510(k) Report

PMA · Approved

Recalled · Class II · 2022

Ventak AV Aicd Ventak Prizm DR/VR, Vitality, Cofient, and Teligen Automatic Implantable Cardioveter Defibrillator System

P960040PMAApproved Jul 18, 1997332-day reviewClass IIIRecalls 14

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Boston Scientific · LWS Implantable Cardioverter Defibrillator (Non-Crt)

Approved on Jul 18, 1997 through premarket approval, which rests on the device’s own clinical evidence after a 332-day review. It has been recalled 14 times.

Decided
Jul 18, 1997
Review
332 days
Class
Class III
Product code
LWS
Recalls
14

PMA approvals rest on their own clinical evidence and do not claim a predicate, so this family tree starts here.

P960040 has 485 supplements since 1997: changes to the approved device, its labelling or its manufacturing, each approved separately.

PMA supplements approved per year

1997: 31998: 51999: 22000: 62001: 62002: 22003: 112004: 122005: 312006: 372007: 192008: 282009: 262010: 242011: 92012: 372013: 242014: 292015: 212016: 232017: 292018: 182019: 152020: 142021: 132022: 152023: 92024: 52025: 52026: 7
1997high 37 (2006) · low 2 (1999)2026
View FIG. 1 as a table
YearSupplements approved
19973
19985
19992
20006
20016
20022
200311
200412
200531
200637
200719
200828
200926
201024
20119
201237
201324
201429
201521
201623
201729
201818
201915
202014
202113
202215
20239
20245
20255
20267
FIG. 1 — Supplements to P960040, 1997–2026Source: FDA PMA database

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA PMA database (decision fields and supplements), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LWS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

Does it have a predicate?

PMA decisions do not claim a predicate device.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 7, 2022Class IIZ-1670-2022Under Investigation by firmThere is potential for unanticipated software anomalies to occur when using the Model 2868 Application loaded on the 3120 ZOOM LATITUTDE programmer when communicating with certain Boston Scientific transvenous defibrillators.
Oct 4, 2017Class IIZ-0077-2018Under Investigation by firmThe devices have an incorrect firmware configuration.
Oct 4, 2017Class IIZ-0079-2018Under Investigation by firmThe devices have an incorrect firmware configuration.
Oct 4, 2017Class IIZ-0078-2018Under Investigation by firmThe devices have an incorrect firmware configuration.
Mar 15, 2010Not classifiedZ-1555-2010PMABoston Scientific Corporation has stopped shipment and is retrieving unimplanted devices from US hospital shelves of all its implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds)…
Dec 1, 2009Not classifiedZ-0610-2010Device DesignBSC CRM provided important patient management information regarding Boston Scientific COGNIS CRT-Ds and TELIGEN ICDs. Engineering simulations and field reports indicate that in some cases, implanted devices may encounter sufficient…
Mar 4, 2009Not classifiedZ-1621-2009Nonconforming Material/ComponentIn April 2007, Boston Scientific CRM communicated with physicians regarding the potential for reduced Elective Replacement Indicator (ERI) to Battery End of Life (EOL) time in a subset of implantable defibrillators due to degradation of a…
Jan 7, 2008Not classifiedZ-1514-2008Device DesignRepetitive mechanical stress applied to the titanium case can induce component damage and device malfunction in certain ICD and CRT-D devices implanted subpectorally with the serial number facing the ribs. Devices implanted subcutaneously…
Jan 7, 2008Not classifiedZ-1513-2008Device DesignRepetitive mechanical stress applied to the titanium case can induce component damage and device malfunction in certain ICD and CRT-D devices implanted subpectorally with the serial number facing the ribs. Devices implanted subcutaneously…
Mar 16, 2007Not classifiedZ-0051-2008Device DesignAtypical Charge Time Behavior: 1)-End of Life (EOL) or ERI is displayed during mid-life (typically 24-48 months), even though battery capacity remains available. 2)- Extended ERI charge time limits: Charge times during mid-life may remain…
Mar 16, 2007Not classifiedZ-0050-2008Device DesignAtypical Charge Time Behavior: 1)-End of Life (EOL) or ERI is displayed during mid-life (typically 24-48 months), even though battery capacity remains available. 2)- Extended ERI charge time limits: Charge times during mid-life may remain…
Jun 24, 2006Not classifiedZ-1297-06OtherGuidant has determined that low-voltage capacitors from a single component supplier may malfunction in a manner that can lead to premature battery depletion in the affected devices.
May 15, 2006Not classifiedZ-1292-06OtherGuidant has identified a pattern of premature battery depletion in devices manufactured with a single lot of capacitors from a single supplier. If function of this low voltage capacitor is compromised, the battery may deplete prematurely.
Apr 7, 2006Not classifiedZ-0828-06Other77 distributed devices may have a cracked layer of insulation in a flexible hybrid circuit, leading to internal shorting during charging such as during routine capacitor formation or for defibrillation. This would likely render the device…

No ancestor or descendant has a Class I recall.

How does its category compare?

All LWS devices

LWS Implantable Cardioverter Defibrillator (Non-Crt): FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LWS

97in 2025

2007: 702008: 942009: 802010: 1192011: 1362012: 1292013: 1482014: 1682015: 1002016: 1452017: 1622018: 1422019: 982020: 1252021: 1112022: 912023: 1012024: 1082025: 972026 (partial year): 93
2007high 168 (2014) · low 70 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LWS

Too few LWS 510(k)s in any year since 2007 to give a median.

Recalls per year · LWS

2in 2025

2007: 42008: 32009: 82010: 62011: 82012: 52013: 12015: 2, 1 Class I2016: 2, 2 Class I2017: 7, 5 Class I2018: 12020: 6, 3 Class I2022: 32023: 22025: 2, 1 Class I
2007high 8 (2009) · low 0 (2014)2026

Class I (12 since 2007) Class II/III not classified

Adverse event reports per year · LWS

24,303in 2025

2007: 10,3302008: 11,3272009: 13,7682010: 17,3212011: 15,0972012: 12,8202013: 17,2422014: 18,5942015: 14,0402016: 16,6382017: 18,4542018: 17,2282019: 16,1842020: 16,9092021: 19,6542022: 23,4542023: 28,7012024: 25,2022025: 24,303
2007high 28,701 (2023) · low 10,330 (2007)2025

MAUDE reports naming product code LWS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200770—410,330
200894—311,327
200980—813,768
2010119—617,321
2011136—815,097
2012129—512,820
2013148—117,242
2014168—018,594
2015100—2 (1)14,040
2016145—2 (2)16,638
2017162—7 (5)18,454
2018142—117,228
201998—016,184
2020125—6 (3)16,909
2021111—019,654
202291—323,454
2023101—228,701
2024108—025,202
202597—2 (1)24,303
202693—0—
Cite this page

510(k) Report. “P960040: Ventak AV Aicd Ventak Prizm DR/VR, Vitality, Cofient, and Teligen Automatic Implantable Cardioveter Defibrillator System.” Lineage record from FDA PMA data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/P960040

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Page last updated Nov 3, 2026