude Medical20091
K080924Mond Rvot Stylet Models 4140, 4141, 4150, 4151St. Jude Medical, Cardiac Rhythm Management Divisi20080

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 3, 2015Class IZ-0660-2016Process controlLeads may have been compromised during the manufacturing process, with cuts in the insulation layer of the SVC shock coil.
Nov 28, 2011Not classifiedZ-0457-2012Device DesignSt. Jude Medical has confirmed the failures associated with all cause insulation failure on the St. Jude Medical Riata and Riata ST Silicone Endocardial Defibrillation Leads with specific emphasis on externalized conductors.
Nov 28, 2011Not classifiedZ-0458-2012Device DesignThe recall was initiated because St. Jude Medical has confirmed the failures associated with all cause insulation failure on the St. Jude Medical Riata and Riata ST Silicone Endocardial Defibrillation Leads with specific emphasis on…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K093858 (St. Jude Medical, Cardiac Rhythm Management Divisi, 2010) has gone furthest, with 1 descendant of its own.

How does its category compare?

All LWS devices

LWS Implantable Cardioverter Defibrillator (Non-Crt): FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LWS

97in 2025

2007: 702008: 942009: 802010: 1192011: 1362012: 1292013: 1482014: 1682015: 1002016: 1452017: 1622018: 1422019: 982020: 1252021: 1112022: 912023: 1012024: 1082025: 972026 (partial year): 93
2007high 168 (2014) · low 70 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LWS

Too few LWS 510(k)s in any year since 2007 to give a median.

Recalls per year · LWS

2in 2025

2007: 42008: 32009: 82010: 62011: 82012: 52013: 12015: 2, 1 Class I2016: 2, 2 Class I2017: 7, 5 Class I2018: 12020: 6, 3 Class I2022: 32023: 22025: 2, 1 Class I
2007high 8 (2009) · low 0 (2014)2026

Class I (12 since 2007) Class II/III not classified

Adverse event reports per year · LWS

24,303in 2025

2007: 10,3302008: 11,3272009: 13,7682010: 17,3212011: 15,0972012: 12,8202013: 17,2422014: 18,5942015: 14,0402016: 16,6382017: 18,4542018: 17,2282019: 16,1842020: 16,9092021: 19,6542022: 23,4542023: 28,7012024: 25,2022025: 24,303
2007high 28,701 (2023) · low 10,330 (2007)2025

MAUDE reports naming product code LWS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200770—410,330
200894—311,327
200980—813,768
2010119—617,321
2011136—815,097
2012129—512,820
2013148—117,242
2014168—018,594
2015100—2 (1)14,040
2016145—2 (2)16,638
2017162—7 (5)18,454
2018142—117,228
201998—016,184
2020125—6 (3)16,909
2021111—019,654
202291—323,454
2023101—228,701
2024108—025,202
202597—2 (1)24,303
202693—0—
Cite this page

510(k) Report. “P950022: TVL (TM) Lead System.” Lineage record from FDA PMA data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/P950022

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026